PK2-5MM-101

PP010-OM-E Rev A 7
Typical Voltage Derating Curve
10
100
1000
0.1 1 10 100 1000
Frequency [MHz]
Voltage [V] DC+AC peak
Typical Input Impedance
10
100
1000
10000
100000
1000000
10000000
0.01 0.1 1 10 100 1000
Frequency [MHz]
Zin [ohms]
8 PP010-OM-E Rev A
Probe Compensation
Proper compensation of the probe is required to assure good amplitude
accuracy in the dynamic portions of the waveform being measured. LF
compensation matches the probe to differences in oscilloscope input
capacitance. The LF compensation should always be checked and adjusted as
needed when first connecting a passive probe to the oscilloscope input.
LF compensation is performed by connecting the input of the probe to a low
frequency square wave, such as the oscilloscope calibrator signal set to 1 kHz.
The compensation is adjusted by rotating the trimmer pot accessible through
the small hole in the center of the housing near the BNC connector. Use the
tool supplied with the probe for this adjustment.
Adjusting Probe Compensation
§ § §

PK2-5MM-101

Mfr. #:
Manufacturer:
Description:
LeCroy, Probe Accessory, Sprung Hook for PP010 Probe