GPM40-12G

GPC40 Commercial/GPM40 Medical
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated output load: 1.3 A
Hold-Up Time
20 ms minimum from loss of ac input at full load, nominal line (115 Vac).
Output Power
40 W continuous, 50 W peak. Peak ratings are for 60 s maximum duration,
10% duty cycle. During peak load condition, output regulation may exceed
total regulation limits.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit on outputs 1 & 2; foldback
type on output 3. Recovery after fault is automatic. See output ratings chart
for additional notes or conditions.
Overvoltage Protection.
Main outputs: 124% ± 12%.
Eciency
70% at full rated load, nominal input voltage, depending on model and load
distribution.
Turn-on Time
Less than 1 second at 120 Vac, 25°C (inversely proportional to input voltage
and thermistor temperature).
Input Protection
Internal ac fuse provided. Designed to blow only if a catastrophic failure oc-
curs in the unit.
Inrush Current
Inrush is limited by internal thermistors. Inrush at 240 Vac under cold start
conditions will not exceed 34 A.
Temperature Coecient
0.03%/°C typical on all outputs.
Environmental
Designed for 0 to 50°C operation at full rated output power; derate output
current and total output power by 2.5% per °C above 50°C.
40 WATT GLOBAL PERFORMANCE SWITCHERS
Wide-range ac input 85-264 Vac
• 2-year warranty
• Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11
Class B (Medical models)
• Single and multiple outputs
• Commercial Approved to UL1950,
EN60950 E3:2000 and CSA-C22.2 No. 234
• Medical Approved to UL2601-1, EN60601-1 and
CSA22.2 No. 601
• Single and multiple outputs
• RoHS compliant models available (G sux)
GLOBAL PERFORMANCE SWITCHERS
All specications are typical at nominal input, full load at 25
O
C unless otherwise stated
Summary:
D C P o w e r S u p p l i e s
SPECIFICATIONS
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Line Frequency Harmonics
EN55022 Class B; FCC Class B
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kH
EN61000-4-5, 1 kV di., 2 kV com.
EN61000-3-2
• marked to LVD
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, dierential mode. Measured with noise
probe directly across output terminals of the power supply.
Transient Response
Main output—500 μs typical response time for return to within 0.5% of nal value
for a 50% load step change. ∆i/ ∆t<0.2 A/μs. Maximum voltage deviation is 3.5%.
Startup/ shutdown overshoot less than 3%.
Remote Sense
Provided as a standard feature on single-output models.
Voltage Adjustment
Built-in potentiometer adjusts main output voltage ±5%.
EMI/EMC Compliance
All models include built-in EMI ltering to meet the
following emissions requirements:
Commercial Safety
Approved to UL1950, CSA22.2 No. 950, IEC950, EN60950. All dc outputs are SELV
under normal and single fault conditions.
Medical Leakage Current
The maximum leakage current is as follows:
Test Condition Normal Single Fault
120 Vac @ 60 Hz input 25 μA 35 μA
264 Vac @ 60 Hz input 40 μA 60 μA
Medical Safety
Approved to UL2601-1, CSA22.2 No. 601 Level 3 and EN60601-1. UL le E116994;
CSA #LR46516. The output(s) are intended for safety earthed Signal Output
and Intermediate Circuits only. The output(s) are not acceptable for patient con-
nection without additional isolation. All dc outputs are SELV under normal and
single fault conditions.
Note: For 48 volt model, GPC40-48 or GPM40-48, contact factory for availability.
A. Total regulation is dened as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. To maintain these regulation conditions, the 5.1V current must be at least 1/4 of V2 and not greater than 5 times the V2 current.
C. Requires +5.1V to be adjusted within ±1% with at least a 0.4A load to maintain regulation on this output.
D. Add “G” sux to model number for RoHS compliant model.
3
3
Commercial
Model
Medical
Model
Output No. Output Output
Minimum
Output
Maximum
Output
Peak
Noise P-P Total
Regulation
Notes
GPC40A GPM40A 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +12 V 0 A 2 A 3 A 120 mV 5% B,C
3 -12 V 0 A 0.4 A 0.7 A 120 mV 3% C
GPC40B GPM40B 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +15 V 0 A 2 A 3 A 150 mV 5% B,C
3 -15 V 0 A 0.4 A 0.7 A 150 mV 3% C
GPC40D GPM40D 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +24 V 0.A 1 A 1.5 A 240 mV 5% B,C
3 -12 V 0.A 0.4 A 0.7 A 120 mV 3% C
GPC40-3.3 GPM40-3.3 1 3.3 V 0 8 A 10 A 35 mV 2%
GPC40-5 GPM40-5 1 5 V 0 8 A 10 A 50 mV 2%
GPC40-9 GPM40-9 1 9 V 0 A 4.5 A 6 A 90mV 2%
GPC40-12 GPM40-12 1 12 V 0 A 3.3 A 4.2 A 120 mV 2%
GPC40-15 GPM40-15 1 15 V 0 A 2.7 A 3.3 A 150 mV 2%
GPC40-24 GPM40-24 1 24 V 0 A 1.7 A 2.1 A 240 mV 2%
GPC40-28 GPM40-28 1 28 V 0 A 1.4 A 1.8 A 280 mV 2%
SL Power Electronics Corp, 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com. Rev. 1/07.
Data Sheet © 2007 SL Power Electronics Corp. The information and specications contained in this data sheet are believed to be correct at time of publication.
However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specications are subject to change without notice.
GPC40/GPM40 MECHANICAL SPECIFICATIONS
ENVIRONMENTAL SPECIFICA-
TIONS
OPERATING NON-OPERATING
Temperature (A) See individual specs. -40 to +85°C
Humidity (A) 0 to 95% RH 0 to 95% RH
Shock (B) 20 g
pk
40 g
pk
Altitude -500 to 10,000 ft -500 to 40,000 ft
Vibration (C) 1.5 g
rms’
0.003 g
2
/Hz
5 g
rms’
0.026 g
2
/Hz
A. Units should be allowed to warm up/operate under non-condensing
onditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction,
3 orthogonal axes, total 6 shocks.
C. Random vibration—10 to 2000Hz, 6dB/octave roll-o from 350 to
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and
1 hr./axis non-operating.

GPM40-12G

Mfr. #:
Manufacturer:
Description:
Switching Power Supplies 40W 12V @ 3.3A