Si7015-A10
Rev. 1.0 13
4.2. Relative Humidity Sensor Accuracy
To determine the accuracy of a relative humidity sensor, it is placed in a temperature and humidity controlled
chamber. The temperature is set to a convenient fixed value (typically 30 °C) and the relative humidity is swept
from 20 to 80% and back to 20% in the following steps: 20% – 40% – 60% – 80% – 80% – 60% – 40% – 20%. At
each set-point, the chamber is allowed to settle for a period of 60 minutes before a reading is taken from the
sensor. Prior to the sweep, the device is allowed to stabilize to 50%RH. The solid top and bottom trace in Figure 6,
“Measuring Sensor Accuracy Including Hysteresis,” shows the result of a typical sweep after non-linearity
compensation.
Figure 6. Measuring Sensor Accuracy Including Hysteresis
The RH accuracy is defined as the center (dashed) line shown in Figure 6, which is the average of the two data
points at each relative humidity set-point. In this case, the sensor shows an accuracy of 0.25%RH. The Si7015
accuracy specification (Table 4) includes the following:
Unit-to-unit and lot-to-lot variation in non-linearity compensation
Accuracy of factory calibration
Margin for shifts that can occur during solder reflow.
The accuracy specification does not include the following:
Hysteresis (typically ±1%)
Effects from long term exposure to very humid conditions
Contamination of the sensor by particulates, chemicals, etc.
Other aging related shifts (“Long-term stability”)
Variations due to temperature
Ͳ5
Ͳ4
Ͳ3
Ͳ2
Ͳ1
0
1
2
3
4
5
10 20 30 40 50 60 70 80 90
%RHAccuracy
%RHSetͲpoint
RHAccuracyvs.RHSetͲPoint
Hysteresis
Si7015-A10
14 Rev. 1.0
4.3. Temperature Compensation
The Si7015 relative humidity sensor is calibrated at a temperature of 30 °C; it is at this temperature that the sensor
will give the most accurate relative humidity readings. For relative humidity measurements at other temperatures,
the RH reading from the Si7015 must be compensated for the change in temperature relative to 30 °C.
Temperature-compensated relative humidity readings can be calculated as follows:
Where:
RH
TempCompensated
is the temperature compensated relative humidity value in %RH.
RH
Linear
is the linear corrected relative humidity value in %RH.
Temperature is the ambient temperature in °C as measured by the Si7015 on chip temperature sensor.
Q
1
and Q
0
are unit-less correction coefficients derived through characterization of Si7015s by Silicon
Laboratories.
This temperature compensation is most accurate in the range of 15–50 °C. The values for the correction
coefficients are shown in Table 9.
4.4. Hysteresis
The moisture absorbent film (polymeric dielectric) of the humidity sensor will carry a memory of its exposure
history, particularly its recent or extreme exposure history. A sensor exposed to relatively low humidity will carry a
negative offset relative to the factory calibration, and a sensor exposed to relatively high humidity will carry a
positive offset relative to the factory calibration. This factor causes a hysteresis effect illustrated by the solid top
and bottom traces in Figure 6. The hysteresis value is the difference in %RH between the maximum absolute error
on the decreasing humidity ramp and the maximum absolute error on the increasing humidity ramp at a single
relative humidity Setpoint and is expressed as a bipolar quantity relative to the average, the center dashed trace in
Figure 6. In the case of Figure 6, the measurement uncertainty due to the hysteresis effect is ±1.05%RH.
4.5. Prolonged Exposure to High Humidity
Prolonged exposure to high humidity will result in a gradual upward drift of the RH reading. The shift in sensor
reading resulting from this drift will generally disappear slowly under normal ambient conditions. The amount of
shift is proportional to the magnitude of relative humidity and the length of exposure. In the case of lengthy
exposure to high humidity, some of the resulting shift may persist indefinitely under typical conditions. It is generally
possible to substantially reverse this affect by baking the device (see Section “4.8. Bake/Hydrate Procedure”).
Table 9. Linearization Coefficients
Coefficient Value
Q
0
0.060162
Q
1
0.000508
RH
TempCompensated
RH
Linear
Temperature 30RH
Linear
Q
1
Q
0
++=
Si7015-A10
Rev. 1.0 15
4.6. PCB Assembly
4.6.1. Soldering
Like most ICs, Si7015 devices are shipped from the factory vacuum-packed with an enclosed desiccant to avoid
any drift during storage and to prevent any moisture-related issues during solder reflow. The following guidelines
should be observed during PCB assembly:
Si7015 devices are compatible with standard board assembly processes. Devices should be soldered
using reflow per the recommended card reflow profile. See Section “10. PCB Land Pattern and Solder
Mask Design” for the recommended card reflow profile.
A “no clean” solder process is recommended to minimize the need for water or solvent rinses after
soldering. Cleaning after soldering is possible, but must be done carefully to avoid impacting the
performance of the sensor. See AN607 for more information on cleaning.
It is essential that the exposed polymer sensing film be kept clean and undamaged. This can be
accomplished by careful handling and a clean, well-controlled assembly process. When in doubt or for
extra protection, a heat-resistant, protective cover such as Kapton
®
KPPD-1/8 can be installed during PCB
assembly.
Si7015s may be ordered with a factory-fitted, solder-resistant protective cover. This cover provides protection
during PCB assembly or rework but without the time and effort required to install and remove the Kapton
®
tape. It
can be left in place for the lifetime of the product, preventing liquids, dust, or other contaminants from coming into
contact with the polymer sensor film. See Section “8. Ordering Guide” for a list of ordering part numbers that
include the cover.
4.6.2. Rehydration
The measured humidity value will generally shift slightly after solder reflow. A portion of this shift is permanent and
is accounted for in the accuracy specifications in Table 4. After soldering, an Si7015 should be allowed to
equilibrate under controlled RH conditions (room temperature, 45-55%RH) for at least 48 hours to eliminate the
remainder of the shift and return the device to its specified accuracy performance.

SI7015-A10-GMR

Mfr. #:
Manufacturer:
Silicon Labs
Description:
Board Mount Humidity Sensors Digital RH ( 4.5% max) & temperature sensor, commercial grade
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union