5360

28
5390 536053705380
Basic Functions
Basic Accuracy 0.025% 0.05% 0.1%
Voltage (Best Resolution) 750VAC/1000VDC 750VAC/1000VDC
(10µV) (100µV)
Current (Best Resolution) 10AAC/DC 10AAC/DC
(10nA) (1µA)
Frequency √√
Temperature Probe
Duty Cycle √√
Event Counter
Backlight √√
Specifications
The 5300 Series handheld DMM is the most advanced digital mul-
timeter available in this price range. These meters incorporate a
special 50,000 count integrated circuit for high-accuracy (0.025%)
basic accuracy, high-reliability design. Measure AC Volts (True RMS
AC, AC + DC), Resistance, Capacitance and Frequency, in addition
to dB, Duty Cycle, Pulse Width, Power Line Disturbances and
Temperature. These meters also feature fuse and battery access
without breaking a calibration seal, great for ISO facilities. All mod-
els offer Peak Hold, Hold, Relative Mode and Backlight.
Precision 50,000 count instruments
0.025% Basic DCV accuracy (Models 5380 and 5390)
True RMS reading
Large LCD display provides easy readability, all test
parameters in clear view.
Best value of any DMM
CE safety approved
AC VOLTAGE
(Auto/Manual Ranging, True RMS)
Modes: AC or AC+DC, selectable
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
Ranges DC to 1 kHz to 4 kHz to 10 kHz to 30 kHz to Input
1kHz 4 kHz 10 kHz 30 kHz 50 kHz Resolution Impedance
500 mV 0.3%R+30D 1%R+3D 5%R+30D Not specified 10µV 10 M/100 pF
or 1 G
5V 0.3%R+30D 1%R+3D 2%R+30D 2%R+30D 3%R+30D 100 µV 11 M/100 pF
50V 0.3%R+30D 1%R+30D 2%R+30D 2%R+30D 3%R+30D1 mV 10 M/100 pF
500V 0.3%R+30D 1%R+30D 2%R+30D 2%R+30D 3%R+30D10 mV 10 M/100 pF
750V 0.3%R+30D Not specified 100 mV 10 M/100pF
50 kHz to 100 kHz (5V, 50 V, 500 V ranges): ±6% FS typical
Crest Factor: 6
models
5390 5380 5370
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Ranges DC to 1kHz 1 kHz to 5 kHz to 10kHz to 20 kHz to Input
5 kHz 10 kHz 20 kHz 30 kHz Resolution Impedance
500 mV 1%R+3D 4%R+3D Not specified 100µV 10 M/100 pF
5V 1%R+3D 1%R+3D 2%R+30D 2%R+3D 3%R+3D 1m V 11 M/100 pF
50V 1%R+3D 1%R+3D 2%R+3D 2%R+3D 3%R+3D 10m 10 M/100 pF
500V 1%R+3D 1%R+3D 2%R+3D 2%R+3D 3%R+3D 100 mV 10 M/100 pF
750V 1%R+3D Not specified 1 V 10 M/100pF
Crest Factor: 3
model
5360
Features
models
5390 5380 5370 5360
For your nearest distributor, call 800-462-9832 or www.bkprecision.com
29
Unless otherwise stated, accuracy specifications apply from 18°C to 28°C. Specifications stated as n%R+nD, mean ±(n% of reading +n least significant digits). Accuracy specifications stated as
n%FS, mean ±(n% of Full Scale)
DC VOLTAGE (Auto Manual Ranging) Input
Ranges Accuracy Resolution Impedance
500 mV 0.025%R+2D 0.025%R+2D 0.05%R+2D 0.1%R+2D 10 µV 10 Mor
1 G*
5V 0.025%R+2D 0.025%R+2D 0.025%R+2D 0.05%R+2D 0.1%R+2D 11 M
50V 0.025%R+2D 0.025%R+2D 0.05%R+2D 0.1%R+2D 1 mV 10 M
500V 0.025%R+2D 0.025%R+2D 0.05%R+2D 0.2%R+2D 10 mV 10 M
1000V 0.025%R+2D 0.025%R+2D 0.05%R+2D 0.3%R+2D 100 mV 10 M
*(10 M only on 5360)
AC/DC VOLTAGE PROTECTION
Overvoltage Protection ±1100 V DC + AC Peak
Transient Protection 6kV for 10 µs
DC CURRENT (Auto/Manual Ranging) Max Burden
Ranges Accuracy Resolution Voltage
500µA 0.2%R+5D Not Available 10 nA 700 mV
5 mA 0.2%R+2D 0.2%R+2D 100 nA 700 mV
50mA 0.05%R+2D 0.2%R+2D 1 µA 700 mV
500 mA 0.2%R+2D 0.2%R+2D 10 µA 1.5 V
10 A 0.5%R+2D 0.3%R+2D 1 mA 500 mV
Specifications
Common Features
Accessories
SUPPLIED:
Test Leads, Spare Fuses, Instruction Manual
OPTIONAL:
TL-130A General Purpose DMM Kit
TP-5300 Platinum Temperature Probe (for 5370)
AK-5350 Calibration Software (AK-5300 Hardware required)
AK-5300 Communications Hardware & LabView
®
drivers
models
5390 5380 5370 5360
AC CURRENT (Auto/Manual Ranging, True RMS) AC or AC + DC
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
1 kHz to 10 kHz to 20 kHz to Input
Ranges DC to 1kHz 10 kHz 20 kHz 30 kHz Resolution Impedance
500 µA 0.75%R+50D 1.0%R+50D 2.0%R+50D 3.5%R+50D 10 nA 700 mV
5 mA 0.6%R+50D 0.75R+50D 2.0%R+50D 3.0%R+50D 100 nA 700 mV
50 mA 0.6%R+50D 0.75%R+50D 2.0%R+50D 3.0%R+50D 1 µA 700 mV
500 mA 0.7%R+50D 0.75%R+50D Not specified 10 µA 1.5 V
10 A 1% of FS Not specified 1 mA 500 mV
Crest Factor: 6
models
5390 5380 5370
AC CURRENT
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Ranges DC to 5 kHz Resolution Max Burden Voltage
5 mA 1.0%R+3D 1 µA 700 mV
50 mA 1.0%R+3D 10 µA 700 mV
500 mA 1.0%R+3D 100 mA 1.5 V
10 A 1% of FS 10 mA 500 mV
Crest Factor: 3
model
5360
RESISTANCE (Auto/Manual Ranging)
Accuracy
Ranges Resolution Test Current
500 0.07%R+5D 0.1%R+5D 10 m 1 mA
5 k 0.07%R+2D 0.1%R+2D 100 m 100 µA
50 k 0.07%R+2D 0.1%R+2D 1 10 µA
500 k 0.07%R+2D 0.1%R+2D 10 1 µA
5 M 0.3%R+2D 0.3%R+2D 100 100 µA
50 M 1.0%R+2D 1.0%R+2D 1 k 10 µA
models
5390 5380 5370 5360
CAPACITANCE (Auto/Manual Ranging)
Max
Accuracy Measurement Measurement
Range Resolution Current TIme
50 nF 1.0%R+2D 1.0%R+2D 10 pF 100 nA 0.5 sec
500 nF 1.0%R+2D 1.0%R+2D 100 pF 1 µA 0.5 sec
5 pF 1.0%R+2D 1.0%R+2D 1 nF 10 µA 0.5 sec
50 µF 1.0%R+2D 1.0%R+2D 10 nF 100 µA 0.5 sec
500 µF 1.0%R+2D 1.0%R+2D 100 nF 1 mA 1.5 sec
5000 µF 1.0%R+2D 2.0%R+2D 1 µF 1 mA 3 sec/ mF
50 mF 1.0%R+2D 2.0%R+2D 10 µF 1mA 3 sec/ mF
FREQUENCY (Autoranging)
Range: 0.6 Hz to 500 kHz
Accuracy: ±(0.03%rdg+1 digit)
500 mV 5V to 500 V 750 V 10 A
Ranges (5390, 5370) 500 µA to 500 mA (5390, 5370)
0.6Hz to 5kHz 2% of FS 2% of FS 100 V 2 A
5 kHz to 50 kHz 5% of FS 5% of FS 250 V Not specified
50 kHz to 500 kHz Not specified 10% of FS Not specified
CONTINUITY - Threshold: 10 to 20
DIODE TEST - Measures forward voltage drop of diode 0 to 1.999V
dB MEASUREMENT - 5390 and 5370 - Measures value relative to a reference from 1 to 9999 , 5380 at 600
PEAK HOLD - Selection: Peak + or Peak- (surge or sag) Acq. Time: <1rms
DUTY CYCLE: 5390, 5370, 5360-Modes: Selectable %+ ,or %- Min
EVENT COUNTER (Model 5390) - Range: 0 to 99999 (Totalize)
MIN PULSE DURATION: 2 µs
PULSE WIDTH (Model 5390) Modes: Selectable +, - Min. Pulse Width: 20µs Max Pulse Period: 12.5 µs
TEMPERATURE (Model 5370) Using Model TP-5300 Probe (Optional)
RANGE: -200° to +800°C
Computer Interface: RS232 capability built-in. Requires AK-5300
GENERAL
Display: 50,000 count resolution LCD, 0.55" (14mm) digit height,
34 segment analog bargraph
Measurement Rate: 2/sec., 20/sec - bargraph
Power: 9V (NEDA 1604) Battery Life: 500 hrs
Dimensions: 1.57 x 3.23 x 7.44" (40 x 82 x 189mm)
Weight: 14 oz (397g)
models
5390 5380 5370 5360
models
5390 5380 5370 5360
Three Year Warranty

5360

Mfr. #:
Manufacturer:
B&K Precision
Description:
TC 5360 4 IN INSUL 90 DEGREE LIQ TI
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet