TPP 150-112A-J

www.tracopower.com Page 1 of 4
AC/DC Medical Power Supply TPP 150A-J Series, 150 Watt
Models
Open frame 150 W power supply with
JST connection in 2.0 x 4.0 package
Certification according to IEC/EN/ES
60601-1 3rd edition for 2MOPP
Low leakage current <75 A rated for
BF applications
Risk management process according to
ISO 14971 including risk management file
Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3
Active power factor correction >0.95
Protection class l and ll prepared
Operating up to 5000 m altitude
Ready to meet ErP directive,
<0.3 W no load power consumption
5 year product warranty
The TPP 150A-J series of 150 Watt AC/DC open frame power supplies feature a
reinforced double I/O isolation system according to latest medical safety stan-
dards (60601-1 3rd edition, 2  MOPP). The earth leakage current is below 75
µA which makes the units suitable for BF (body floating) applications. The excel-
lent efficiency of up to 92% allows a high power density for the standard 2.0”
x 4.0” packaging format. The full load operating temperature range is –25°C to
+55°C while it goes up to 8C with 50% load derating. The EMC characteristic
is dedicated for applications in industrial and domestic fields. High reliability is
provided by the use of industrial quality grade components and an excellent ther-
mal management. It makes the products an ideal solution for medical devices
and for demanding safety and space critical applications.
prepared
C B
Scheme
ES 60601-1
U L 6 0 950 -1
Order code Output voltage
Output current max.
natural convection
Output current max.
forced air cooling
Efficiency typ.
TPP 150-112A-J
12 VDC
(10.8 - 13.2 VDC)
8.34 A 12.5 A 91 %
TPP 150-115A-J
15 VDC
(13.5 - 16.5 VDC)
7.34 A 10 A 92 %
TPP 150-124A-J
24 VDC
(21.6 - 26.4 VDC)
4.59 A 6.25 A 92 %
TPP 150-128A-J
28 VDC
(25.2 - 30.8 VDC)
3.93 A 5.36 A 92 %
TPP 150-136A-J
36 VDC
(32.4 - 39.6 VDC)
3.06 A 4.17 A 92 %
TPP 150-148A-J
48 VDC
(43.2 - 52.8 VDC)
2.09 A 3.13 A 92 %
Models
Encased version with screw terminal
connection see TPP 150 Series
www.tracopower.com/overview/tpp150
TPP 150A-J Series, 150 Watt
www.tracopower.com Page 2 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Input Specifications
Input voltage range – AC range (universal input) 85 – 264 VAC
– DC range 120 – 370 VDC
Input frequency 47 – 63 Hz
Input current at full load – at 115 VAC / 230 VAC 1.7 A max. / 0.8 A max.
Input protection T 3.15 A / 250 VAC (internal fuse)
Input inrush current – at 230 VAC 60 A max.
Zero load power consumption 0.3 W max. (acc. ErP directive)
Power factor 0.95 min.
Output Specifications
Voltage set accuracy ±1%
Output voltage adjustment ±10% (by trim potentiometer)
Regulation – Input variation (Vin min. to Vin max.) 0.2% max.
– Load variation (0 to 100%) 0.5% max.
Minimum load not required
Temperature coefficient 0.02 %/K max.
Hold-up time – Vin = 115 VAC 16 ms min.
Start-up time 1 s max.
Rise time 20 ms typ.
Ripple and noise 12 Vout models: 120 mVp-p typ.
w. cap. 1F/25V 1206 X7R MLCC
(20 MHz bandwidth) 15 Vout models: 150 mVp-p typ.
w. cap. 1F/25V 1206 X7R MLCC
24 Vout models: 220 mVp-p typ.
w. cap. 1F/50V 1206 X7R MLCC
28 Vout models: 220 mVp-p typ.
w. cap. 1F/50V 1206 X7R MLCC
36 Vout models: 250 mVp-p typ.
w. cap. 1F/50V 1206 X7R MLCC
48 Vout models: 250 mVp-p typ.
w. cap. 0.1F/100V 1206 X7R MLCC
Overvoltage protection 115 – 135% of nominal Vout
Overload protection 115 – 150% Iout typ.
Short circuit protection Hiccup mode, continuous (automatic recovery)
Transiente response – Peak deviation (25% load step change) 3% max.
– Recovery time 500 s typ.
Fan power supply – Temperature-sensitive speed control 12 VDC / 500 mA max.
Capacitive load 12 Vout models: 10‘400 F max.
15 Vout models: 6‘600 F max.
24 Vout models: 2‘600 F max.
28 Vout models: 1‘900 F max.
36 Vout models: 1‘150 F max.
48 Vout models: 650 F max.
General Specifications
Temperature ranges – Operating temperature –25°C to +85°C (with derating, see page 3)
– Storage temperature –40°C to +85°C
Humidity (non condensing) 5 – 95 % rel. H.
Altitude during operation 5000 m max.
Switching frequency (at 230 VAC) 60 kHz typ. (pulse frequency modulation)
Isolation voltage (60 s) – Input to Output 4000 VAC
(2  MOPP insulation) – Input/Output to Case 2000 VAC
Isolation resistance (at 500 VDC) 100 MOhm min.
TPP 150A-J Series, 150 Watt
www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Derating
General Specifications (continued)
Leakage current (at 264 VAC / 60 Hz) 100 A max.
Reliability – calculated MTBF at +25°C acc. MIL-HDBK-217F 786100 h
Protection class class I and II prepared
EMC emissions – Conducted / radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
Conducted: EN 55032 class B (internal filter)
Radiated: EN 55032 class A (internal filter)
– Harmonic current emissions IEC/EN 61000-3-2, class A & D
– Voltage flicker IEC/EN 61000-3-3
EMC immunity IEC/EN 60601-1-2, EN 55024
– Electrostatic discharge ESD IEC/EN 61000-4-2, ±15kV/8kV perf. criteria A
– RF field immunity IEC/EN 61000-4-3, 20V/m perf. criteria A
– Electrical fast transients/burst immunity IEC/EN 61000-4-4, ±2kV perf. criteria A
– Surge IEC/EN 61000-4-5, ±1kV/2kV perf. criteria A
– Conducted RF IEC/EN 61000-4-6, 20 Vrms perf. criteria A
– Magnetic field IEC/EN 61000-4-8, 10A/m perf. criteria A
– Voltage dip and interruptions IEC/EN 61000-4-11, see below
Voltage dip and interruptions according to EN 60601-1-2 30%, 500ms perf. criteria A
Reference: 230 VAC / 50Hz 60%, 100ms perf. criteria A
> 95%, 10ms perf. criteria A
> 95%, 5000ms perf. criteria B
Safety standards and certification IEC/EN/UL 60950-1,
IEC/EN 60601-1 3rd edition,
ANSI/AAMI ES60601-1:2005(R)2012
– Certification documents www.tracopower.com/overview/tpp150a
Environment – Vibration acc. IEC 60068-2-6
– Shock acc. IEC 60068-2-27
– Thermal shock acc. MIL-STD-810F
Environmental compliance – Reach
www.tracopower.com/info/reach-declaration.pdf
– RoHS RoHS directive 2011/65/EU
Weight 187 g (6.60 oz)
40 50 80
150
0
Output Power [W]
Ambient Temperature [°C]
-25
12 & 48 Vout models at input voltage 230 VAC
other output models at input voltage 230 VAC
all models at input voltage 115 or 230 VAC
90
60 70
5040 60 80
Output Power [W]
Ambient Temperature [°C]
-25
40
140
70
90
12 & 48 Vout models at input voltage 115 VAC
other output models at input voltage 115 VAC
Forced air cooling of 10 CFM (external fan)
Natural convection
60
80
100
120
20
160

TPP 150-112A-J

Mfr. #:
Manufacturer:
TRACO Power
Description:
AC/DC Power Modules 150W 12V 12.5A 2X4 Med. Open
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union