GSM220A12-R7B

-A12GSM220
Rated wattage
IEC320-C14 AC inlet
Output voltage
GSM220A is a highly reliable, 220W desktop style single-output green medical adaptor series. This product
is equipped with a 3-pin ( FG) standard IEC320-C14 power plug, adopting the input range from 80VAC to with
264VAC. The entire series supplies different output voltages between 12VDC and 48VDC that can satisfy
the demands for various kinds of medical electrical devices. The circuitry design meets the international
medical standards (2*MOPP), having an ultra low leakage current (<100 A), fitting the medical devices in
direct electrical contact with the patients.
With the efficiency up to 94.5% and the extremely low no-load power consumption below 0.15W,
GSM220A is
under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case. GSM220A is approved with the international medical safety certificates.
R7B
Series name
Description
Model Encoding
Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to save the energy when it is either
compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meet
R7B: Standard model,Power DIN 4Pin with lock type
DC plug type
Other options available by customer requested (see Page 4~5)
{
μ
EN 60601-1
IEC 60601-1
8
+
EN60601-1
+
ANSI/AAMI ES60601-1
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
ApplicationsFeatures
3 pole AC inlet IEC320-C14, Class power unit
Medical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
Extremely low leakage current
No load power consumption<0.15W
Energy efficiency level VI and meet CoC Version 5
-30~+70 wide range working temperature
/ Over temperatureProtections: Short circuit / Overload / Over voltage
LED indicator for power on
Lifetime > 80 K hours
3 years warranty
GSM220A series
220W AC-DC le MedicalAdaptorReliab Green
IEC60601-1
File Name:GSM220A-SPEC 2018-01-12
SPECIFICATION
ORDER NO.
DC VOLTAGE
Note.2
SAFETY MODEL NO.
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
RIPPLE & NOISE (max.)
Note.3
OUTPUT
VOLTAGE RANGE
Note.7
FREQUENCY RANGE
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (max.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 10)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION
OTHERS
NOTE
PACKING
PROTECTION
GSM220A15
GSM220A15-R7B
GSM220A12
GSM220A12-R7B
GSM220A20
GSM220A20-R7B
GSM220A24
GSM220A24-R7B
GSM220A48
GSM220A48-R7B
15V12V 20V 24V 48V
80mVp-p80mVp-p 120mVp-p 120mVp-p 150mVp-p
VOLTAGE TOLERANCE
Note.4
LINE REGULATION
Note.5
LOAD REGULATION
±5.0%±5.0% ±4.0% ±3.0% ±2.0%
±1.0%±1.0% ±1.0% ±1.0% ±1.0%
±5.0%±5.0% ±4.0% ±3.0% ±2.0%
80 ~ 264VAC
47 ~ 63Hz
PF>0.91 / 230VAC PF>0.98 / 115VAC at full load
90%90% 92% 93.5%
94.5%
4A / 115VAC 2A / 230VAC
Cold start 55 A / 115AVC 110A / 230VAC
WITHSTAND VOLTAGE
Note.9
ISOLATION RESISTANCE
210.79K hrs min. MIL-HDBK-217F(25 )
-30 ~ +70 (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH non-condensing
±℃0.03% / (0~40 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
210*85*46mm (L*W*H)
1.1Kg; 12pcs/14.2Kg/0.91CUFT
CONNECTOR
PLUG
CABLE
See page 4~5 ; Other type available by customer requested
See page 4~5 ; Other type available by customer requested
OVERLOAD
OVER VOLTAGE
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
13.4A15A 11A 9.2A 4.6A
0 ~ 13.4A0 ~ 15A 0 ~ 11A 0 ~ 9.2A
0 ~ 4.6A
201W180W 220W 221W 221W
Protection type : Shut down o/p voltage, re-power on to recover
SETUP, RISE TIME
Note.6
HOLD UP TIME (Typ.)
2000ms, 50ms / 230VAC 2000ms, 50ms / 115VAC at full load
20ms / 230VAC 20ms / 115VAC at full load
OVER TEMPERATURE
Shut down o/p voltage, recovers automatically after temperature goes down
Earth leakage current < 115 A/264VAC , Touch current <100 A/264VAC
1. All parameters are specified at 230VAC input, rated load, 25 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1 f & 47 f capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltages. Pleas check the derating curve for more details.
8. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
9. Optional for 1.5KVAC with BF rated.
10 .The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC
directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
ISOLATION LEVEL
Primary-Secondary: 2xMOPP
, Primary-Earth: 1xMOPP
OPERATING ALTITUDE Note.8
3000 meters
Parameter
Parameter
ESD
Standard
Standard
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
EN55024 , EN60601-1-2, EN61204-3
Level 4, 15KV air ; Level 4, 8KV contact
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
Level 3, 1KV/Line-Line , 2KV/Line-FG
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
IEC60601-1, TUV EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved
,
EAC TP TC 004
μ
μ
μ
μ
GSM220A series
220W AC-DC le MedicalAdaptorReliab Green
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
File Name:GSM220A-SPEC 2018-01-12
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
Derating Curve Static Characteristics
Mechanical Specification
INPUT VOLTAGE (VAC) 60Hz
80 10090 120110 160140 200180 240220 264
LOAD (%)
90
100
80
70
60
50
40
For 12~15V
For others
Power LED
46
85
210
30 3mm±
UL2464 16AWGx4C 1000 50mm±
Case No. 961A Unit:mm
AMBIENT TEMPERATURE ( )
LOAD (%)
0-30 2010 30 40
20
50
80
100
6050 70
(HORIZONTAL)
40
60
PIN NO.
1
2
3
4
OUTPUT
+
Vo
-Vo
-Vo
+Vo
2
3
1
4
AC FG
Outer shell connected to AC FG
-V not connected to AC FG
2
14
3
KYCON KPPX 4P equivalent-
DC output plug
Standard plug: R7B
R7B
Pin Assignment
AC Inlet
IEC320-C14
95
GSM220A series
220W AC-DC le MedicalAdaptorReliab Green
File Name:GSM220A-SPEC 2018-01-12

GSM220A12-R7B

Mfr. #:
Manufacturer:
Description:
SPS Desktop 180W 12V/15A Medical
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet