AWM720P1

Mass Airflow Sensors
AWM720P1 Airflow
AWM700 Series
Sensing and Control
FEATURES
Flow tubes for ranges up to
200 SLPM
Highly stable null and full-
scale
Compact package design
Extremely low hysteresis
and repeatability errors,
less than 0.35% of reading
Fast response time, 6 ms
typical
Low power consumption,
less than 60 mW
TYPICAL APPLICATIONS
Oxygen concentrators
Oxygen conservers
Respirators and ventilators
Nebulizers
Continuous positive airway
pressure (CPAP) equipment
Anesthesia delivery
Leak detection
Spectroscopy
Mass flow controllers
Telecommunication
systems
Environmental climate
controls
Fuel cell controls
WARNING
PERSONAL INJURY
DO NOT USE these
products as safety or
emergency stop devices or
in any other application
where failure of the product
could result in personal
injury.
Failure to comply with these
instructions could result in
death or serious injury.
AWM700 Series microbridge mass airflow sensors provide in-line flow
measurement with a specially designed bypass flow housing. The sensors
measure flow as high as 200 standard liters per minute (SLPM) while
inducing a pressure drop of 1 inch H
2
O, typically. The AWM700 has a high
flow range capability in a small package.
The AWM700 has a 6 millisecond response time, requires a 10 Vdc
supply, but consumes only 60 mW of power. The compact plastic package
withstands overpressures of 25 psi without compromising performance.
The snap-in AMP compatible connector provides reliable connection. The
sensor is also well suited for use in portable devices and battery-powered
applications.
The AWM700 Series provides a combination of time proven reliability,
high accuracy, and precision operating characteristics. This inherent
accuracy over life reduces need for recalibration. AWM700 sensor circuitry
performs amplification and temperature compensation.
The AWM720P1 200 LPM Mass Airflow Sensor, developed primarily for
the medical ventilation market, meets the high performance requirements
of many medical and analytical instrumentation applications.
WARNING
MISUSE OF DOCUMENTATION
The information presented in this product sheet is for reference only. Do
not use this document as product installation information.
Complete installation, operation, and maintenance information is
provided in the instructions supplied with each product.
Failure to comply with these instructions could result in death or serious
injury.
Mass Airflow Sensors
AWM720P1 Airflow
AWM700 Series
2 Honeywell Sensing and Control www.honeywell.com/sensing
PERFORMANCE SPECIFICATIONS
Flow Range (Full Scale) + 200 SLPM
Min. Typ. Max. Units
Excitation (1) 9.990 10.000 10.010 Vdc
Power Supply 8.000 10.000 15.000 Vdc
Power Consumption 60 mW
Output Load
Sinking 10 mA
Sourcing 20 mA
Calibration gas Air
Null Voltage Shift
+25°C to -25°C,
+25°C to +85°C
± .025 typ.
Vdc
Full Scale Output Shift
+25°C to +10°C
-2.0 % Reading
+25°C to +40°C
+2.0 % Reading
Ratiometricity Error (1)
± 0.30 typ.
% Reading
Repeatability and Hysteresis (2)
± 0.50
% Reading
Response Time 6 typ. ms
Pressure Drop
@ Full Scale
1.0 typ.
2,5 typ.
inch H
2
O
mBar
Overpressure 25 max. psi
Temperature Range
Operating
-25°C to +85°C [-13°F to +185°F]
Storage
-40°C to +90°C [-40°F to +194°F]
Weight 34 [1.20 oz] gram
Connector—4 pin receptacle AMP 103956-3 (provided with sensor)
FLOW SPECIFICATIONS
Flow Nominal
±Tolerance
Pressure Drop
(SLPM) (Vdc) Typical (Vdc) (inch H
2
O) (mBar)
0 1.00 0.05
0
0
25 2.99
0.04
0.10
50 3.82 0.18
0.13
0.33
75 4.30
0.21
0.53
100 4.58
0.34
0.85
150 4.86
0.65
1.64
200 5.00 0.36
1.09
2.74
Notes: 1. Output voltage is ratiometric to supply voltage.
2. Repeatability and Hysteresis tolerances reflect inherent inaccuracies of the measurement equipment.
Mass Airflow Sensors
AWM720P1 Airflow
AWM700 Series
www.honeywell.com/sensing Honeywell Sensing and Control 3
Voltage Output vs. Flow
0.00
0
1.000
2.000
3.000
4.000
5.000
6.000
0 25 50 75 100 125 150 175 200
Airflow (SLPM)
Vo
(Vdc)
Pressure Drop vs. Flow
0.000
0.200
0.400
0.600
0.800
1.000
1.200
02550
75
100 125 150 175 200
Airflow (SLPM)
Pressure
(inch H
2
O)
Figure 1. Voltage Output vs. Airflow Figure 2. Pressure Drop vs. Airflow
OUTPUT CURVES
Figure 1, Voltage Output vs. Airflow and Figure 2,
Pressure Drop vs. Airflow depict performance
characteristics for the AWM700 Series sensors at
10.0 ±0.01 Vdc at 25ºC.
NOTICE
LAMINAR FLOW
Due to the fast response time of the sensor, these
specifications were generated using laminar flow.
Airflow instability or “turbulence” present in the
airstream will result in an increase in measurement
uncertainty.
The turbulent flow problem can be corrected by either
straightening the airflow using flow laminarizing or by
slowing the response of the sensor using a simple RC
time constant on the output of the sensor. This, of
course, slows down the sensor response time. The
values needed depend on the amount of turbulence
present in the application.
Several techniques for laminarizing the flow include
adding hex shaped honeycombs, foam, screen
materials or adding constrictors (frits) to the flow
stream. There are various commercial laminar flow
elements that can be purchased. Unfortunately the
greater the efficiency of the laminarizer, the greater
the increase in pressure drop in order to establish a
given flow rate. Plastic honeycomb material probably
gives the most improvement for the least pressure
drop. In any test fixture, the avoidance of sharp radii is
an absolute requirement.
Failure to comply with these application
instructions may result in product failures.
ELECTRICAL CONNECTION
The AWM700 Series accepts a latch detent
connector, such as: AMP part number 103956-3.
Information on latch detent connectors is available
from the AMP Product Information Center, 1-800-
522-6752 or the AMP Customer Hotline,
1-800-722-1111.
RELATED AMP LITERATURE
82160 MTE Interconnection System
(AMPMODU) Catalog
108-25034 Product Specification
(technical performance information)
114-25026 Application Specification
(describes product, proper assembly,
full tooling information)
IS 6919 Instruction Sheet for assembly
procedure
MAKING ELECTRICAL CONNECTIONS
1. Remove (unlatch) the connector from the
AWM700.
2. Hand-crimp the interface wire to the appropriate
pin on the connector. One possible tool: AMP
Hand-Crimp Tool, part number 58342-1.
3. Insert the terminal contacts into the connector
housing after carrier strip (lead-frame) is
removed.
4. Reconnect (latch) connector to AWM700 device.
MOUNTING INSTRUCTIONS
Mount AWM700 Series sensors with 6-32 screws.
Honeywell recommends use of washers below
screw head. Mounting torque is 0.68 N m [6.0 in lb]
max.

AWM720P1

Mfr. #:
Manufacturer:
Description:
Flow Sensors Amplified; 200 SLPM Tapered 22.0 mm
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet