ME40A1202B01

ME40
40W Single Output External Power
Medical Series
www.slpower.com ME40 Rev 2.4 15-NOV-16 1
Features
Meets UL/EN/IEC60601-1-2, 4th edition for EMC
*
Approved to EN/IEC/UL60601-1, 3rd edition
2 MOPP Input to Output Isolation
Meets DoE Efficiency Level VI Requirements
No load input power
Average Efficiency
Up to 40W of AC-DC Power
Universal Input 90-264Vac Input Range
Desktop and Wall-Plug versions
Meets EN55011/CISPR11, FCC Part 15.109 Class B
Conducted & Radiated Emissions, with 6db margin
IP22 Rated Enclosure
E-cap life of >8 years
>1,000,000 hours MTBF
3 Year Warranty
Description
A high performance AC to DC external power supply family designed for medical applications. The ME40A Medical Series low power
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels which satisfy the 2 MOPP
requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC
*
. The ME40A Series models will operate at universal input
range of 90 to 264Vac over the wide temperature range of -20°C to +70°C, delivering full rated output power up to +40°C and
applicable output power derating at 70°C. These models are available in desktop and wall-plug versions, include an IP22 rating per
IEC60529 for the enclosure, and output cable terminated at a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model Output Output Ripple & Line Load
Overvoltage Output Input
Number Volts Current Power
Noise¹
Regulation Regulation
Trip Range Connector Configuration
ME40A0503F01 5.0V 5.00A 25W 100mV pk-pk
±1% ±5%
120%-150%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
ME40A0903F01 9.0V 4.00A 36W 100mV pk-pk
±1% ±5%
120%-150%
ME40A1203F01 12.0V 3.40A 40W 120mV pk-pk
±1% ±5%
120%-150%
ME40A1803F01 18.0V 2.22A 40W 180mV pk-pk
±1% ±5%
120%-150%
ME40A2403F01 24.0V 1.70A 40W 240mV pk-pk
±1% ±5%
120%-140%
ME40A0503N01 5.0V 5.00A 25W 100mV pk-pk
±1% ±5%
120%-150%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
ME40A0903N01 9.0V 4.00A 36W 100mV pk-pk
±1% ±5%
120%-150%
ME40A1203N01 12.0V 3.40A 40W 120mV pk-pk
±1% ±5%
120%-150%
ME40A1803N01 18.0V 2.22A 40W 180mV pk-pk
±1% ±5%
120%-150%
ME40A2403N01 24.0V 1.70A 40W 240mV pk-pk
±1% ±5%
120%-140%
ME40A0503Q01 5.0V 5.00A 25W 100mV pk-pk
±1% ±5%
120%-150%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
ME40A0903Q01 9.0V 4.00A 36W 100mV pk-pk
±1% ±5%
120%-150%
ME40A1203Q01 12.0V 3.40A 40W 120mV pk-pk
±1% ±5%
120%-150%
ME40A1803Q01 18.0V 2.22A 40W 180mV pk-pk
±1% ±5%
120%-150%
ME40A2403Q01 24.0V 1.70A 40W 240mV pk-pk
±1% ±5%
120%-140%
ME40A0503B01 5.0V 5.00A 25W 100mV pk-pk
±1% ±5%
120%-150%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades²
ME40A0903B01 9.0V 4.00A 36W 100mV pk-pk
±1% ±5%
120%-150%
ME40A1203B01 12.0V 3.40A 40W 120mV pk-pk
±1% ±5%
120%-150%
ME40A1803B01 18.0V 2.22A 40W 180mV pk-pk
±1% ±5%
120%-150%
ME40A2403B01 24.0V 1.70A 40W 240mV pk-pk
±1% ±5%
120%-140%
ME40A0503C01 5.0V 5.00A 25W 100mV pk-pk
±1% ±5%
120%-150%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North
American Blades³
ME40A0903C01 9.0V 4.00A 36W 100mV pk-pk
±1% ±5%
120%-150%
ME40A1203C01 12.0V 3.40A 40W 120mV pk-pk
±1% ±5%
120%-150%
ME40A1803C01 18.0V 2.22A 40W 180mV pk-pk
±1% ±5%
120%-150%
ME40A2403C01 24.0V 1.70A 40W 240mV pk-pk
±1% ±5%
120%-140%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Standard models are fitted with North American blades. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE40B1203F01).
ME40
40W Single Output External Power
Medical Series
www.slpower.com ME40 Rev 2.4 15-NOV-16 2
General Specifications
AC Input
100-240Vac, ±10%, 47-63Hz, 1
Turn On Time
Less than 700mS @115Vac, full load
Input Current
115Vac: 1.2A, 230Vac: 01.6A
Hold-up Time
20mS min., at full Load, 100Vac input
Inrush Current
264Vac, cold start: will not exceed 40A
Overtemperature
Protection
Will shutdown upon an over-temperature
condition, auto-recovery.
Input Fuses
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
Overload Protection
130 to 180% of rating, Hiccup Mode
Earth Leakage
Current
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Short Circuit
Protection
Hiccup Mode, auto recovery.
Efficiency
>87%, typical
Overvoltage
Protection
Hiccup mode. See models chart for trip
ranges.
Output Power
40W continuous – See models chart for
specific voltage model ratings.
Isolation
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
No Load Input
Power
<0.1W per DoE Efficiency Level VI
Requirements
Safety Standards
EN/IEC/UL60601-1, 3rd edition and
Ripple and Noise
See models chart on pg 1.
Operating
Temperature
-20°C to +70°C
Output Voltage
See models chart on pg 1.
Temperature
Derating
See derating chart
Transient
Response
500
µ
s response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, i/t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Storage Temperature
-40°C to +85°C
Regulation
See models chart on pg 1.
Altitude
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Drop Test
1.4m from table top to wooden platform, 6
faces.
Relative Humidity
5% to 95%, non-condensing
Vibration
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
Dimensions
See outline drawings
MTBF
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
Weight
250g
E-Cap Life
>8 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
ME40
40W Single Output External Power
Medical Series
www.slpower.com ME40 Rev 2.4 15-NOV-16 3
EMI/EMC Compliance
Conducted Emissions:
EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4
th
Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4
th
Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4
th
Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4
th
Edition requirements.
Conducted Disturbances induced by RF
Fields
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4
th
Edition, Table 5.
Rated Power frequency magnetic fields
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Mechanical Drawing
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
Notes: 1. All dimensions in mm.
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
9
6
95.5
55
54
1450±50
1450±50
33.6

ME40A1202B01

Mfr. #:
Manufacturer:
Description:
Wall Mount AC Adapters Class II/Wall/Change 12Vdc/40W/Medical
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment: