U5200
Industrial Pressure Transducer
SENSOR SOLUTIONS ///U5200
CONNECTION TYPES
Note: Transmitter of gage pressure type requires vent to atmosphere on the pressure reference side. This is accomplished via cable from the
transmitter (the end of the cable should be terminated to clean and dry area) or through the customer mating connector/cable assembly
which has internal vent path.
Suggested vented M12 mating connector P/N MB12FWAFF04ST-4 and MB12FWAFF04ST-3 at www.finecables.com for 0.157”~0.236”
and 0.236”~0.315” diameter cable respectively.
WEATHERPROOF
Note: Weatherproof ratings are met when the mating connectors are installed properly and the cable termination is to dry and clean area.
OUTPUTS
MATING HOUSING
P/N
MATING TERMINAL
P/N
RUBBER SEAL
P/N
Bayonet BAYONET PTIH-10-6P OR EQUIV PT06A-10-6S MIL-C-26482 - -
Packard 3-PIN METRI-PACK 150 12078090 12103881, QTY 3 -
4-WIRE,22 AWG, SHIELDED,
PVC JACKET, 105 DEGC
BINDER SERIES 713,
09 3431 77 04 OR EQUIV
4-POS FEMALE CONNECTOR - -
AMP/TE
AMP / TE 3-PIN ECONOSEAL
J SERIES
171630-1 (AWG 20~24)
171662-1 (AWG 16~20)
QTY 3
172746-1 (AWG 20~24)
172888-2 (AWG 16~20)
QTY 3
INDUSTRIAL STANDARD 9.4MM
FORM C
HIRSCHMANN 933 024-100,OR,
ATAM KD046000B7 (SEAL INCL.)
- HIRSCHMANN 730 185-002
FORM A DIN EN 175 301-803-A 18MM
HIRSCHMANN 931 969-100,OR,
ATAM KA245000B4 (SEAL INCL.)
- HIRSCHMANN 730 801-002
Sumitom o SUMITOMO 3-PIN HV040
8100-3067 (AWG 20~22)
8100-3068 (AWG 16~18)
QTY 3
7165-1075 (INS. DIA 1.1~1.6MM)
7176-0621 (INS. DIA 1.6~1.9MM)
7165-0622 (INS. DIA 1.8~2.2MM)
QTY 3
CONNECTION IP CODE
Bayonet IP67
Packard IP66
Cable IP67
1/2NPT CONDUIT IP67
M12 IP67
AMP/TE IP67
FORM C IP65
FORM A IP65
Sumitom o IP67
OUTPUT SIGNAL SUPPLY VOLTAGE
5 ± 0.25V
PROTECTED to 30V