GSM7-24

14
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
GSM7 Medical
7 Watt Single Output Global Performance Switchers
FEATURES:
Worlds smallest medically approved 7 W supply
2.00" x 2.28" x 0.56"
Wide-range ac input (85-264 Vac)
Medical Approval to UL2601-1 and EN60601-1
EMI conducted to CISPR11 Class B
Ideal for ultra-compact applications
marked to LVD
SPECIFICATIONS:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Holdup Time
16 ms minimum from loss of ac.
Output Power
Normal continuous output power is 7 W, 9 W peak for 60 s
maximum duration, 10% duty cycle. Factory set to begin
power limiting at approximately 9 W.
Output Regulation
Regulation from initial setpoint measured by changing
load from no-load to 50% load or 50% load to full load in
either direction. Initial sepoint tolerance is measured at
50% load. No minimum load is required to maintain
proper regulation.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit.
No Load Turn-on/Standby
Full regulation at no load.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential
mode. Measured with scope probe directly across output
terminals of the power supply with load terminated with
0.1 uF capacitor.
Voltage Adjustment
Factory set with fixed resistors to maximize reliability.
Efficiency
72% minimum for the 5.0 V model at full rated load,
nominal input voltage. Efficiency increases as output
voltage increases.
Input Protection
Internal ac fuse provided on all units. Designed to open
only if catastrophic failure occurs in the unit; fuse does not
blow on overload or short circuit. Internal 320 Vac MOV
included. Units will not be damaged when tested for input
transients in accordance with the following standards:
ESDIEC1000-4-2 Level III
Fast TransientIEC1000-4-4 Level III
Environmental
Designed for 0 to 50°C operation at full rated output
power; derate output current and total output power by
2.5% per °C above 50°C. See Environmental specifica-
tion on next page.
Medical EMI/EMC Compliance
All models include built-in EMI filtering to meet the
following emissions requirements:
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions EN55011 Class B; FCC Class B
Static Discharge EN61000-4-2, 6 kV contact, 8 kV air
RF Field Susceptibility EN61000-4-3, 3 V/meter
Fast Transients/Bursts EN61000-4-4, 2 kV, 5 kHz
Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com.
Medical Safety
Condor D.C. Power Supplies, Inc. declares under our sole
responsibility that all GSM7 models are in conformity with
the applicable requirements of EN60601-1 following the
provisions of the Low Voltage Directive 73/23/EEC.
All GSM7 models are certified to be in compliance with
the applicable requirements of UL2601-1 (Patient Care
Equipment), CSA 22.2 No. 234 (with additional tests to
C22.2 No 601.1 per T.I.L. CA-08), EN 60601-1. Classification
(in accordance with sub-clause 5 of EN60601-1):
(5.1) Protection against electric shock = class II
(5.2) Degree of protection against electric shock = signal
output or intermediate
(5.3) Protection against harmful ingress of water =
ordinary (no protection)
(5.5) Not suitable for use in the presence of flammable
anesthetic mixture
(5.6) Mode of operation = continuous
Earth Leakage Current
Leakage current measured in ground wire connection to
output common.
ledoM
*tnerruCegakaeL **tnerruCegakaeL
egatloVtseT
dohteMtseT
7MSG Aµ23Aµ05zH06/caV2311-1062LU
7MSG Aµ55Aµ58zH05/caV4621-106CEI
* Normal condition **Single fault condition
15
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 805-486-4565 FAX 805-487-8911 www.condorpower.com
GSM7 MECHANICAL SPECIFICATIONS
GSM7 Medical 7 Watt Single Output
A. Units should be allowed to warm up/operate under non-condensing conditions
before application of power.
B. Random vibration10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
C. Shock testinghalf-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal
axes, total 6 shocks.
AC NEUTRAL
latnemnorivnE
noitacificepS
gnitarepO gnitarepo-noN
)A(erutarepmeTCº05ot0Cº58+ot04-
)A(ytidimuHHR%59ot0HR%59ot0
)B(kcohSg02
kp
g04
kp
edutitlAtf000,01ot005-tf000,04ot005-
)C(noitarbiVg5.1
smr
g300.0,
2
zH/g5
smr
g620.0,
2
zH/
ledoM tuptuO tnerruC noitalugeRdaoL
tniopteSlaitinI
ecnareloT
tniopteSPVO esioNdnaelppiR
5-7MSG V1.5A4.1%57.0%5.2V6.0±2.6%1
21-7MSG V21A6.0%57.0%5.2V1.1±41%1
51-7MSG V51A5.0%57.0%5.2V5.1±5.81%1
42-7MSG V42A3.0%57.0%5.2V5.2±82%1
82-7MSG V82A3.0%57.0%5.2V8.2±43%1