5 INCH-D-DO

Digital Output Pressure Sensors
0404
0404
04
16-Mar-03
Command Summary Table:
Command Description Response
RA<cr> Read Accuracy String RA=[Accuracy String]<cr>
RC<cr> Read Captured Pressure RC=[hhhh]<sp>[eeee]<cr>
RH<cr> Read High Resolution Pressure RH=[hhhh]<sp>[eeee]<cr>
RL<cr> Read Low Resolution Pressure RL[llll]<sp>[eeee]<cr>
RM<cr> Read Model RM=[Model String]<cr>
RR<cr> Read Pressure Range RR=[pressure Range String]<cr>
RS<cr> Read Serial Number RS=[S/N String]<cr>
RT<cr> Read Temperature RT=[Temperature Range String]<cr>
U[S/N String][Command]<cr> Unique Command For Matching S/N U[S/N String],sp>[Response String]
For Non-matching S/N {null}
WC<cr> Capture Pressure {null}
Notations:
<> indicates a single ascii character
[] indicates an ascii string
{} text within the braces describes the response (this is essentially a comment)
“” text within quotes represents a literal ascii text string
Definitions:
Term Name Description
<cr> Carriage Return ascii Carrage Return. This is a command/response delimiter
<sp> Space ascii Space Character
[Accuracy String] Accuracy String Part accuracy string. Given in % full scale output.
Example: 0.250 %FSO
Notes:
1.) There is a space between the numeric accuracy “0.250”
and units “%FSO.”
2.) FSO stands for Full Scale Output (full scale output is
determined by the Pressure Range String.)
[hhhh] High Resolution Output This is a four character ascii string representing a
hexidecimal value.
Example: 3F7C
Represents an output count of 16,252
Note: the output is forced to “8000” upon an error.
except error bit 8, see error bit codes.
[llll] Low Resolution Output This is a four character ascii string representing a
hexidecimal value.
Example: 1D58
Represents an output count of 7,512
Note: the output is forced to “8000” upon an error.
except error bit 8, see error bit codes.
[Model String] Model String Part Model as given in the data sheet (also order number.)
The general model syntax is
[Full Scale Pressure]<sp>[Pressure Units]-[Pressure Model]-DO
Example: 100 PSI-D-DO
Where:
Full Scale Pressure=100
Pressure Units=PSI (inH2O, mbar or mmHg available)
DO represents Digital Output
Notes:
1.) Exception to this syntax is the Barometer.
2.) Custom models may be different.
a 16035 Vineyard Blvd. Morgan Hill, CA 95037 p 408 225 4314 f 408 225 2079 e www.allsensors.com
all sensorsall sensors
all sensorsall sensors
all sensors
All Sensors DS-0012 Rev A
All Sensors reserves the right to make changes to any products herein. All Sensors does not assume any liability arising out of the application or use of any product or circuit
described herein, neither does it convey any license under its patent rights nor the rights of others.
® MICROWIRE/PLUS IS A REGISTERED TRADEMARK OF NATIONAL SEMICONDUCTOR.
Definitions:
Term Name Description
[Pressure Range String] Pressure Range String This is the compensated pressure range of the part.
Syntax: [low limit]<sp>”to”<sp>[high limit]<sp>[units}[mode]
Esample: 20 to 32 mmHgA
Where:
Low limit = 20
High limit = 32
Units = mmHg
Mode = A (absolute pressure) (D differential and G gage
pressure also available.)
[Response String] A Fully Formed Response String Example:RL=1E43<sp>0000<cr>
[S/N String] Serial Number String YMDD-NN-BSPP (12 character String)
Where:
Y : Year (0~9)
M : Month (A~M, excluding I)
DD: Day of Month
NN: Lot (lot sequence for a given day)
B : Lot Batch# (A~Z)
S : Test Oven Slot# (1~5)
P : Position on Slot (1~15)
Example: 3D23-03-A103
April 23, 2003
3rd lot of the day
Batch A, Slot 1, Position 03
This allows traceability to original test data
[Temperature Range String] Temperature Range String This is the compensated temperature range of the part.
Syntax: [low limit]<sp>[high limit]<sp>”C”
Example: -20 to 85 C
-40 to 125 C and custom ranges also available.
[eeee] Error Codes The error codes are bits packed within a double byte.
The four character string is an ascii hex expression.
The error code bits are:
Bit 0 : Part not factory compensated
Bit 1 : Tdex Overflow
Bit 2 : Tdex Over-range
Bit 3 : Pdex Overflow
Bit 4 : Pdex Over-range
Bit 5 : PWL Overflow
Bit 6 : Scale Overflow
Bit 7 : High Resolution Overflow
Bit 8 : Pressure Output Limited to Specific Value
Bits 9 through 15 : Reserved
Example: 0100
Represents erro bit 8 set and the Pressure Output Limited
Notes:
Bit 0 : This should not appear if the part has been calibrated
Bits 1 thru 7 : indicate computational error when compensated
Bit 8: indicates that the pressure applied to the part exceeds the
range of the part and is limit to either the high or low limit
{null} Null Response No response from the part. In the event of a serial number
mismatch the part will not respond (to avoid buss contention.)
Month Codes
January : A
February : B
March : C
April : D
May : E
June : F
July : G
August : H
September : J
October : K
November : L
December : M

5 INCH-D-DO

Mfr. #:
Manufacturer:
Amphenol All Sensors
Description:
Board Mount Pressure Sensors Differential Digital A Package
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment: