3EHQ3

37
RFI Power Line Filters
Highest Performance RFI Filters for Medical Equipment
1
HQ Series
HQ Series
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
For email, phone or live chat, please go to
te.com/help
corcom.com
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz: 2 A
@250 VAC 50 Hz: 5 A
Hipot rating (one minute):
Line to Ground: 2250 VDC
Line to Line: 1450 VDC
Rated Voltage (max): 250 VAC
Operating Frequency: 50/60 Hz
Rated Current: 3 & 6A
Operating Ambient Temperature Range
(at rated current I
r
): -10°C to +40°C
In an ambient temperature (T
a
) higher than +40°C
the maximum operating current (I
o
) is calculated as
f
ollows
: I
o
= I
r
(85-Ta)/45
• Designed to provide the highest available
attenuation of RFI noise in the frequency range
from 10kHz to 30MHz for low leakage current
applications
• Size and cost-effective
Ordering Information
6 E HQ 8 M
Suffix
M – Metric inserts (style 8 only)
Omit for standard or none
I
nput / Output Style
1 – .250 [6.3] spade terminals
3 – Wire leads
8 – IEC inlet*
(line side)
Wire leads (load side)
HQ Series
Leakage current designation
E – Low leakage (<0.5 mA)
Current Rating
3 or 6A
Specifications
UL Recognized
CSA Certified
VDE Approved
3EHQ3
Available Part Numbers
6EHQ8
*IEC 60320-1 C14 inlet mates with C13 connector
3EHQ1 6EHQ1
3EHQ3 6EHQ3
3EHQ8 6EHQ8
3EHQ8M 3EHQ8M
3EHQ1
GA400
: NEMA 5-15P to IEC 60320-1 C-13 line cord
Accessories
38
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
For email, phone or live chat, please go to
te.com/help
corcom.com
Highest Performance RFI Filters for Medical Equipment (continued)
HQ Series
Case Styles
HQ1
D
E
A
C
B
C
B
3A Models
6A Models
HQ3
A
D
E
Brown
Blue
Brown
Blue
Grn/Yel
C
B
C
B
Blue
Grn/Yel
Brown
Blue
Grn/Yel
Brown
3A Models
6A Models
3EHQ8 & 3EHQ8M
A
Tapped insert
D
B
E
C
Brown
Grn/Yel
Blue

 
 
 
6EHQ8 & 6EHQ8M
E
A
D
C
B
Grn/Yel
Blue Brown
Tapped inserts
0.63
16.0
Ref

 
 

 
 
 
 

 
 
 
 
Electrical Schematics
LINE
LOAD
3EHQ
LINE
LOAD
6EHQ
39
Highest Performance RFI Filters for Medical Equipment (continued)
RFI Power Line Filters
1
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
For email, phone or live chat, please go to
te.com/help
corcom.com
HQ Series
Recommended Panel Cutout
.235
5.96
1.575
40.01
.157
3.98
.910
23.11
Dia. (2)
1.295
32.89
R

Part No.
ABCDE
  



3EHQ1
    
    
3EHQ3
    
    
3EHQ8,
3EHQ8M
    
    
6EHQ1
    
    
6EHQ3
    
    
6EHQ8,
6EHQ8M
    
    
Case Dimensions

Performance Data
Typical Insertion Loss
Minimum Insertion Loss
Common Mode / Asymmetrical (Line to Ground)
3EHQ
db
100
90
80
70
60
50
40
30
20
10
0
0.01 0.1 1 10 30
Frequency in MHz
6EHQ
100
90
80
70
60
50
40
30
20
10
0
0.01 0.1 1 10 30
Frequency in MHz
Measured in closed 50 Ohm system
Common Mode / Asymmetrical (L-G)
Differential Mode / Symmetrical (L-L)
Measured in closed 50 Ohm system
 
        
            
            
Differential Mode / Symmetrical (Line to Line)
 
        
           
           

3EHQ3

Mfr. #:
Manufacturer:
TE Connectivity
Description:
AC Power Line Filters Power Line Filters 3A WIRE-WIRE LEAD FLANGE MOUNT
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet