THM 6-2413WI

www.tracopower.com Page 1 of 4
DC/DC Converter THM 6WI Series, 6 Watt
The THM-6WI series is a range of medical 6 Watt DC/DC converters in DIP-24
plastic package and with ultra-wide 4:1 input voltage range. They provide a rein-
forced isolation system for 5000 VACrms isolation and a very low leakage current
of less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition for
2  MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk
management file. Design and production conform to the quality management
system ISO 13485. With a high efficien¬cy of up to 87% and highest grade com-
ponents the converters can reliably operate in an ambient temperature range
of –4C up to +90°C. They constitute a reliable solution not only for medical
equipment but also for demanding ranges of application such as transportation,
control & measurement or IGBT drivers.
Ultra wide 4:1 input voltage 6 W DC/DC
converter in a compact DIP-24 plastic case
I/O isolation 5000 VACrms rated for
250 VACrms working voltage
Certification according to IEC/EN/ES
60601-1 3rd edition for 2MOPP
Risk management process according to
ISO 14971 including risk management file
Acceptance criteria for electronic
assemblies according to IPC-A-610 Level 3
Low leakage current < 2A
Extended operating temperature range
–40°C to 90°C.
EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
Operating up to 5000m altitude
5 year product warranty
Order code Input voltage range Output voltage Output current max. Efficiency typ.
THM 6-0510WI
3.3 VDC 1800 mA 82.5 %
THM 6-0511WI
5.0 VDC 1200 mA 86.0 %
THM 6-0512WI
12 VDC 500 mA 86.0 %
THM 6-0513WI
4.5 – 9 VDC
15 VDC 400 mA 86.0 %
THM 6-0515WI (5 VDC nominal)
24 VDC 250 mA 86.5 %
THM 6-0521WI
±5.0 VDC ±600 mA 84.0 %
THM 6-0522WI
±12 VDC ±250 mA 86.5 %
THM 6-0523WI
±15 VDC ±200 mA 87.5 %
THM 6-2410WI
3.3 VDC 1800 mA 83.0 %
THM 6-2411WI
5.0 VDC 1200 mA 86.5 %
THM 6-2412WI
12 VDC 500 mA 89.0 %
THM 6-2413WI
9. 0 – 36 VDC
15 VDC 400 mA 89.0 %
THM 6-2415WI (24 VDC nominal)
24 VDC 250 mA 88.5 %
THM 6-2421WI
±5.0 VDC ±600 mA 85.0 %
THM 6-2422WI
±12 VDC
±250 mA
88.5 %
THM 6-2423WI
±15 VDC
±200 mA
88.0 %
THM 6-4810WI
3.3 VDC
1800 mA
82.5 %
THM 6-4811WI
5.0 VDC 1200 mA 86.5 %
THM 6-4812WI
12 VDC 500 mA 88.0 %
THM 6-4813WI
18 – 75 VDC
15 VDC 400 mA 89.0 %
THM 6-4815WI (48 VDC nominal)
24 VDC 250 mA 88.0 %
THM 6-4821WI
±5.0 VDC ±600 mA 85.0 %
THM 6-4822WI
±12 VDC ±250 mA 88.0 %
THM 6-4823WI
±15 VDC ±200 mA 88.0 %
Models
ES 60601-1
C B
Scheme
IEC 606010-1
THM 6WI Series, 6 Watt
www.tracopower.com Page 2 of 4
Input Specifications
Input current no load 5 Vin models: 20 mA typ.
24 Vin models: 6 mA typ.
48 Vin models: 4 mA typ.
Surge voltage (3 s max.) 5 Vin models: 16 V max.
24 Vin models: 50 V max.
48 Vin models: 100 V max.
Start-up voltage 5 Vin models: 4. 5 VDC (or lower)
24 Vin models: 9 VDC (or lower)
48 Vin models: 18 VDC (or lower)
Startup time 30 ms
Under voltage shut down 5 Vin models: 4 VDC typ.
24 Vin models: 8 VDC typ.
48 Vin models: 16 VDC typ.
EMC emissions – Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
EN 55032 class A (internal filter)
– Application note for filter class B proposal www.tracopower.com/overview/thm6wi
EMC immunity – Generic for Medical equipment IEC/EN 60601-1-2 4th edition
– ESD (electrostatic discharge) EN 61000-4-2, air ±8 kV, contact ±6 kV,
perf. criteria A
– Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A
– Fast transient / surge EN 61000-4-4, ±2 kV, perf. criteria A
(with external input capacitor / diode) EN 61000-4-5, ±2 kV perf. criteria A
5 Vin models: Nippon chemi-con KY 1000 µF/ 25 V and
reverse diode (Vishay V10P45) in parallel
24 Vin models: Nippon chemi-con KY 470 µF/ 50 V
48 Vin models: Nippon chemi-con KY 330 µF/ 100 V
Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A
– Magnetic field immunity EN 61000-4-8
100 A/m, continuous, perf. criteria A
1000 A/m, 1 sec., perf. criteria A
External input fuse required 5 Vin models: 5 A (slow blow type)
24 Vin models: 2.5 A (slow blow type)
48 Vin models: 1.5 A (slow blow type)
Output Specifications
Voltage set accuracy ±1 % max.
Regulation Input variation single output: 0.2 % max.
dual output: 0.5 % max.
– Load variation (0 – 100 %) single output: 0.2 % max.
dual output: 1.0 % max.
– Cross regulation dual output: 5.0 % max. (asymmetrical load 25/100 %)
Minimum load not required
Ripple and noise (20 MHz Bandwidth) 3.3 & 5.0 Vout models: 30 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
12 & 15 Vout models: 40 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
24 Vout models: 50 mVp-p typ. with cap. 4.7 µF/50 V X7R MLCC
Transient response – Recovery time (25% load step change) 250 s typ.
Over load protection at 150 % typ. of Iout rated (hiccup mode)
Short circuit protection Continuous, automatic recovery
Over voltage protection – Single output 3.3 Vout models: 3.7 – 5.4 VDC
5.0 Vout models: 5.6 – 7.0 VDC
12 Vout models: 13.5 – 19.6 VDC
15 Vout models: 18.3 – 22.0 VDC
24 Vout models: 29.1 – 32.5 VDC
– Dual output ±5 Vout models: 5.6 – 7.0 VDC
±12 Vout models: 13.5 – 18.2 VDC
±15 Vout models: 17.0 – 22.0 VDC
THM 6WI Series, 6 Watt
www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
General Specifications
Capacitive load – Single output 3.3 Vout models: 2‘100 F max.
5.0 Vout models: 1‘500 F max.
12 Vout models: 260 F max.
15 Vout models: 210 F max.
24 Vout models: 75 F max.
– Dual output ±5 Vout models: 860 F max. (each output)
±12 Vout models: 150 F max. (each output)
±15 Vout models: 110 F max. (each output)
Temperature ranges – Operating (designed for) –40°C to +90°C (without derating)
– Rated according to IEC/EN 60601-1 –40°C to +70°C (without derating)
– Case temperature +105°C max.
– Storage temperature –55°C to +125°C
Thermal impedance 18 °C/W
Humidity (non condensing) 5 % to 95 % rel H max.
Isolation voltage – to meet ES/IEC/EN 60601-1 5000 VACrms, rated for 250 VACrms working
(50 Hz, 60 s) voltage, 2  MOPP
Clearance/creepage 8 mm min.
Leakagecurrent ( at 240 VAC, 60 Hz) 2 A max.
Isolation capacitance (input/output) 17 pF max.
Altitude during operation 5000 m
Temperature coefficient ±0.02 %/K typ.
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign) 4700‘000 h
Switching frequency 250 kHz ±25 kHz (pulse width modulation)
Vibration and thermal shock resistance according to MIL-STD-810F
Safety standards/approvals – Medical equipment ANSI/AAMI ES60601-1:2005/(R)2012,
IEC/EN60601-1 3rd edition
– Certification documents www.tracopower.com/overview/thm6wi
Environmental compliance Reach www.tracopower.com/info/reach-declaration.pdf
RoHS RoHS directive 2011/65/EU
Physical Specifications
Casing material non-conductive black plastic
Base material non-conductive black plastic
Potting material silicone (UL94 V-0 rated)
Package weight 14 g (0.48 oz)
Soldering temperature 265°C / 10 s max.
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.

THM 6-2413WI

Mfr. #:
Manufacturer:
TRACO Power
Description:
Isolated DC/DC Converters Product Type: DC/DC; Package Style: DIP-24; Output Power (W): 6; Input Voltage: 9-36 VDC; Output 1 (Vdc): 15; Output 2 (Vdc): N/A; Output 3 (Vdc): N/A
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union