THM 10-0515

ht tp://www.tracopower.com Page 1 of 4
DC/DC Converter THM 10 Series, 10 Watt
The THM-10 series is a range of medical 10 Watt DC/DC converters in DIP-24
plastic package and with wide 2:1 input voltage range. They provide a reinforced
isolation system for 5000 VACrms isolation and a very low leakage current of
less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2
x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk
management file. Design and production conform to the quality management
system ISO 13485. With a high efficiency of up to 89% and highest grade com-
ponents the converters can reliably operate in an ambient temperature range
of –4C up to +90°C. They constitute a reliable solution not only for medical
equipment but also for demanding ranges of application such as transportation,
control & measurement or IGBT drivers.
Wide 2:1 input voltage 10 W DC/DC
converter in a compact DIP-24 plastic case
I/O isolation 5000 VACrms rated for
250 VACrms working voltage
Certification according to IEC/EN/ES
60601-1 3rd edition for 2MOPP
Risk management process according to
ISO 14971 including risk management file
Acceptance criteria for electronic
assemblies according to IPC-A-610 Level 3
Low leakage current < 2A
Extended operating temperature range
–40°C to 90°C.
EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
Operating up to 5000m altitude
5 year product warranty
Order code Input voltage range Output voltage Output current max. Efficiency typ.
THM 10-0510
3.3 VDC 2500 mA 80.0 %
THM 10-0511
5.0 VDC 2000 mA 84.0 %
THM 10-0512
12 VDC 830 mA 86.5 %
THM 10-0513
4.5 – 9 VDC
15 VDC 670 mA 87.0 %
THM 10-0515 (5 VDC nominal)
24 VDC 416 mA 85.5 %
THM 10-0521
±5.0 VDC ±1000 mA 83.0 %
THM 10-0522
±12 VDC ±416 mA 85.5 %
THM 10-0523
±15 VDC ±333 mA 86.5 %
THM 10-1210
3.3 VDC 2500 mA 83.0 %
THM 10-1211
5.0 VDC 2000 mA 85.5 %
THM 10-1212
12 VDC 830 mA 88.0 %
THM 10-1213
9. 0 – 18 VDC
15 VDC 670 mA 89.0 %
THM 10-1215 (12 VDC nominal)
24 VDC 416 mA 89.0 %
THM 10-1221
±5.0 VDC ±1000 mA 84.0 %
THM 10-1222
±12 VDC ±416 mA 89.0 %
THM 10-1223
±15 VDC ±333 mA 88.0 %
THM 10-2410
3.3 VDC 2500 mA 83.0 %
THM 10-2411
5.0 VDC 2000 mA 86.5 %
THM 10-2412
12 VDC 830 mA 89.0 %
THM 10-2413
18 – 36 VDC
15 VDC 670 mA 89.0 %
THM 10-2415 (24 VDC nominal)
24 VDC 416 mA 89.0 %
THM 10-2421
±5.0 VDC ±1000 mA 85.0 %
THM 10-2422
±12 VDC ±416 mA 89.0 %
THM 10-2423
±15 VDC ±333 mA 88.0 %
THM 10-4810
3.3 VDC 2500 mA 82.5 %
THM 10 - 4811
5.0 VDC 2000 mA 86.5 %
THM 10-4812
12 VDC 830 mA 89.0 %
THM 10-4813
36 – 75 VDC
15 VDC 670 mA 89.0 %
THM 10-4815 (48 VDC nominal)
24 VDC 416 mA 88.5 %
THM 10-4821
±5.0 VDC ±1000 mA 85.0 %
THM 10-4822
±12 VDC ±416 mA 88.0 %
THM 10-4823
±15 VDC ±333 mA 88.0 %
Models
ES 60601-1
C B
Scheme
IEC 606010-1
THM 10 Series, 10 Watt
http://www.tracopower.com Page 2 of 4
Input Specifications
Input current no load 5 Vin models: 20 mA typ.
12 Vin models: 10 mA typ.
24 Vin models: 6 mA typ.
48 Vin models: 4 mA typ.
Surge voltage (3 sec. max.) 5 Vin models: 16 V max.
12 Vin models: 25 V max.
24 Vin models: 50 V max.
48 Vin models: 100 V max.
Start-up voltage 5 Vin models: 4.5 VDC (or lower)
12 Vin models: 9 VDC (or lower)
24 Vin models: 18 VDC (or lower)
48 Vin models: 36 VDC (or lower)
Startup time 30 ms typ.
Under voltage shut down 5 Vin models: 3 4.4 VDC
12 Vin models: 7 8.8 VDC
24 Vin models: 15 – 17.5 VDC
48 Vin models: 31 – 34.5 VDC
Conducted noise – Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
EN 55032 class A (internal filter)
EN 55032 class B with external components
EMC immunity – Generic for Medical equipment IEC/EN 60601-1-2 4th edition
– ESD (electrostatic discharge) EN 61000-4-2, air ±8 kV, contact ±6 kV,
perf. criteria A
– Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A
– Fast transient / surge EN 61000-4-4, ±2 kV, perf. criteria A
(with external input capacitor / diode) EN 61000-4-5, ±2 kV perf. criteria A
5 Vin models: Nippon chemi-con KY 1000 µF/ 25 V and
reverse diode (Vishay V10P45) in parallel
12 & 24 Vin models: Nippon chemi-con KY 470 µF/ 50 V
48 Vin models: Nippon chemi-con KY 330 µF/ 100 V
Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A
– Magnetic field immunity EN 61000-4-8
100 A/m, continuous, perf. criteria A
1000 A/m, 1 sec., perf. criteria A
Output Specifications
Voltage set accuracy ±1 % max.
Regulation Input variation single output: 0.2 % max.
dual output: 0.5 % max.
– Load variation 0 – 100 % single output: 0.2 % max.
dual output: 1.0 % max.
– Cross regulation dual output: 5.0 % max. (asymmetrical load 25/100 %)
Minimum load not required
Ripple and noise (20 MHz Bandwidth) 3.3 & 5.0 VDC models: 30 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
12 & 15 VDC models: 40 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
24 VDC models: 50 mVp-p typ. with cap. 4.7 µF/50 V X7R MLCC
Transient response (25% load step change)
– Recovery time 250 s typ.
Over current limitation at 150 % typ. of Iout rated (hiccup mode)
Short circuit protection Continuous, automatic recovery
THM 10 Series, 10 Watt
http://www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Output Specifications (continued)
Over voltage protection –Single output 3.3 VDC models: 3.7 – 5.0 VDC
5.0 VDC models: 5.6 – 7.0 VDC
12 VDC models: 13.5 – 16.0 VDC
15 VDC models: 18.3 – 22.0 VDC
24 VDC models: 29.1 – 34.5 VDC
–Dual output ±5 VDC models: 5.6 – 7.0 VDC
±12 VDC models: 13.5 – 18.2 VDC
±15 VDC models: 17.0 – 22.0 VDC
Capacitive load –Single output 3.3 VDC models: 3‘000 F max.
5.0 VDC models: 2‘500 F max.
12 VDC models: 430 F max.
15 VDC models: 350 F max.
24 VDC models: 125 F max.
–Dual output ±5 VDC models: 1440 F max. (each output)
±12 VDC models: 250 F max. (each output)
±15 VDC models: 180 F max. (each output)
General Specifications
Temperature ranges – Operating –40°C to +90°C (with derating)
– Rated according to IEC/EN 60601-1 –40°C to +50°C (without derating)
– Case temperature +105°C max.
– Storage temperature –55°C to +125°C
Derating 3.5 %/K above 75°C
Thermal impedance 18 K/W typ.
Humidity (non condensing) 5 % to 95 % rel H max.
Isolation voltage – to meet ES/IEC/EN 60601-1 5000 VACrms, rated for 250 VACrms working
(50Hz, 60sec) voltage, 2  MOPP
Clearance/creepage 8 mm min.
Leakage current ( at 240 VAC, 60 Hz) 2 A max.
Isolation capacitance (input/output) 17 pF max.
Altitude during operation 5000 m
Temperature coefficient ±0.02 %/K typ.
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign) 3‘850‘000 h
Switching frequency 300 kHz ±30 kHz (pulse width modulation)
Vibration and thermal shock resistance according to MIL-STD-810F
Safety standards/approvals
– Medical equipment ANSI/AAMI ES60601-1:2005/(R)2012,
IEC/EN60601-1 3rd edition
– Certification documents www.tracopower.com/products/overview/thm10
Environmental compliance Reach www.tracopower.com/info/reach-declaration.pdf
RoHS RoHS directive 2011/65/EU
Physical Specifications
Casing material non-conductive black plastic
Base material non-conductive black plastic
Potting material silicone (UL94 V-0 rated)
Package weight 14 g (0.48oz)
Soldering temperature max. 265°C / 10 sec
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.

THM 10-0515

Mfr. #:
Manufacturer:
TRACO Power
Description:
Isolated DC/DC Converters 10W 4.5-9Vin 24Vout 416mA DIP Med Iso
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union