Operator’s Manual
5
Do not overload. Do not apply any potential to the probe leads that exceeds
the maximum rating of the probe.
Always comply with the Voltage vs.
Frequency Derating Curve.
Observe all terminal ratings. Observe all markings on the oscilloscope
before connecting. Consult the oscilloscope product manual for further
ratings information.
Do not remove probe casing. Touching exposed connections may result in
electric shock.
Use indoors only within operational environment listed. Do not use in wet or
explosive atmospheres. Keep product surfaces clean and dry.
Handle with care. Probe tips are extremely sharp and may puncture skin or
cause other bodily injury if not handled properly.
Keep fingers behind the finger guard of probe body and accessories.
Do not operate with suspected failures. Before each use, inspect the probe
and accessories for any potential damage such as tears or other defects in
the probe body, cable jacket, accessories, etc. If any part is damaged, cease
operation immediately and sequester the probe from inadvertent use.
Cleaning
Clean only the exterior of the probe, cables, and accessories. Use a
soft cotton cloth moistened with a mild detergent and water solution.
Do not allow any portion of the probe to be submerged at any time.
Dry the probe and accessory thoroughly before attempting to make
any voltage measurement.
Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body, cables, and accessories.
PP023 Passive Probe
6
Probe Compensation
Proper compensation of the probe is required to assure amplitude accuracy
of the waveform being measured by matching the probe to the
oscilloscope’s input capacitance. Compensation should be adjusted
whenever the probe is connected to or transferred between oscilloscopes.
Low Frequency
1. Apply a 1 kHz square wave to the probe or connect to the
oscilloscope’s calibrator output.
2. Adjust the single LF trimmer located on the PP023 BNC Box until you
achieve a flat-topped square wave (see figure below):
High Frequency
1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time).
2. Remove the plastic caps from the two, high-frequency trimmers on
the PP023 BNC Compensation Box.
3. Adjust HF2 then HF1 until you achieve a flat-topped square wave (see
figure below).
Operator’s Manual
7
Over-compensated, correct, and under-compensated waveforms
Probe Accessories
Probe Tips
Both the rigid probe tip and the spring tip can be screwed on to the probe
body by hand.
Be sure not to overtighten the tips to avoid damaging
the threads.
Ground Accessories
The ground lead and the short collar are attached to the probe at the time
of shipping. To replace the ground lead with the swivel lead adapter,
unscrew the short collar from the probe and remove the ground lead.
Replace the short collar to hold the swivel lead adapter in place.
Alternatively, the long collar can be pushed onto the probe body and used in
conjunction with the ground spring, BNC adapter or PCB adapter.
Service
Refer all service to qualified personnel.
For the most complete and up-to-date list of sales and service centers by
country, visit teledynelecroy.com/support/contact.

PP023-2

Mfr. #:
Manufacturer:
Description:
OSCOPE PROBE X10 500MHZ 10M 2PK
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet