DCM20-1-DC5-R-C

Murata Power Solutions’ DCM20 series DC power meters display the most critical mea-
surements in DC-powered equipment: voltage, amperage and power. A front panel selector-
switch provides two modes of operation: a fixed reading of any of the three parameters, or
a continuous cycling through all three measurements. In fixed-reading mode, the function
to be displayed (V/A/W/kW) is selected by a capacitive touch sensor, avoiding the wear-out
issues often seen with membrane or other mechanical control switches.
The DCM20 typically requires 6 mA of operating current at a supply voltage of 12 VDC, and
only 2 mA at 72 VDC. This feature allows it to be added to existing systems with minimal
electrical loading effects.
The DCM20 can be powered from a DC power supply ranging from 9 to 72 VDC. The meter
supports two fundamental voltage measurement configurations. In the first configuration,
the DCM20 is powered from the DC power supply line which it is monitoring. When used in
this way, it can provide voltage measurements in the range of 9 - 72 VDC, as it is unpow-
ered at lower voltages. The DCM20 may also be configured such that it is powered from a
separate power supply than that which is monitoring. When used in this way, it can support
voltage measurements over the entire 0.5 to 72 VDC range.
For current and power measurements, DCM20 series DC power meters require the use of
an external shunt resistor. The DCM20’s wide common-mode input range (0 to 72 V) sup-
ports both high-side and low-side current sensing in a variety of applications. A 4-position
DIP switch supports field selection of 16 shunt calibration factors. Additionally, a 12-turn
trim potentiometer is also provided on the rear of the meter to support precision field
calibration of the meter to the actual shunt value.
A large (0.36 in/9.2 mm), bright LED display makes DCM20 DC power meters easily
readable from as far as 15 feet (5 meters). All units are packaged in a one-piece, vibration
resistant, polycarbonate housing that can be installed in 0U rack components and vertical
rack spaces as small as 1U. Their miniature size is perfect for laboratory instrumentation,
alternative energy applications, standby power supplies and any other products that require
precise monitoring of DC voltage, current, and power.
PRODUCT OVERVIEW
FEATURES
Displays DC volts, amperes, watts and kilowatts.
Two display modes: continuous auto cycling or fixed.
9 - 72 VDC voltage operation.
0.5 - 72 VDC voltage measurement range.
0 -1200 A current measurement range (dependent on
external shunt).
0 - 96 kW power measurement range.
Supports both high-side and low-side current monitoring.
Long-life contactless touch-sensor function selector
switch.
Low current consumption (< 10 mA at 12 VDC).
Internal self-resetting fuse
One-piece polycarbonate housing fits ‘0U’ & ‘1U’ racks.
Compact surface mount design occupies minimal panel
space.
Bright LED-display with four function annunciators.
Choice of red, green, or blue display.
SIMPLIFIED SCHEMATIC DIAGRAM
J1
VOLTS AMPS WATTS kW
DC-DC
CONVERTER
THERMAL
FUSE
ADC
(VOLTAGE)
MICROCONTROLLER
DISPLAY
DCM20
ADC
(CURRENT)
SHUNT RANGE
SELECTOR SWITCH
SHUNT TRIM
ADJUST
TOUCH
SENSOR
SEL
SIGNAL
CONDITIONER
1
2
3
4
+V
–V
+IN
–IN
+5 V
–5 V
3
2
1
TB1
+
+
+
9 – 72
VDC
CURRENT-LIMITED
EXTERNAL POWER SOURCE
EXTERNAL
SHUNT
Figure 1. Simplified Schematic
www.murata-ps.com
www.murata-ps.com/support
For full details go to
www.murata-ps.com/rohs
DCM20 Series
Three-Function DC Power Meters
MPM_DCM20Series.A01.D5 Page 1 of 6
Performance/Functional Specifications
Typical at TA = +25ºC , V+ = 12 VDC unless otherwise specified
ORDERING INFORMATION
DCM20-1-DC5-R-C (Red Display)
DCM20-1-DC5-G-C* (Green Display)
DCM20-1-DC5-B-C* (Blue Display)
*Consult factory for availability
TECHNICAL NOTES
IMPORTANT! To ensure safe and reliable operation, DCM20 power
meters must be installed and serviced by qualified technical personnel.
Contact Murata Power Solutions if there is any doubt regarding their
installation or operation.
1. Measurement Type: DCM20 series multifunction DC power meters
employ precision ADCs and a low-power microcontroller to measure and
display DC voltage, current and power. The choice of measurement is
selectable through the front panel. The meters measure DC current using
an external series shunt resistor which is inserted into the circuit to be
measured. Because the DCM20 has a wide common-mode input range
(0 - 72 VDC), the shunt resistor may be inserted into either the high-side
(supply) or the low-side (return) of the circuit being monitored.
DC Voltage may be monitored either at the meter’s +IN (TB1-3) input
terminal or at the meter’s +V (TB1-1) power input terminal, based on
the setting of jumper block J1. The choice of voltage monitoring input is
selected based on the application requirements.
The DCM20 derives its DC power measurement by multiplying the
measured current by the measured voltage. Power is always reported
as a positive quantity, regardless of the polarity of the measured current.
(Power = | Voltage × Current | ).
2. Basic Operation: Upon first application of power to the meter, the unit
will first perform a self-test routine and then continuously display DC
volts with the VOLTS LED annunciator illuminated. The display will remain
in the VOLTS mode as long as the front panel ‘SEL’ button is not touched.
After the unit powers up to normal operation in the VOLTS reading mode,
momentarily (approximately one second) touching the ‘SEL’ button on
the unit’s front panel twice in succession will cycle the display to AMPS,
and then to WATTS (or kW, depending on the power measurement value).
Momentarily touching ‘SEL’ a third time will return the display back to
the VOLTS reading mode. Holding the ‘SEL’ button down for 5 seconds
will place the unit in a continuous auto-cycling mode, and the display will
repetitively scroll through all three measurements, with each measure-
ment remaining displayed for 3 seconds.
When the continuous auto-cycling mode is initially selected, the unit will
briefly display ‘Auto On’ before continuous cycling begins. Momentarily
touching the ‘SEL’ button again will cause the unit to briefly display ‘Auto
OFF’ before it returns to the fixed VOLTS reading mode.
The power-up default display mode (Volts/Amps/Watts/Auto-cycle) may
be selected by simply leaving the DCM20 in a given display mode for 60
seconds or more. The next time the meter powers up, it will do so in the
new default mode. The default mode may be reset by selecting a new
operating mode and leaving the meter in that mode for 60 seconds or
longer.
3. Calibration: The DCM20’s voltage measurement function is factory-
calibrated and cannot be field-calibrated. Because the meter is designed
to work with a wide range of shunt resistors for current measurement,
however, field configuration and fine adjustment of the current measure-
ment function is supported. The DCM20’s shunt scaling factor is selected
by a 4-position DIP switch SW1 on the rear of the unit (refer to figure 9,
Mechanical Specifications). A total of 16 ranges are supported. Table 1
lists supported current ranges and their associated switch settings, and
also includes recommended Murata shunt resistors for selected ranges.
Additionally, a trimmer potentiometer on the rear of the unit supports fine
adjustment of the shunt scaling factor to accommodate tolerance errors
!
Measurement Limits Min Typ Max Units
DC Voltage
(A) 0.5 V 72 V DC
DC Current
(B)
Range ‘1’ - 10 mΩ external shunt
Range ‘16’ – 41.67 µΩ external shunt
-5
-1200
5
1200
A DC
Current Shunt Differential Voltage
Range
(C)
-100 +100 mV DC
Current Shunt Common-mode Voltage
Range
(D)
0 72 V
Current Fine Trim Range using Trim
Potentiometer
±1%
Power
Watts Display Mode
Kilowatts Display Mode
0
0
9999
96
W DC
kW DC
Performance
Display Update Rate 2.5 updates/sec.
Voltage Accuracy ± 1% ± 2%
% Full-
scale
Current Accuracy ± 2% ± 3%
Power Accuracy ± 2% ± 3%
Temperature Drift (0 - 60) +/- 0.5 Counts/°C
Zero Current Reading (within 1 sec.) +/- 4 Count
Operating Supply Voltage 9 72 V DC
Operating Supply Current
(E)
12 V operation
72 V operation
6
2
10
5
mA
Display
Display Type and Size
4 Digit LED, 9.14 mm (0.36 in) Height
4 LED Annunciators (‘Volts,’ ‘Amps,’
‘Watts,’ ‘kW’)
Over-range Indication Flashing Display
Decimal Point Selection Automatically Set
TB1 Terminal Block
Wire Size
Copper Wire
14-24 AWG
Insulation Strip Length 6mm (0.24 in)
Screw Tightening Torque 4.6 in - lb (0.51 N-m)
Physical/Environmental
Operating Temperature 0 +60 °C
Storage Temperature -40 +75 °C
Humidity (non-condensing) 0 85 %
Dimensions See figure 6
Weight
Unit Weight: 1.02 oz / 29g
Net Package Weight: 12.28 oz / 348g
Gross Package Weight: 22.05 oz / 625g
Hex Nut Tightening Torque 20 in - oz (0.14 N-m)
(A) Voltage measured across +V (TB1-1) and –V (TB1-2) OR +IN (TB1-3) and –V (TB1-2) terminals.
(B) Current measurement range dependent on external shunt selection.
(C)
Applied to +IN (TB1-3) and –IN (TB1-4) terminals.
(D)
Voltage applied across +IN (TB1-3) and –V (TB1-2) terminals.
(E)
Current drawn by +V (TB1-1) terminal when operating voltage applied across +V (TB1-1) and
–V (TB1-2) terminals.
www.murata-ps.com/support
DCM20 Series
Three-Function DC Power Meters
MPM_DCM20Series.A01.D5 Page 2 of 6
in the shunt value. Because adjusting the meter using this potentiometer
requires that the DCM20 be powered up and measuring current, this
operation should only be performed by qualified personnel using an
insulated adjustment tool.
4. Shunt Selection: Table 1 lists recommended current shunts avail-
able from Murata Power Solutions for current ranges supported by the
DCM20. Please note that a given shunt may not necessarily support
continuous operation at the extreme limits of its measurement range.
The maximum current that a shunt may be safely operated at is also
highly dependent on application conditions such as ambient temperature,
airflow and mounting position. Before selecting and installing a shunt,
you should refer the shunt’s datasheet for details on correct operation.
5. Jumper Configuration: The setting of jumper J1 determines the source
of voltage to be measured. When the shorting block is placed over posts
J1-1 and J1-2, the DCM20 measures the voltage between the +IN (TB1-
3) connector terminal and the -V (TB1-2) connector terminal. When the
shorting block is placed over posts J1-2 and J1-3, the DCM20 measures
the voltage between the +V (TB1-1) and -V (TB1-2) connector terminals.
The meter must have the shorting block placed on jumper J1 for normal
operation. Jumper block J1 is located on the rear of the meter as shown
in figure 9 (mechanical drawing). The DCM20 is shipped with the shorting
block placed across jumper pins J1-2 and J1-3.
6. Wiring: Power supply and input wiring must be rated for the electrical
and environmental conditions under which the meter will be operated.
They must also comply with any regulatory or application-mandated
requirements pertaining to the user’s installation. Connections to DCM20
meters must be made with all power sources de-energized. Refer to the
Functional Specifications section for detailed wire type and wire gauge
information.
Wiring insulation should be stripped to 6 mm (0.24 in). All wires must be
inserted into the terminal block openings such that the screw terminal
does not pinch any insulation. After final assembly, inspect all terminal
block connections for shorts between adjacent conductors; this step is
particularly important when using stranded wire.
7. Protection and Fusing: DCM20 meters’ power supply leads are protect-
ed against momentary overvoltage ((+V - -V) > 72 V) and reverse polarity
conditions ((+V - -V) < 0 V ). The +V (TB1-1) power supply line is also
protected with an internal self-resetting thermal fuse. The meter’s +IN
(TB1-3) and –IN (TB1-4) terminals are protected against brief voltage
transients. Neither set of inputs is protected against sustained conditions
exceeding the limits listed in in the Performance & Functional Specifica-
tions table. External fusing must be supplied by the user in accordance
to applicable safety and regulatory requirements for the system in which
the DCM20 is installed.
8. Shunt Polarity: While the external current shunt has no intrinsic ‘polar-
ity’, the way in which it is connected into the circuit being measured will
influence the polarity of current as measured by the meter. The relation-
ship between current direction and meter reading is shown in figure 2.
9. Terminal Block Wire Size & Torque Ratings: It is important to tighten
all screw-terminals to their torque specification of 4.5 in - lb (0.51 N-m).
Proper tightening will help ensure reliable operation. Applications subject
to vibration should use stranded wire.
10. Typical Wiring Diagrams: The DCM20 is a flexible voltage, current, and
power measurement device, supporting both high-side and low-side
shunt configurations. This section describes four of the most common
ways in which the DCM20 may be used.
In figure 3, the current shunt is inserted into the load’s positive supply
line, which is known as high-side measurement. The DCM20 is also
powered from this line. In this configuration the shorting block on jumper
J1 should be connected across pins J1-1 and J1-2. Because the meter
is powered from the line being monitored, useful measurements can
only be made on DC line voltages in the range of 9 - 72 VDC, as at lower
voltages the meter will not be powered. In this configuration, the DCM20
measures the sum of both the load power and the power dissipated by
the shunt.
Range SW1 Setting
Full-scale
Current
Range (A)
Full-scale
Power
Range
Shunt
Voltage
(mV)
Recommended
Murata Shunt
Resistor P/N
1
1
5.000 400.0 W 50 3020-01097-0
2
1
10.00 800 W 100 3020-01107-0
3
1
20.00 1600 W 50 3020-01098-0
4
1
50.00 4000 W 50 3020-01096-0
5
1
50.00 4000 W 100 --
6
1
75.00 6000 W 50 --
7
1
100.0 8000 W 50 3020-01099-0
8
1
100.0 8000 W 100 3020-01108-0
9
1
150.0 2.00 kW 50 3020-01100-0
10
1
200.0 6.00 kW 50 3020-01101-0
11
1
300.0 24.00 kW 50 3020-01102-0
12
1
400.0 32.00 kW 50 --
13
1
500.0 40.00 kW 50 3020-01103-0
14
1
800.0 64.00 kW 50 3020-01104-0
15
1
1000 80.00 kW 50 3020-01105-0
16
1
1200 96.00 kW 50 3020-01106-0
Table 1. Shunt Range Selection
Figure 2. Relationship between shunt voltage polarity
and meter reading
–IN+IN
+
SHUNT VOLTAGE
METER READS
POSITIVE AMPS
METER READS
NEGATIVE AMPS
–IN+IN
+
SHUNT VOLTAGE
VOLTS AMPS WATTS kW VOLTS AMPS WATTS kW
www.murata-ps.com/support
DCM20 Series
Three-Function DC Power Meters
MPM_DCM20Series.A01.D5 Page 3 of 6