ETMA120250UD-P5P-IC

cui.com
date 07/10/2015
page
1 of 4
SERIES: ETMA 30W DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
MODEL
output
voltage
output
current
output
power
ripple
and noise
1
efciency
level
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
ETMA050400UD 5 4 20 50 V
ETMA090300UD 9 3 27 90 V
ETMA120250UD 12 2.5 30 120 V
ETMA150200UD 15 2 30 150 V
ETMA180165UD 18 1.65 30 180 V
ETMA240125UD 24 1.25 30 240 V
1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
FEATURES
• up to 30 W power
• universal input (90~264 Vac)
• single regulated output from 5~24 Vdc
• over voltage and short circuit protections
• full medical safety approvals
• level V efficiency
• custom designs available
PART NUMBER KEY
ETMA050400UD - XX - IC - CXX
DC Plug Type:
See output plug options
on page 3 for more details
Factory
Designation
Base Number
example of 5 Vdc, 4 A
Reserved for Custom
Congurations
For more information, please visit the product page | For information on related products, view Dc Power Jacks
cui.com
date 07/10/2015 page 2 of 4CUI Inc SERIES: ETMA 30W DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
INPUT
parameter conditions/description min typ max units
voltage 90 264 Vac
frequency 47 63 Hz
input current 0.6 A
inrush current at 240 Vac 50 A
no load power consumption 0.3 W
OUTPUT
parameter conditions/description min typ max units
line regulation
1
±1 %
load regulation
2
5 Vdc output
9 Vdc output
12, 15 Vdc outputs
18, 24 Vdc outputs
±6
±4
±3
±2
%
%
%
%
voltage accuracy ±2 %
hold-up time at 115 Vac 16 ms
switching frequency 70 kHz
temperature coefcient ±0.05 %/°C
Note: 1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
PROTECTIONS
parameter conditions/description
over voltage protection TVS component to clamp
short circuit protection continuous, auto restart
SAFETY & COMPLIANCE
parameter conditions/description min typ max units
isolation voltage input to output 5,656 Vdc
safety approvals IEC 60601-1, EN 60601-1, UL 60601-1
EMI/EMC EN 55011, EN 60601-1-2, EN 61000-3-(2,3)
MTBF as per MIL-HDBK-217F, 115 Vac, 25 °C 300,000 hours
RoHS 2011/65/EU
ENVIRONMENTAL
parameter conditions/description min typ max units
operating temperature 0 40 °C
storage temperature -20 85 °C
humidity non-condensing 93 %
For more information, please visit the product page | For information on related products, view Dc Power Jacks
cui.com
date 07/10/2015 page 3 of 4CUI Inc SERIES: ETMA 30W DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
DC OUTPUT PLUG OPTIONS / DC CORD
MECHANICAL DRAWING
50
0.5
110
0.5
LED
30
0.5
cord length
MODEL NO. CABLE GAUGE CORD LENGTH
ETMA050400UD 18 AWG 720 mm ±50
ETMA090300UD 18 AWG 1,220 mm ±50
ETMA120250UD 18 AWG 1,800 mm ±50
ETMA150200UD 18 AWG 1,800 mm ±50
ETMA180165UD 18 AWG 1,800 mm ±50
ETMA240125UD 18 AWG 1,800 mm ±50
MECHANICAL
parameter conditions/description min typ max units
dimensions 110 x 50 x 30 (4.331 x 1.969 x 1.181 inch) mm
input plug IEC320 / C8
weight 220 g
units: mm
A B C Unit
P5/P5R 5.5 2.1 9.5 mm
P6/P6R 5.5 2.5 9.5 mm
ł
ń
Ń
ł
Ń
ń
Right Angle PXRStandard PX
Stripped and Tinned
*Tip positive
ø3.5 +0/-0.1
14±0.5
ø3.5 +0/-0.1
14±0.5
A
C
B
A
B
C
D
C
A
B
C
C
A
B
A
B
P1RP1
Right Angle PXRStandard PX
Right Angle PXXR
ST
Stripped and Tinned
PXX X X
Standard PXX
N = Center Negative
Plug Polarity:
“Blank” = N/A
P = Center Positive
Plug Angle:
“Blank” = Standard
R = Right Angle
Plug Type
DC PLUG TYPE
EIAJ Plugs
Locking DC Plug
Standard DC Plug
3.5 mm Phono Plug
25±5
2
*Contact CUI for additional output plug options.
For more information, please visit the product page | For information on related products, view Dc Power Jacks

ETMA120250UD-P5P-IC

Mfr. #:
Manufacturer:
Description:
Plug-In AC Adapters Desktop AC Adapters ac-dc, 12 Vdc, 2.5 A, SW, C8 desk-top, P5 center pos, no cord, level V, MED
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union