GSM120B12-R7B

-B12GSM120
Rated wattage
IEC320-C8 AC inlet
Output voltage
R7B
Series name
Description
Model Encoding
R7B: Standard model,Power DIN 4Pin with lock type
DC plug type
{
GSM120B is a highly reliable, 120W desktop style single-output green medical adaptor series. This product
is equipped with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80VAC to
264VAC. The entire series supplies different output voltages between 12VDC and 48VDC that can satisfy
the demands for various kinds of medical electrical devices. The circuitry design meets the international medical
standards (2*MOPP), having an ultra low leakage current (<100 A), fitting the medical devices in direct electrical
contact with the patients.
With the efficiency up to 91.5% and the extremely low no-load power consumption below 0.15W, GSM120B is
under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case, providing the double insulation that effectively prevents electrical shock. GSM120B is approved with
the international medical safety certificates.
Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to
save the energy when it is either
compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meet
GSM120B series
120W AC-DC le MedicalAdaptorReliab Green
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
ApplicationsFeatures
2 pole AC inlet IEC320-C8 , Class power unit
Medical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
Extremely low leakage current
No load power consumption<0.15W
Energy efficiency level VI and meet CoC Version 5
-30~+70 wide range working temperature
Protections: Short circuit / Overload / Over voltage / Over temperature
LED indicator for power on
Lifetime > 105 K hours
3 years warranty
+
+
μ
Other options available by customer requested (see Page 4~5)
EN 60601-1
IEC 60601-1
8
EN60601-1
ANSI/AAMI ES60601-1
IEC60601-1
File Name:GSM120B-SPEC 2018-01-12
SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
Note.2
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
RIPPLE & NOISE (max.)
Note.3
OUTPUT
VOLTAGE RANGE 7
Note.
FREQUENCY RANGE
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 9)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION
OTHERS
NOTE
PACKING
PROTECTION
GSM120B15-R7BGSM120B12-R7B GSM120B20-R7B GSM120B24-R7B GSM120B48-R7B
15V12V
GSM120B15GSM120B12
20V
GSM120B20
24V
GSM120B24
48V
GSM120B48
120mVp-p100mVp-p 150mVp-p 180mVp-p 200mVp-p
VOLTAGE TOLERANCE
Note.4
LINE REGULATION Note.5
LOAD REGULATION
±5.0% ±5.0% ±3.0% ±2.5%
±1.0%±1.0% ±1.0% ±1.0% ±1.0%
±5.0%±5.0% ±4.0% ±3.0% ±2.5%
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
PF>0.93 / 230VAC PF>0.97 / 115VAC at full load
89%88% 89.5% 90% 91.5%
1.4A / 115VAC 0.7A / 230VAC
35A / 115VAC 70A / 230VACCold start
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P: 4KVAC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
372K hrs min. MIL-HDBK-217F(25 )
-30 ~ +70 (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH non-condensing
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
167*67*35mm (L*W*H)
0.6Kg; 20pcs/13,0Kg/0.9CUFT
CONNECTOR
PLUG
CABLE
See page ; Other type available by customer requested4~5
See page ; Other type available by customer requested4~5
OVERLOAD
OVER VOLTAGE
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
7A8.5A 6A 5A 2.5A
0 ~ 7A0 ~ 8.5A 0 ~ 6A 0 ~ 5A
0 ~ 2.5A
105W102W 120W 120W 120W
Protection type : Shut down o/p voltage, re-power on to recover
SETUP, RISE TIME
Note.6
HOLD UP TIME (Typ.)
1500ms, 30ms / 230VAC 2000ms, 30ms / 115VAC at full load
40ms / 230VAC 24ms / 115VAC at full load
OVER TEMPERATURE
Shut down o/p voltage, re-power on to recover
Touch current < 100 A/264VAC
±5.0%
±0.03% / (0 ~ 40 )℃℃
ISOLATION LEVEL
Primary-Secondary: 2xMOPP
1. All parameters are specified at 230VAC input, rated load, 25 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1 f & 47 f capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltage. Please check the derating curve for more details.
8. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
GSM120B series
120W AC-DC le MedicalAdaptorReliab Green
OPERATING ALTITUDE Note.8
Parameter
Parameter
ESD
Standard
Standard
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
EN55024 , EN60601-1-2, EN61204-3
Level 4, 15KV air ; Level 4, 8KV contact
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
Level 3, 1KV/Line-Line
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
IEC60601-1, EN60601-1 EN , ANSI/AAMI ES60601-1 (3.1 version),/ 60601-1-11 / ES60601-1-11
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved
, EAC TP TC 004
μ
3000 meters
μ
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
File Name:GSM120B-SPEC 2018-01-12
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
Derating Curve Static Characteristics
Mechanical Specification
Case No. Unit:mmGS120A
INPUT VOLTAGE (VAC) 60Hz
80 10090 120110 160140 200180 240220 264
LOAD (%)
90
100
80
70
60
50
40
35
167
67
PLEDower
30 3mm±
UL2464 16AWGx4C 1000 50mm for 12V±
UL2464 18AWGx4C 1200 50mm for 15 ~ 48V±
GSM120B series
120W AC-DC le MedicalAdaptorReliab Green
AC Inlet
IEC320-C8
PIN NO.
1
2
3
4
OUTPUT
+
Vo
-Vo
-Vo
+Vo
2
3
1
4
2
14
3
KYCON KPPX 4P equivalent-
Standard plug: R7B
R7B
Pin Assignment
DC output plug
AMBIENT TEMPERATURE ( )
LOAD (%)
0-30 2010 30 40
20
50
80
100
6050 70
(HORIZONTAL)
40
60
File Name:GSM120B-SPEC 2018-01-12

GSM120B12-R7B

Mfr. #:
Manufacturer:
MEAN WELL
Description:
Desktop AC Adapters 102W 12V 8.5A 2pole Medical Adaptr
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union