561-800-526-2324 • www.us.tdk-lambda.com/lp/ Disclaimer: us.tdk-lambda.com/lp/legal.htm
DTM110-C Series
Features
u CEC, EISA
(1)
& ErP Stage 2 Compliant
u Wide Range AC Input
u Medical Certifi cations
u > 87% Average Effi ciency
90W to 110W Medical AC-DC External Power Supplies
Model DTM110PWxxx-C
Input Voltage range VAC 90 - 264VAC (47 - 63Hz)
Inrush Current A <100A at 230VAC input, 25°C ambient cold start
Input Current A 2A at 90VAC, Full Load
Power Factor - Meets EN61000-3-2. >0.9 PF
Leakage Current (264VAC 60Hz) µA <300µA
Hold Up Time (Typ) ms >10ms
Temperature Coeffi cient %/°C ±0.05%/°C
Voltage Accuracy % ±1%
Adjustment Range V None
Minimum Load A None
Total Regulation % 12V to 15V Models:±5%, >15V models ±3%
Ripple & Noise % 1%
Short Circuit Protection - <160% - hiccup mode
Overvoltage Protection V 110 - 150% of nominal (Cycle input power to reset)
Effi ciency % >87% average effi ciency
Operating Temperature °C 0 to +40°C
Storage Temperature °C -10 to +70°C
Humidity (non condensing) %RH 10 - 90%RH
Cooling - Convection
Withstand Voltage VAC Input to Ground 1.5kVAC, Input to Output 4kVAC (Output is connected to Ground)
Vibration (non operating) - 23.52m/s
2
(10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
Shock - < 196.1 m/s
2
(20G)
Safety Agency Approvals - UL60601-1, EN60601-1, IEC60601-1
California Energy Commission (CEC) - See model selector
EISA - See model selector
Offl oad Power Consumption W < 0.5W
Conducted & Radiated EMI - EN60601-1-1-2, FCC Part 18 Class B
Immunity - EN55011:2007+A2:2007
Weight (Typ) g 605g
Size (WxLxH) Inches 2.55 x 6.7 x 1.51”
Cable Length mm 1050mm
AC Input Connector - IEC 320-C14 (Accepts IEC 320-C13)
Output Connector - Kycon KPPx-4P or equivalent Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Warranty yrs Two Years
Specifi cations
(1) EISA - Energy Independence and Security Act of 2007
Key Market Segments & Applications