ME90A1803F01

ME90
90W Single Output External Power
Medical Series
www.slpower.com ME90 Rev 2.1 9-NOV-17 1
Features
Meets UL/EN/IEC60601-1-2, 4th edition for
EMC
*
Approved to EN/IEC/UL60601-1, 3rd edition
with isolation levels which satisfy the 2 MOPP
requirements
Meets DoE Efficiency Level VI Requirements
No load input power
Average Efficiency
Up to 90W of AC-DC Power
Universal Input 90-264Vac Input Range
Desktop Style Package
Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin
E-cap life of >7 years
3 Year Warranty
IP22 Rated Enclosure
Description
A high performance AC to DC external power supply family designed for medical applications. The ME90 Medical Series
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels which satisfy the 2
MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC
*
. The ME90 Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up
to +40°C and applicable output power derating at 50°C. These models are available in desktop versions, include an IP22
rating per IEC60529 for the enclosure, and the output cable can be terminated at a variety of output connectors.
*Professional Equipment only. Consult Factory for Table 9 compliance information.
Model Selection
Model
Output
Output
Ripple &
Line
Load
Output
Input
Number Volts Current Power
Noise
¹
Regulation Regulation
Connector
ME90A1251F01 12.0V 7.50A 90W 120mV pk-pk
±1% ±5%
6 pin Molex Type
2
Class I Desktop,
IEC60320 C14
Receptacle
ME90A1503F01 15.0V 6.00A 90W 150mV pk-pk
±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
ME90A1803F01 18.0V 5.00A 90W 180mV pk-pk
±1% ±5%
ME90A2403F01 24.0V 3.75A 90W 240mV pk-pk
±1% ±5%
ME90A1251N01 12.0V 7.50A 90W 120mV pk-pk
±1% ±5%
6 pin Molex Type
2
Class II Desktop,
IEC60320 C8
Receptacle
ME90A1503N01 15.0V 6.00A 90W 150mV pk-pk
±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
ME90A1803N01 18.0V 5.00A 90W 180mV pk-pk
±1% ±5%
ME90A2403N01 24.0V 3.75A 90W 240mV pk-pk
±1% ±5%
ME90A1251Q01 12.0V 7.50A 90W 120mV pk-pk
±1% ±5%
6 pin Molex Type
2
Class II Desktop,
IEC60320 C18
Receptacle
ME90A1503Q01 15.0V 6.00A 90W 150mV pk-pk
±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
ME90A1803Q01 18.0V 5.00A 90W 180mV pk-pk
±1% ±5%
ME90A2403Q01 24.0V 3.75A 90W 240mV pk-pk
±1% ±5%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Molex p/n 39-01-2060 or equivalent. See outline drawing for pinout information.
3. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME90B1251F01).
ME90
90W Single Output External Power
Medical Series
www.slpower.com ME90 Rev 2.1 9-NOV-17 2
General Specifications
Parameter Specification Parameter Specification
AC Input 100-240Vac, ±10%, 47-63Hz, 1 Turn On Time
Less than 1 sec @115Vac, full load
Input Current
115Vac: 2.0A, 230Vac: 1.0A
Hold-up Time
20mS min., at full Load, 100Vac input
Inrush Current
264Vac, cold start: will not exceed 60A
Overtemperature
Protection
Will shutdown upon an overtemperature
condition, auto-recovery.
Input Fuses
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
Overload Protection
130 to 180% of rating, Hiccup Mode
Earth Leakage
Current (Input to Ground)
<500μA@264Vac, 60Hz, NC
<1mA@264Vac, 60Hz, SFC
Short Circuit
Protection
Hiccup Mode, auto recovery.
Patient Leakage
Current (Output to Earth)
<100μA@264Vac, 60Hz, NC
<500µA@264Vac, 60Hz, SFC
Overvoltage
Protection
130 to 150% of output voltage, hiccup mode
Efficiency
>88%, typical
Isolation
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
Output Power
90W continuous – See models chart for
specific voltage model ratings.
Safety Standards
EN/IEC/UL60601-1, 3rd edition
No Load Input
Power
<0.210W (meets DoE Efficiency Level VI
Requirements)
Operating
Temperature
-20°C to +50°C. Derate above 40°C.
Start Up at -40°C, full load, (warmup period
before all parameters are within published specifications).
Ripple and Noise
See models chart on pg 1.
Storage Temperature
-40°C to +85°C
Output Voltage
See models chart on pg 1.
Altitude
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Transient
Response
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, i/t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Relative Humidity
5% to 95%, non-condensing
Regulation
See models chart on pg 1.
Drop Test
1.4m from table top to wooden platform, 4
faces.
Vibration
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of
10 sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
Dimensions
W: 2.58” x L: 5.9” x H: 1.34”
W: 65.5mm x L: 150.5Mm x H: 34mm
MTBF
>250,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
Weight
600g
E-Cap Life
>7 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient
25°C at 24 hrs per day, 365 days/year, 6
power up cycles per day. (80% load on 5V, 12V
model)
ME90
90W Single Output External Power
Medical Series
www.slpower.com ME90 Rev 2.1 9-NOV-17 3
EMI/EMC Compliance
Parameter
Specification
Conducted Emissions:
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4
th
Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55024/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4
th
Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4
th
Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4
th
Edition requirements.
Conducted Disturbances induced by RF
Fields
EN55024/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4
th
Edition, Table 5.
Rated Power frequency magnetic fields
EN55024/IEC1000-4-8, Level 4: 30A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Voltage Interruptions, Dips, Sags & Surges
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225, 270 and 315
degrees, Criteria A; 100% dip for 20mS, Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Note: Performance criteria are based are defined as following:
A – Normal performance during and after the test
B – Temporary degradation, self-recoverable
C – Temporary degradation, operator intervention required to recover the operation
D – Permanent damage

ME90A1803F01

Mfr. #:
Manufacturer:
Description:
Power Supply, AC-DC, Desktop, Medical, 90W, 18V@5A, Class I, C14
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet