968-038

DGS-SO2 968-038
March 2017
PRODUCTION DATA information is current as of publication date.
Products conform to specifications per the terms of the SPEC Copyright © 2011-2017, SPEC Sensors LLC
Sensors standard warranty. Production processing does not
necessarily include testing of all parameters.
EEPROM EXAMPLE OUTPUT
EEPROM Values
nA_per_PPM_x100 = 3947
ADC_OC = 32712
ADC_Zero = 32712
ADC_Span = 65535
PPM_Span_x100 = 65535
Temperature_Offset_x1000 = -1
T_Zero = 25360
RH_Zero = 31218
T_Span = 65535
RH_Span = 65535
LMP91000 Register 0x10 = 3
LMP91000 Register 0x11 = 182
LMP91000 Register 0x12 = 3
Barcode = 090115011033 110601 SO2 1509 39.47
Serial_Number_ = 090115011033
Part_Number = 110601
Gas = SO2
Date_Code = 1509
Sensitivity_Code = 39.47
IMPORTANT PRECAUTIONS
All sensor designs are made for air monitoring @ 1 atm +/- 0.2 atm. Because applications of use and
device implementation are outside our control, SPEC Sensors cannot guarantee performance in a given
device or application, and disclaims any and all liability therefore. Customers should test under their
own conditions to ensure the sensors are suitable for their requirements.
Contact the factory to discuss specific concerns that might damage the sensor performance or life.
Condensation and Water (1)
High Temperature Operation (> 40C) for more than 1 month
Low Humidity Operation (< 15% RH) for more than 3 months
Highly contaminated air over a prolonged period
High levels of particles or soot (unless proper filtering is provided)[2]
(1) Use of porous PTFE membrane or filter cap may address this concern)
(2) Use of replaceable filter recommended where dust and particulate is expected.
DGS-SO2 968-038
March 2017
PRODUCTION DATA information is current as of publication date.
Products conform to specifications per the terms of the SPEC Copyright © 2011-2017, SPEC Sensors LLC
Sensors standard warranty. Production processing does not
necessarily include testing of all parameters.
SENSOR STORAGE, HANDLING AND SOLDERING
This information embodies various general recommendations concerning the storage, handling, and manual soldering
conditions for SPEC SENSORS CSPEC Modules. It is only applicable for modules guaranteed by SPEC SENSORS
stated in SPEC SENSORS Sensor Specification Sheet. Moreover, SPEC SENSORS' modules are NOT warranted
and should NOT be used in high temperature soldering (reflow) or pre-tinning baths.
Sensor & Module Handling
Handle sensors with care. Take precautions, including but not limited to the following:
A. DO NOT apply excessive pressure to the top or bottom of the sensor module.
B. Whenever possible, handle or make contact with the sensor module from the sides of the PCB or substrate.
C. Light vacuum pressure is possible during handling, DO NOT apply vacuum over gas sensor port.
D. If the sealed sensor package is opened, DO NOT re-seal using vacuum or nitrogen gas. DO NOT reseal with
desiccant.
E. DO NOT obstruct the gas sensor port by making direct contact with any tape, apparatus, weights, etc.
F. DO NOT use silicone or other conformal coatings around the sensor or gas port-holes.
G. Operators are requested to wear powder free antistatic gloves.
Manufacturing Assembly Floor Environment
SPEC SENSORS recommends that the manufacturing assembly floor environment be maintained at controlled
conditions:
A. Temperature: 18 - 26°C
B. Relative Humidity: 40 to 60%
C. Pressure: 1.0 ± 0.2 atm
Sensor & Module Storage Conditions
The shelf life for sealed, packaged components is 12 months from the pack seal date, when stored in the factory-
sealed bag under the following conditions:
A. Temperature: 5 to 25 °C
B. Relative Humidity: 20 to 80%
C. Pressure: 1.0 ± 0.2 atm
D. Storage Time: 12 months
When moving from Storage Conditions to the Manufacturing Assembly Floor Environment, the sensors should be
allowed to equilibrate at the new conditions for at least 24 hours prior to manufacturing.
Module Attach Soldering Process
Hand solder only. Keep the soldering iron or solder process tool away from the sensor. The sensor should not see
pre-heat temperatures above 70 °C. There have been suggested cases where a heat sink cover over the sensor may
be applicable to protect the sensor during processing. No Application notes to this approach avai lable. Only to be
used as reference only.
A. DO NOT heat sensor above 70 °C
B. Hand or peripheral process type approach
C. Use solder wire alloy with the lowest possible eutectic temperature
D. Use lowest possible soldering iron temperature
E. Contact the host board with the soldering iron at a 45° angle on the solder pad
F. Keep the soldering iron away from the top and bottom of the sensor module
G. DO NOT place in reflow, wave or IR reflow type processes
H. DO NOT place mounted board In a wash

968-038

Mfr. #:
Manufacturer:
Description:
DGS-SO2 DIGITAL GAS SENSOR MODUL
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union