951-800-526-2324 • www.us.tdk-lambda.com/lp/ Disclaimer: us.tdk-lambda.com/lp/legal.htm
CSS150 Series
Features
Wide Range AC Input
Low profi le, Industry Standard Footprint
Global Safety Agency Compliance
Dual input fuses
150W, 3 x 5” AC-DC Medical Power Supplies
(1) Derate linearly to 90W convection rating from 90 to 100VAC input.
Model CSS150
Input Voltage range V 90 - 264VAC (47 - 400Hz)(1) or 120 - 180VDC*
Inrush Current A <60A maximum at 264VAC input, 25°C ambient cold start
Input Current (115/230VAC) A 1.7 / 0.85
Power Factor Correction - Meets EN61000-3-2, >0.9
Leakage Current μA <250μA 264VAC 63Hz
Hold Up Time (Typ) ms 16ms at 115VAC input
Temperature Coeffi cient - ±0.05%/°C
Voltage Accuracy % ±1%
Minimum Load A None
Total Regulation % ±3%
Ripple & Noise % 2% peak to peak
Overcurrent Protection - 110 - 150%
Overvoltage Protection V 110 - 150% of nominal (Cycle input power to reset)
Remote On/Off - Unit on: Floating or low <1.2V, Unit off: Apply 5V with respect to 0V
Effi ciency % 89% typical
Off load power draw W <0.5W
Operating Temperature °C 0 to +70°C derate linearly to 50% load from 50 to 70°C
Storage Temperature °C -20 to +85°C
Humidity (non condensing) - 10 - 95% RH
Cooling - Convection or forced air (2.5m/s)
Withstand Voltage - Input to Ground 1.5kVAC, Input to Output 3kVAC (2 x MOOPs 3rd Edition),
Output to Ground 500VDC for 1 min.
Isolation Resistance - >100M at 25C & 70%RH, Output to Ground 500VDC
Vibration (non operating) - 19.6m/s
2
(10~55Hz:2G Constant, X,Y,Z 60min each.)
Shock - < 196.1 m/s
2
(20G)
Safety Agency Approvals - UL/CSA/IEC/EN 60950-1, UL/CSA/IEC/EN 60601-1, ANSI/AAMI ES60601-1, CE Mark
Conducted & Radiated EMI - EN55011-B, FCC Class B
Immunity - EN60601-1-2
MTBF - 180,000 (MIL-217F-HDBK)
Weight (Typ) g 340g
Size (WxLxH) mm 3 x 5 x 1.3” (including underside components)
Warranty yrs Two Years
Specifi cations
Key Market Segments & Applications
*Safety certifi ed for AC input only