E3X-DA
E3X-DA
7
SENSITIVITY SETTING (TEACHING)
Set the mode selector to “SET” to start teaching. The red level
display will flash if teaching error occurs. In this case, repeat the
entire teaching procedure. The E3X-DA has the following four
teaching modes.
Maximum Sensitivity Setting
1. Set the mode selector to “SET”.
2. Press the “TEACH” button for 3 seconds (min).
3. Teaching is complete when the level display changes from
red to green. Digital incident level will appear shortly after the
display changes from red to green.
(Red)
(Green)
1234
4. Set to “RUN” mode.
ONE-POINT TEACH MODE
(WITHOUT OBJECT)
1. Set the mode selector to “SET”.
2. Press the “TEACH” button for approximately 1 second.
3. Teaching is complete when the red level display is lit.
Digital incident level will appear shortly after the red level
display is lit.
(Red)
1234
4. Set to “RUN” mode.
5. The threshold is automatically set.
Note: Try the two-point teaching mode if the difference in level is
too fine.
TWO-POINT TEACH MODE
1. Set the mode selector to “SET”.
2. Press the “TEACH” button for approximately 1 second when
the object is at the sensing position.
Object
3. The red level display is lit.
(Red)
4. Remove the object from the sensing area, and press the
“TEACH” button for approximately 1 second.
5. Teaching is complete when the green level display is lit.
Digital incident level will appear shortly after the green dis-
play is lit.
(Green)
1234
6. Set to “RUN” mode.
E3X-DA
E3X-DA
8
ONE-POINT TEACH MODE
(POSITIONING)
1. First set to “One-point Teach Mode (without object)”.
2. Set the sensor to “ADJ” mode.
3. While in the “ADJ” mode, adjust the threshold by pressing the
“TEACH” button until the operation indicator is lit. Pressing
the “TEACH” button increases the sensitivity, but lowers the
threshold level. Pressing the “MODE” button decreases the
sensitivity, but increases the threshold level.
Object
Output
4. Set to “RUN” mode.
DISPLAY SETTING
E3X-DA can be programmed to display digital incident level,
digital percent level or analog display.
1. Set the mode selector to “RUN”. Digital incident level should
appear in the display.
Mode selector
Factory-set to RUN
2. Press the “MODE” button until “123P” is displayed. Digital
percent level is now displayed.
Digital percent
3. Press the “MODE” button to activate the “Analog Incident
Level” display. Two color (red and green) bar graph should
appear in the display.
Analog incident level and threshold
4. Press the “MODE” button to return to the “Digital Incident
Level” display.
SPECIAL FUNCTION SETTING
E3X-DA series has various functions that give users more
flexibility in solving a variety of application requirements such as
the following:
Long sensing distance (Sensing Function)
Faster response time (Sensing Function)
Off-delay timer (Timer Function)
Optical axis alignment (Flashing Function)
Digital data display hold setting (Display Hold Setting)
Data display orientation (Display Orientation Setting)
Sensing Function
1. Set the mode selector to “SET”.
Mode
selector
2. Press the “MODE” button until “fs t” is displayed. The “Stan-
dard” sensing function (fs t) is the default setting. The
“Standard” sensing function provides the standard detecting
distance (400 mm). Response time is 1 ms.
3. Press the “TEACH” button to obtain the “Long Distance”
sensing function. “F Ld” should appear on the display. The
“Long Distance” function provides a sensing distance that is
about 1.3 times that of the “Standard” detecting distance
(diffuse fiber in use). The “Long Distance” setting has a re-
sponse time of 4 ms.
4. Press the “TEACH” button to obtain the “High Speed” sens-
ing function. “F HS” Should appear on the display. The “High
Speed” function provides a detection distance that is about
one-third of the standard distance (diffuse fiber in use). Re-
sponse time is 0.25 ms.
5. Press the “TEACH” button to return to the “Standard” sensing
function display.
E3X-DA
E3X-DA
9
Timer Function
1. Set the mode selection to “SET”.
Mode
Selector
2. Press the “MODE” button until “t
––
0” is displayed. “t
––
0 is the
default setting.
0 (Factory-set)
3. Press the “TEACH” button to set to the desired time setting
(5 ms increment).
200
Adjustable in
5-ms units
Flashing Function
1. Set the mode selector to “SET”.
Mode
Selector
2. Press the “MODE” button until “LOFF” is displayed. “LOFF is
the default setting.
Off (Factory-set)
3. Press the “TEACH” button to activate the “flashing” function.
Display should read “L On”
4. Press the “TEACH” button to return to the “LOFF” mode.
Display Hold Setting Function
1. Set the mode to “SET”.
Mode
Selector
2. Press the “MODE” button until “HOFF” is displayed. “HOFF is
the default setting.
Off (Factory-set)
3. Press the “TEACH” button to activate “Peak Hold” display.
Displayed data get updated every 2 seconds. The maximum
value is displayed.
Peak
4. Press the “TEACH” button to activate “Bottom Hold” display.
Displayed data gets updated every 2 seconds. The minimum
value is displayed.
Bottom
5. Press the “TEACH” button to return to the default setting.
“HOFF”.
Off
Display Orientation Setting Function
1. Set the mode selector to “SET”.
Mode
Selector
2. Press the “MODE” button until “d123” is displayed. “d123 is
the standard setting.
Standard (Factory-set)
3. Press the “TEACH” button to change the orientation of the
display. “d321 should now appear in the display.
4. Press the “TEACH” button to return to the standard setting,
“d123”.