TPP 40-124A-J

www.tracopower.com Page 1 of 4
AC/DC Medical Power Supply TPP 40A Series, 40 Watt
Models
The TPP 40A Series of 40 Watt AC/DC power supplies feature a reinforced dou-
ble I/O isolation system according to latest medical safety standards IEC/EN ES
60601-1 3rd edition for 2  MOPP up to 5000 m MSL. The earth leakage current
is below 75 A what makes the units suitable for BF (body floating) applications.
The excellent efciency of up to 92% allows a high power density for the stan-
dard 2.44” x 3.0” packaging format. The full load operating temperature range
is –4C to +70°C while it goes up to 85°C with 50% load derating. The EMC
characteristic complies to IEC 60601-1-2 ed.4 and is dedicated for applications
in industrial and domestic fields. High reliability is provided by use of industrial
quality grade components and an excellent thermal management. It makes the
products an ideal solution for medical devices and for demanding safety and
space critical applications.
Order code Output voltage Output current max. Efficiency max.
TPP 40-105A-J
5 VDC 8.0 A 90 %
TPP 40-112A-J
12 VDC 3.34 A 92 %
TPP 40-124A-J
24 VDC 1.67 A 92 %
TPP 40-148A-J
48 VDC 0.84 A 92 %
Models
prepared
C B
Scheme
ES 60601-1
Open frame power supply pin connector
Certification according to IEC/EN/ES
60601-1 3rd edition for 2MOPP
Low leakage current <75 A rated for
BF applications
Risk management process according to
ISO 14971 including risk management file
Acceptance criteria for electronic assem-
blies according to IPC-A-610 Level 3
EMC emission to IEC 60601-1-2 ed.4
Protection class I & ll
Approved for operation up to 5000 m MSL
Ready to meet ErP directive, < 0.15 W
no load power consumption
5 year product warranty
Encased version with screw terminal
connection see TPP 40 Series
www.tracopower.com/overview/tpp40
Note: - Output is ajustable by ±10% with internal potentiomet
- Other output voltages are available on request
TPP 40A Series, 40 Watt
www.tracopower.com Page 2 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Input Specifications
Input voltage range – AC range (universal input) 85 – 264 VAC
– DC range 120 – 370 VDC
Input frequency 47 – 63 Hz
Input current at full load – at 115 VAC / 230 VAC 1.05 A max. / 0.55 A max.
Input protection T3.15 A/250 VAC (internal fuse in both line &
neutral)
Input inrush current – at 230 VAC 60 A max.
Zero load power consumption 0.15 W max. (acc. ErP directive)
Output Specifications
Voltage set accuracy ±1%
Regulation – Input variation 0.2% max.
– Load variation (0 - 100%) 5 VDC model: 0.7% max.
other models: 0.5% max.
Minimum load not required
Temperature coefficient 0.02%/K
Hold-up time – Vin = 115 VAC 25 ms typ.
Start-up time <1 s
Rise time 20 ms typ.
Ripple and noise 5 & 12 VDC output: 75 mVp-p typ.
w. cap. 10F/25V 1206 X7R MLCC
(20MHz Bandwidth) 24 VDC output: 75 mVp-p typ.
w. cap. 1F/50V 1206 X7R MLCC
48 VDC output: 150 mVp-p typ.
w. cap. 0.1F/100V 1206 X7R MLCC
Overvoltage protection 125 – 140% of nominal Vout
Current limitation at 145% Iout typ.
Short circuit protection hiccup mode (automatic recovery)
Transiente response – Peak deviation 3% max. (25% load step change)
– Recovery time 600 s typ.
Capacitive load 5 Vout model: 16‘000 F max.
12 Vout model: 2‘785 F max.
24 Vout model: 700 F max.
48 Vout model: 175 F max.
TPP 40A Series, 40 Watt
www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
General Specifications
Operating temperature –40°C to +85°C
with derating
Output power de-rating 2.67 %/K above +70°C at 230 VAC
2.75 %/K above +65°C at 115 VAC
Storage temperature –40°C to +85°C
Humidity (non condensing) 5 – 95 % rel. H max.
Altitude during operation 5000 m
Switching frequency - single output 5 VDC model: 70 kHz typ. (pulse frequency modulation)
(at 230 VAC) other models: 120 kHz typ. (pulse frequency modulation)
Isolation voltage – Input / Output (60 s) 4000 VAC
(2  MOPP insulation) – Input / Case (60 s) 2500 VAC
Leakage current (at 264 VAC/60Hz) 75 A max.
Isolation resistance (at 500 VDC) 100 MOhm min.
Reliability – calculated MTBF at +25°C acc. to IEC 61709 3000‘000 h
Protection class class II prepared
Electromagnetic compatibility – Conducted & Radiated input surpression EN 55011 limits to IEC 60601-1-2 4th editon
(EMC), emissions EN 55032 class B (internal filter)
– Harmonic current emissions IEC / EN 61000-3-2, class B
– Voltage flicker IEC / EN 61000-3-3, class B
Electromagnets compatibility (EMC), immunity IEC / EN 60601-1-2
– Electrostatic discharge ESD IEC / EN 61000-4-2, 8kV/6kV perf. criteria A
– RF field immunity IEC / EN 61000-4-3, 20V/m perf. criteria A
– Electrical fast transients/burst immunity IEC / EN 61000-4-4, ± 2kV perf. criteria A
– Surge IEC / EN 61000-4-5, ± 1kV/± 2kV perf. criteria A
– Conducted RF IEC / EN 61000-4-6, 20 Vrms perf. criteria A
– Magnetic field IEC / EN 61000-4-8, 10A/m perf. criteria A
Voltage dip and interruptions according to EN 60601-1-2 30%, 500ms perf. criteria A
reference: 100 VAC / 50Hz 60%, 100ms perf. criteria B
> 95%, 10ms perf. criteria A
> 95%, 5000ms perf. criteria B
Safety standards and certification IEC/EN 60601-1 3rd edition,
ANSI/AAMI ES60601-1:2005(R)2012
– Certification documents www.tracopower.com/overview/tpp40a
Environment – Vibration acc. IEC 60068-2-6 3 axis, sine sweep, 10–55Hz, 1g, 1oct/min
– Shock acc. IEC 60068-2-27 3 axis, 10g half sine, 11msShock
20 G (3 directions each 3 times)
Environmental compliance – Reach www.tracopower.com/products/reach-declaration.pdf
– RoHS RoHS directive 2011/65/EU
Connection JST pin connector