JHL0324S05

JHL03 Series
DC-DC Converter
xxx Series
www.xppower.com
1
Models & Ratings
Notes
1. Input current measured at nominal input voltage.
2. Input current measured at lowest input voltage.
3. Maximum capacitive load is per output.
4. Typical values.
International Medical Approvals
4000 VAC Reinforced Insulation
Meets IEC60601-1, 3rd Edition
2 MOPP Isolation at 250 VAC
2 µA Patient Leakage Current
DIP24 Package
EN55011 Level A With No External Components
3 Year Warranty
D
imensions:
J
HL03:
1
.25 x 0.80 x 0.40” (31.15 x 20.32 x 10.20 mm)
3 Watts
Input Voltage Output Voltage Output Current
Input Current
Maximum
Capacitive Load
(
3)
Efficiency
(
4)
Model Number
No Load
(
1)
Full Load
(
2)
1
0-17 V
5.0 V 600 mA 56 mA 325 mA 720 µF 76% JHL0312S05
12.0 V 250 mA 72 mA 316 mA 300 µF 78% JHL0312S12
1
5.0 V
2
00 mA
6
7 mA
3
15 mA
2
40 µF
7
8%
J
HL0312S15
±
12.0 V
±
125 mA
4
3 mA
3
04 mA
±
140 µF
8
1%
J
HL0312D12
±15.0 V ±100 mA 56 mA 303 mA ±120 µF 80% JHL0312D15
20-30 V
5.0 V 600 mA 38 mA 167 mA 720 µF 74% JHL0324S05
12.0 V 250 mA 37 mA 165 mA 300 µF 78% JHL0324S12
15.0 V 200 mA 23 mA 146 mA 240 µF 82% JHL0324S15
±12.0 V ±125 mA 29 mA 150 mA ±140 µF 80% JHL0324D12
±15.0 V ±100 mA 42 mA 166 mA ±120 µF 80% JHL0324D15
1. All dimensions are in inches (mm)
2. Weight: 0.04 lbs (20 g) approx.
3. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
4. Pin pitch tolerance: ±0.014 (±0.35)
5. Case tolerance: ±0.02 (±0.5)
Notes
Mechanical Details
Pin Connections
Pin Single Dual
2 -Vin -Vin
3 -Vin
-Vin
9 No Pin
Common
10 Trim
Trim
11 No Pin
-Vout
14 +Vout +Vout
16 -Vout Common
22 +Vin +Vin
23 +Vin +Vin
Bottom View
1.25
(31.75)
0.90
(22.86)
0.80
(20.32)
0.60
(15.24)
0.18
(4.57)
0.10
(2.54)
0.40 (10.20)
0.12 (3.05) Min.
0.20
(5.08)
2
10
3
22
1614
23
9
11
Notes:
1. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
2. Pin pitch tolerance: ±0.014 (±0.35)
3. Case Tolerance: ±0.02 (±0.5)
JHL03 Series
www.xppower.com
2
DC-DC Converter
Input
Output
Characteristic Minimum Typical Maximum Units Notes & Conditions
O
utput Voltage 5 30 V See Models and Ratings table
O
utput Voltage Trim ±10 % Via external resistors, see Application Notes
Initial Set Accuracy
±
1 % on V1
±2 % on V2 of dual output models
Minimum Load 0ANo minimum load required
Start Up Delay 5 ms
Start Up Rise Time 2 ms
Line Regulation ±0.3 %
Load Regulation ±1 % 0 - 100% load
Cross Regulation ±4 %
On dual output models with one output set to 50% load and
the other varied from 10% to 100% load (D05 20% to 100%)
Transient Response 4 % deviation
Recovery to within 1% in <500 µs for a 50% load change
at 0.25 A/µs rate
Ripple & Noise 1 % pk-pk 20 MHz bandwidth
Short Circuit Protection Trip & Restart (hiccup mode), auto recovery
Overload Protection 120 200 % Trip & Restart (hiccup mode)
Overvoltage Protection 115 140 %
Temperature Coefficient 0.03 %/ ˚C
C
haracteristic
M
inimum
T
ypical
M
aximum
U
nits
N
otes & Conditions
I
nput Voltage Range 10 17 VDC 12 V nominal
20 30 VDC 24 V nominal
Input Current See Models and Ratings table
Inrush Current 25 A At 30 VDC input
I
nput Filter Pi type
P
atient Leakage Current 2 µA
U
ndervoltage Lockout
On at >8.8 V. Off <8.3 V 12 V models
On at >17.5 V. Off <17.0 V 24 V models
Input Surge
25 VDC 12 V models for 3 s
5
0 VDC 24 V models for 3 s
General
Characteristic Minimum Typical Maximum Units Notes & Conditions
Efficiency 80 % See Models and Ratings table
Isolation
4000 VAC
For 1 min. Double/reinforced with a working voltage of 250
VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000
VAC for 10 ms in accordance with IEC60664-1
Patient Leakage Current 2 µA
Input to Output Capacitance 20 pF
Switching Frequency 250 kHz
Power Density 7.5 W/in
3
Mean Time Between Failure >1 MHrs MIL-HDBK-217F, +25 °C GB
Weight 0.04 (20.0) lb (g)
Environmental
Characteristic Minimum Typical Maximum Units Notes & Conditions
Operating Temperature -20 +80 °C See derating curve
Storage Temperature -40 +100 °C
Case Temperature +100 °C
Humidity 5 90 %RH Non-condensing
Cooling Natural convection
Shock ±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47
Vibration 10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6
0
-20 -10 0 10 20 30 40 50 60 70 80 90
20
40
60
80
100
Ambient Temperature (°C)
Load (%)
JHL03 Series
www.xppower.com
DC-DC Converter
3
10 Mar 16
EMC: Immunity
Phenomenon Standard Test Level Criteria Notes & Conditions
I
mmunity IEC60601-1-2 Ed 4.0: 2014 As Below
ESD Immunity EN61000-4-2 ±8 kv Contact, ±15 kv Air A
Radiated Immunity EN61000-4-3 10 V/m A
8
0 MHz - 2.7 GHz plus discrete communication
proximity field frequencies
EFT/Burst EN61000-4-4 2A
Surges EN61000-4-5 1A
C
onducted Immunity EN61000-4-6 3 Vm A
M
agnetic Fields EN61000-4-8 30 A/m A
S
afety Approvals ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition
EMC: Emissions
P
henomenon
S
tandard
T
est Level
N
otes & Conditions
Conducted EN55011 Level A
R
adiated EN55011 Level A
Application Notes
For 5 V output:
Trim +10%, R = 3.4 k typical
Trim -10%, R = 1.1 k typical
For 12 V output:
Trim +10%, R = 5.9 k typical
Trim -10%, R = 11.3 k typical
For 15 V output:
Trim +10%, R = 8.4 k typical
Trim -10%, R = 10.4 k typical
For ±12 V output:
Trim +10%, R = 12.8 k typical
Trim -10%, R = 9.5 k typical
For ±15 V output:
Trim +10%, R = 18 k typical
Trim -10%, R = 14.8 k typical
Trim
Trim Up
-Vout
+ Vout
Trim Down
Trim
1
2
3
External Output Trim
Derating Curve
Safety Approvals
Safety Agency Safety Standard Notes & Conditions
CB Report IEC60601-1 Including Risk Management Medical
UL ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008 Medical
EN EN60601-1 Medical

JHL0324S05

Mfr. #:
Manufacturer:
Description:
DC DC CONVERTER 5V 3W