JHM1512D05

JHM15 Series
DC-DC Converter
xxx Series
www.xppower.com
1
Models & Ratings
Notes
1. Input current measured at nominal input voltage.
2. Input current measured at lowest input voltage.
3. Maximum capacitive load is per output.
4. Typical values.
International Medical Approvals
4000 VAC Reinforced Insulation
Medical Approval, IEC60601-1, 3rd Edition
2 µA Patient Leakage Current
Compact 1 x 1.6” Footprint
EN55011 Level A With No External Components
3 Year Warranty
D
imensions:
J
HM15:
1
.60 x 1.00 x 0.40” (40.60 x 25.40 x 10.20 mm)
15 Watts
Input Voltage Output Voltage Output Current
Input Current
Maximum
Capacitive Load
(
3)
Efficiency
(
4)
Model Number
No Load
(
1)
Full Load
(
2)
9-18 V
5.0 V 3000 mA 9.2 mA 1930 mA 3000 µF 87% JHM1512S05
12.0 V 1250 mA 6.5 mA 1938 mA 1330 µF 86% JHM1512S12
1
5.0 V
1
000 mA
8
.0 mA
1
944 mA
1
000 µF
8
6%
J
HM1512S15
±
5.0 V
±
1500 mA
6
.6 mA
1
955 mA
±
1470 µF
8
4%
J
HM1512D05
±12.0 V ±625 mA 11.2 mA 1911 mA ±660 µF 87% JHM1512D12
±15.0 V ±500 mA 11.0 mA 1879 mA ±550 µF 88% JHM1512D15
18-36 V
5.0 V 3000 mA 5.6 mA 972 mA 3000 µF 86% JHM1524S05
12.0 V 1250 mA 6.1 mA 968 mA 1830 µF 85% JHM1524S12
15.0 V 1000 mA 6.4 mA 966 mA 1000 µF 87% JHM1524S15
±5.0 V ±1500 mA 5.4 mA 981 mA ±1470 µF 83% JHM1524D05
±12.0 V ±625 mA 7.3 mA 954 mA ±660 µF 87% JHM1524D12
±15.0 V ±500 mA 8.5 mA 943 mA ±550 µF 87% JHM1524D15
BOTTOM VIEW
126
54 3
0.200 (5.08)
0.100 (2.54)
0.100 (2.54)
0.40 (10.2) 0.22 (5.6) min
ø 0.040 (1.02)
1.60 (40.6)
1.40 (35.56)
1.00 (25.4)
0.200 (5.08)
0.400 (10.16)
0.400 (10.16) 0.400 (10.16)
1. All dimensions are in inches (mm)
2. Weight: 0.04 lbs (20 g) approx.
3. Pin diameter: 0.04 ±0.002 (1.02 ±0.05)
4. Pin pitch tolerance: ±0.01 (±0.25)
5. Case tolerance: ±0.02 (±0.5)
Notes
Mechanical Details
Pin Connections
Pin Single Dual
1 +Vin +Vin
2 -Vin
-Vin
3 +Vout
+Vout
4 -Vout
Common
5 Trim
-Vout
6 No Pin
No Pin
JHM15 Series
www.xppower.com
2
DC-DC Converter
Input
Output
Characteristic Minimum Typical Maximum Units Notes & Conditions
Output Voltage 5 15 V See Models and Ratings table
Output Voltage Trim ±10 % Via external resistors, see Application Notes
Initial Set Accuracy
±1 % on V1
±
2 % on V2 of dual output models
M
inimum Load 0ANo minimum load required
S
tart Up Delay 5 ms
Start Up Rise Time 2 ms
Line Regulation ±0.3 %
Load Regulation
±2 % 0 - 10% load
±1 % 10 - 100% load
Cross Regulation ±4 %
On dual output models with one output set to 50% load and
the other varied from 10% to 100% load (D05 20% to 100%)
Transient Response 4 % deviation
Recovery to within 1% in <500 µs for a 50% load change
at 0.25 A/µs rate
Ripple & Noise 1 % pk-pk 20 MHz bandwidth
Short Circuit Protection Trip & Restart (hiccup mode), auto recovery
Overload Protection 120 200 % Trip & Restart (hiccup mode)
Overvoltage Protection 115 140 % Non latching, auto recovery
Temperature Coefficient 0.03 %/ ˚C
C
haracteristic
M
inimum
T
ypical
M
aximum
U
nits
N
otes & Conditions
I
nput Voltage Range 9 18 VDC 12 V nominal
18 36 VDC 24 V nominal
Input Current See Models and Ratings table
Inrush Current 20 A at 36 V
I
nput Filter Pi type
P
atient Leakage Current 2 µA
U
ndervoltage Lockout
On at >8.8 V. Off <8.3 V 12 V models
On at >17.5 V. Off <17.0 V 24 V models
Input Surge
25 VDC 12 V models for 3 s
5
0 VDC 24 V models for 3 s
General
Characteristic Minimum Typical Maximum Units Notes & Conditions
Efficiency 86 % See Models and Ratings table
Isolation
4000 VAC
For 1 min. Double/reinforced with a working voltage of 250
VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000
VAC for 10 ms in accordance with IEC60664-1
Input to Output Capacitance 20 pF
Switching Frequency 250 kHz
Power Density 23 W/in
3
Mean Time Between Failure >1 MHrs MIL-HDBK-217F, +25 °C GB
Weight 0.04 (20.0) lb (g)
Environmental
Characteristic Minimum Typical Maximum Units Notes & Conditions
Operating Temperature -40 +80 °C See derating curve
Storage Temperature -55 +100 °C
Case Temperature +100 °C
Humidity 5 90 %RH Non-condensing
Cooling Natural convection
Shock ±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47
Vibration 10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6
0
-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90
20
40
60
80
100
Ambient Temperature (°C)
Load (%)
JHM15 Series
www.xppower.com
DC-DC Converter
3
28 Mar 17
EMC: Immunity
Phenomenon Standard Test Level Criteria Notes & Conditions
Immunity IEC60601-1-2, EN61204-3
E
SD Immunity EN61000-4-2 2A
Radiated Immunity EN61000-4-3 10 V/m A
EFT/Burst EN61000-4-4 2A
Surges EN61000-4-5 1A
C
onducted Immunity EN61000-4-6 10 Vm A
M
agnetic Fields EN61000-4-8 3 A/m A
S
afety Approvals ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition
EMC: Emissions
P
henomenon
S
tandard
T
est Level
N
otes & Conditions
Conducted EN55011 & EN55022 Level A
R
adiated EN55011 & EN55022 Level A
Application Notes
For 5 V output:
Trim +10%, R = 3.4 k typical
Trim -10%, R = 1.1 k typical
For 12 V output:
Trim +10%, R = 5.9 k typical
Trim -10%, R = 11.3 k typical
For 15 V output:
Trim +10%, R = 8.4 k typical
Trim -10%, R = 10.4 k typical
For ±5 V output:
Trim +10%, R = 12.0 k typical
Trim -10%, R = 8.0 k typical
For ±12 V output:
Trim +10%, R = 12.8 k typical
Trim -10%, R = 9.5 k typical
For ±15 V output:
Trim +10%, R = 18 k typical
Trim -10%, R = 14.8 k typical
Trim
Trim Up
-Vout
+ Vout
Trim Down
Trim
External Output Trim
Derating Curve
Safety Approvals
Safety Agency Safety Standard Notes & Conditions
CB Report IEC60601-1 Ed 3 Including Risk Management Medical
UL ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008 Medical

JHM1512D05

Mfr. #:
Manufacturer:
Description:
DC DC CONVERTER +/-5V 15W
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet