GSM28-24G

Temperature Coecient
0.03%/°C typical on all outputs.
Environmental
Designed for 0 to 50°C operation at full rated output power; derate output current
and total output power by 2.5% per °C above 50°C.
Medical EMI/EMC Compliance
All models include built-in EMI ltering to meet the following emissions require-
ments:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated output load is 0.85 A.
Hold-up Time
15 ms minimum from loss of ac input at full load, nominal line (120 Vac).
Output Power
Normal continuous output power is 28 W, 32 W peak for 60 s maximum dura-
tion, 10% duty cycle. Factory set to begin power limiting at approximately 35
W.
Overload Protection
Fully protected against short circuit and output overload. Short circuit pro-
tection is cycling type power limit.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, dierential mode. Measured with
scope probe directly across output terminals of the power supply with load
terminated with 0.1 μF capacitor.
Overvoltage Protection
Built in with ring point set per ratings table. OVP ring reduces voltage to
less than 50% of nominal voltage in 50 ms.
Voltage Adjustment
Factory set with xed resistors to maximize reliability.
Eciency
70% minimum for the 5.1 V model at full rated load, nominal input voltage.
Eciency increases as output voltage increases.
Input Protection
Internal ac fuse provided on all units. Designed to open only if a catastrophic
failure occurs in the unit.
Inrush Current
Inrush limited by internal thermistor. The inrush at 230 Vac, averaged over the
rst ac half-cycle under cold start conditions will not exceed 32 A.
Minimum Load Requirement 5% of full load rating
Transient Response 3.5% max. dev. 50% load step at 0.2 A/
μs. Recovery to
0.5% within 500 μs.
GSM28 Medical
28 Watt Global Performance Switchers
• Industrys smallest 28 W medically approved switcher
• Compact size (4.00” x 2.59” x 0.92”)
Wide-range ac input: 85-264 Vac
• Less than 25 μA leakage current @ 120 Vac
• Approved to UL2601-1, EN60601-1
• EMI to FCC, CISPR 11 Class B/IEC601-1-2
• Overvoltage protection standard
• RoHS Compliant Models Available (G sux)
GLOBAL PERFORMANCE SWITCHERS
All specications are typical at nominal input, full load at 25
O
C unless otherwise stated
FEATURES:
D C P o w e r S u p p l i e s
SPECIFICATIONS
• marked to LVD
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Line Frequency Harmonics
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV di., 2 kV com.
EN61000-3-2 Class A
Earth Leakage Current
Leakage current measured in the Gnd wire connection when measured per
EN60601-1 or UL2601-1 is as follows:
Medical Medical Safety
SL Power Electronics Corp. declares under our sole responsibility that all GSM
models are in conformity with the applicable requirements of UL2601-1 Patient
Care Equipment, CSA-C22.2 No. 234 (with additional tests to C22.2 No. 601.1 per
T.I.L. CA-08), EN60601-1.
Medical Normal SingleFault Test Test
Model Leakage Leakage Voltage Method
GSM28 25μA 45 μA 132 Vca/60 Hz UL2601-1
GSM28 50 μA 90 μA 264 Vca/50 Hz IEC60601-1
SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 1/07.
Data Sheet © 2007 SL Power Electronics Corp. The information and specications contained in this data sheet are believed to be correct at time of publication.
However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specications are subject to change without notice.
ENVIRONMENTAL SPECIFICATIONS OPERATING NON-OPERATING
Temperature (A) See individual specs -40 to +85°C
Humidity (A) 0 to 95% RH 0 to 95% RH
Shock (B) 20 g
pk
40 g
pk
Altitude -500 to 10,000 ft -500 to 40,000 ft
Vibration (C) 1.5 g
rms’
0.003 g
2
/Hz
5 g
rms’
0.026 g
2
/Hz
GSM28 MECHANICAL SPECIFICATIONS
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 or-
thogonal axes, total 6 shocks.
C. , Random vibration—10 to 2000Hz, 6dB/octave roll-o from 350 to
2000Hz 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis
non-operating.
Medical Model Output Current Load Regulation Initial Setpoint Tolerance OVP Setpoint Ripple and Noise
GSM28-5 5.1 V 5.5 A 0.75% 2.5% 6.2 ± 0.6 V 1.4%
GSM28-12 12 V 2.3 A 0.75% 2.5% 14 ± 1.0 V 1%
GSM28-15 15 V 1.9 A 0.75% 2.5% 18.5 ± 1.5 V 1%
GSM28-24 24 V 1.2 A 0.75% 2.5% 28 ± 2.5 V 1%
GSM28-28 28 V 1.0 A 0.75% 2.5% 34 ± 2.8 V 1%
* Add “G” sux to model number for RoHS compliant model.

GSM28-24G

Mfr. #:
Manufacturer:
Description:
Switching Power Supplies 28W 24V @ 1.2A