SLG-0150 FMK

www.sensirion.com Version 6 December 2016 4/7
3 Communication with the Sensor
The SLG flow meter shows bidirectional, linear
transfer characteristics. The product comes fully
calibrated for water.
Digital Sampling Time, 16 bit 74 ms
Digital Sampling Time, 9 bit 1 ms
3.1 Electrical Specifications
Table 2: DC Characteristics
Parameter
Conditions
Min.
Typ.
Max.
Units
Power Supply
DC, VDD
Sensor only
4
5
6
V
w. SCC1-RS485
4
5
6
V
w. SCC1-Analog
12
24
36
V
Operating
Current
VDD = 4-6 V
5.1
mA
w. SCC1-RS485
20
70
mA
w. SCC1-Analog
4.3
mA
3.2 Electrical Connector and Pinout
The flow meter is equipped with a male connector
type M8, 4-pin, threaded lock according to
IEC 61076-2-101 (Ed. 1)/ IEC 60947-5-2, and is
compatible with Sensirion’s SCC1 interface cables.
Table 3: Electrical pinout
Pin
1
SDA (data)
2
GND
3
VDD
4
SCL (clock)
3.3 Communication via USB cable
The Sensirion USB Sensor Cable provides an easy to
use USB Interface for laboratory and desktop use.
For further information please see the SCC1-USB
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.4 Digital Communication via RS485-Bus
The SCC1-RS485 Sensor Cable for flow sensors
allows the communication via RS485 interface for use
in a demanding industrial automation environment. In
addition to the standard commands available in the
I²C interface of the sensor, the incorporated
microcontroller of the cable provides more complex
logic such as a dispense volume totalizer, automatic
dispense detection, automatic heater control and data
buffer for asynchronous read-out.
For further information please see the SCC1-RS485
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.5 Analog Communication
The SCC1-ANALOG Sensor Cable allows simple and
quick readout of Sensirion’s liquid flow meters by
converting the digital sensor reading to a 0...10.5 V
analog voltage output. Additionally, a digital (high/low)
output with two different modes of operation is
available (Flow Switch / Volume Counter).
For further information please see the SCC1-Analog
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.6 Digital Communication via I²C-Bus
Digital communication between a master and the SLI
sensor runs via the standard I²C-interface. The
physical interface consists of two bus lines, a data line
(SDA) and a clock line (SCL) which need to be
connected via pull-up resistors to the bus voltage of
the system.
These lines can be used on 3.3V or 5.0V level with a
clock frequency of 100 kHz. For the detailed
specifications of this I²C communication, please refer
to specific I²C Application Notes from Sensirion.
Table 4: I²C Output Characteristics
Parameter
Min.
Typ.
Max.
Units
I
2
C Bus Clock Frequency
100
kHz
Output Voltage Low
(SDA/SCL), I
sink
= 6mA
0.1
0.5
V
Low Level Output Current
(SDA/SCL)
6
mA
High Level Input Voltage
(SDA/SCL)
2.0
V
Low Level Input Voltage
(SDA/SCL)
1.0
V
www.sensirion.com Version 6 December 2016 5/7
4 Fluidic Connection
Table 5: Fluidic Specifications
Parameter
SLG-0025
SLG-0075
SLG-0150
Wetted Materials:
Internal sensor tube material
Fused Silica
Fitting material
Stainless Steel
Sealing material
PEEK
Fluid connector ports (Fittings)
Stainless steel 10-32 coned port for Upchurch ferrules,
nanoViper™ compatible
Pressure drop (at full scale flow rate, H
2
O, 23°C)
2 bar
100 mbar
10 mbar
Total internal volume
100 nl
350 nl
1200 nl
For more information on the fluidic connection please find: “Application Note Sensor Ports and Tubing Connections” in the
Download Center on our webpage (www.sensirion.com/liquidflow-download).
5 Mechanical Specifications
Table 6: Mechanical Specifications
Parameter
SLG-0025
SLG-0075
SLG-0150
Largest dimensions
80 x 30 x 20 mm
Total mass
150 g
Inner diameter d
25 µm
75 µm
150 µm
Protection class
IP 53
All dimensions in mm
2x coned port 10-32
UNF, Upchurch and
nanoViper™
compatible
SLG-0025 / SLG-0075 / SLG-0150
M8 4-pin
www.sensirion.com Version 6 December 2016 6/7
Ordering Information
For OEM applications, the flow meter can be
purchased in larger quantities without any additional
parts. For optimum performance, Sensirion
recommends using the SLG flow meters in
combination with the SCC1 interface cables.
For laboratory use and technology evaluation, the
Flow Meter Kit SLG-xxxx can be ordered.
This laboratory-package contains:
Liquid Flow Meter SLG-0xxx
Fitting Material for UHP fittings
PC Software (Viewer & Data Export Tool)
SCC1-USB Sensor Cable with USB connector
for plug-and-play connection to a PC
SCC1-Analog Sensor Cable with 0-10 V voltage
output.
Table 7: Ordering information
Product
Article Number
SLG-0025 UHPLC Liquid Flow Meter
1-101316-01
SLG-0075 UHPLC Liquid Flow Meter
1-100592-02
SLG-0150 Liquid Flow Meter
1-101374-01
Flow Meter Kit SLG-0025
1-101334-01
Flow Meter Kit SLG-0075
1-100896-02
Flow Meter Kit SLG-0150
1-101389-01
Interface Cables:
SCC1-RS485 Sensor Cable Pigtail 2m
1-100804-01
SCC1-RS485 Sensor Cable Pigtail 5m
1-101122-01
SCC1-ANALOG Sensor Cable Pigtail 2m
1-101072-01
SCC1-ANALOG Sensor Cable Pigtail 10m
1-101219-01
SCC1-USB Sensor Cable 2m
1-101007-01

SLG-0150 FMK

Mfr. #:
Manufacturer:
Sensirion
Description:
Multiple Function Sensor Development Tools Ultra-High Pressure Liquid Flow Meter Kit
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet