THM 30-4815

www.tracopower.com Page 1 of 4
DC/DC Converter THM 30 Series, 30 Watt
The THM-30 series is a range of medical 30 Watt DC/DC converters in 2.0”
x 1.0” plastic package and with wide 2:1 input voltage range. They provide a
reinforced isolation system for 5000 VACrms isolation and a very low leakage
current of less than 2.5 A. The units are approved to IEC/EN/ES 60601-1 3rd
edition for 2  MOPP (Means Of Patient Protection) and come along with an
ISO 14971 risk management file. Design and production conform to the quality
management system ISO 13485. With a high efficiency of up to 90% and highest
grade components the converters can reliably operate in an ambient tempera-
ture range of –40°C up to +80°C. They constitute a reliable solution not only for
medical equipment but also for demanding ranges of application such as trans-
portation, control & measurement or IGBT drivers.
Wide 2:1 input voltage 30 W DC/DC
converter in a 2  1 ” plastic case
I/O isolation 5000 VACrms rated for
250 VACrms working voltage
Certification according to IEC/EN/ES
60601-1 3rd edition for 2MOPP
Risk management process according to
ISO 14971 including risk management file
Acceptance criteria for electronic
assemblies according to IPC-A-610 Level 3
Low leakage current < 2.5 A
Extended operating temperature range
–40°C to 80°C.
EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
Operating up to 5000m altitude
5 year product warranty
Order code Input voltage range Output voltage Output current max. Efficiency typ.
THM 30-1211
5.0 VDC 6000 mA 88.5 %
THM 30-1212
12 VDC 2500 mA 88.5 %
THM 30-1213
9. 0 – 18 VDC
15 VDC 2000 mA 89.5 %
THM 30-1215 (12 VDC nominal)
24 VDC 1250 mA 89.0 %
THM 30-1221
±5 VDC ±3000 mA 86.0 %
THM 30-1222
±12 VDC ±1250 mA 88.5 %
THM 30-1223
±15 VDC ±1000 mA 89.0 %
THM 30-2411
5.0 VDC 6000 mA 88.5 %
THM 30-2412
12 VDC
2500 mA
89.0 %
THM 30-2413
18 – 36 VDC
15 VDC 2000 mA 90.5 %
THM 30-2415 (24 VDC nominal)
24 VDC 1250 mA 89.5 %
THM 30-2421
±5 VDC ±3000 mA 86.0 %
THM 30-2422
±12 VDC ±1250 mA 90.0 %
THM 30-2423
±15 VDC ±1000 mA 90.0 %
THM 30-4811
5.0 VDC 6000 mA 89.0 %
THM 30-4812
12 VDC 2500 mA 89.0 %
THM 30-4813
36 – 75 VDC
15 VDC 2000 mA 90.0 %
THM 30-4815 (48 VDC nominal)
24 VDC 1250 mA 89.0 %
THM 30-4821
±5 VDC ±3000 mA 86.5 %
THM 30-4822
±12 VDC ±1250 mA 90.0 %
THM 30-4823
±15 VDC ±1000 mA 89.5 %
Models
ES 60601-1
C B
Scheme
IEC 606010-1
THM 30 Series, 30 Watt
www.tracopower.com Page 2 of 4
Input Specifications
Input current no load 12 Vin models: 11 mA typ.
24 Vin models: 9 mA typ.
48 Vin models: 9 mA typ.
Surge voltage (3 sec. max.) 12 Vin models: 25 V max.
24 Vin models: 50 V max.
48 Vin models: 100 V max.
Start-up voltage 12 Vin models: 9 VDC (or lower)
24 Vin models: 18 VDC (or lower)
48 Vin models: 36 VDC (or lower)
Startup time 60 ms max. (30 ms typ.)
Under voltage shut down (lock-out circuit) 12 Vin models: 7.8 - 8.6 VDC
24 Vin models: 15.8 - 17.4 VDC
48 Vin models: 32 - 34 VDC
Input filter Pi-type
Conducted noise – Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
EN55032 class A (internal filter)
EN55032 class B with external components
– Filter proposal www.tracopower.com/overview/thm30
EMC immunity – Generic for Medical equipment IEC/EN 60601-1-2 4th edition
– ESD (electrostatic discharge) EN 61000-4-2, air ±15 kV, contact ±8 kV,
perf. criteria A
– Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A
– Fast transient / surge EN 61000-4-4, ±2 kV, perf. criteria A
(with external input capacitor / diode) EN 61000-4-5, ±2 kV perf. criteria A
12 Vin models: 2 pcs. Nippon chemi-con KY 220 µF / 100 V
1 pcs. TVS - SMDJ36A, 36V, 3000 W)
24 Vin models: 2 pcs. Nippon chemi-con KY 220 µF / 100 V
1 pcs. TVS - SMDJ58A, 58V, 3000 W)
48 Vin models: 2 pcs. Nippon chemi-con KY 220 µF / 100 V
1 pcs. TVS - SMDJ120A, 120V, 3000 W)
Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A
– Magnetic field immunity EN 61000-4-8
100 A/m, continuous, perf. criteria A
1000 A/m, 1 sec., perf. criteria A
External input fuse required 12 Vin models: 6.3 A
(recommended values, slow blow type) 24 Vin models: 3.15 A
48 Vin models: 1.6 A
Output Specifications
Voltage set accuracy ±1 % max.
Output voltage adjustment range 5 & 12 VDC models: ±10 %
(single output models only) 15 & 24 VDC models: 10 / +20 %
Regulation Input variation single output: 0.2 % max.
dual output: 0.5 % max.
– Load variation 0 – 100 % single output: 0.2 % max.
dual output: 1.0 % max.
– Cross regulation dual output: 5.0 % max. (asymmetrical load 25/100 %)
Temperature coefficient ±0.02 %/K typ.
Minimum load not required
Ripple and noise (±)5.0 VDC models: 50 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
(20 MHz Bandwidth) (±)12 VDC models: 75 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
±15 VDC models: 75 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
15 VDC models: 100 mVp-p typ. with cap. 10 µF/25 V X7R MLCC
24 VDC models: 100 mVp-p typ. with cap. 4.7 µF/50 V X7R MLCC
Transient response – Recovery time (25% load step change) 250 s typ.
THM 30 Series, 30 Watt
www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Output Specifications
Over current limitation at 150 % typ. of Iout rated (hiccup mode)
at 185 % max. of Iout rated (hiccup mode)
Short-circuit protection Continuous, automatic recovery
Overvoltage protection (±)5.0 VDC models: 6.2 VDC typ.
(±)12 VDC models: 15 VDC typ.
(±)15 VDC models: 20 VDC typ.
24 VDC models: 30 VDC typ.
Capacitive load –Single output 5.0 VDC models: 7‘200 F max.
12 VDC models: 1‘200 F max.
15 VDC models: 1‘000 F max.
24 VDC models: 375 F max.
–Dual output ±5 VDC models: 3‘600 F max. (each output)
±12 VDC models: 750 F max. (each output)
±15 VDC models: 500 F max. (each output)
General Specifications
Temperature ranges – Operating –40°C to +80°C
– Case temperature +105°C max.
– Storage temperature –55°C to +125°C
Derating (±)5 VDC models: 1.67 %/K above 45°C
other models: 2 %/K above 55°C
Overtemperature protection at 115°C typ.
Thermal impedance 12.9 K/W typ.
Humidity (non condensing) 5 % to 95 % rel H max.
Isolation voltage (50 Hz, 60 s) 5000 VACrms reinforced
Clearance/creepage 8 mm min.
Leakage current ( at 240 VAC, 60 Hz) 2.5 A max.
Isolation capacitance (input/output) 20 pF typ.
Altitude during operation 5000 m
Temperature coefficient ±0.02 %/K typ.
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign) 1‘137‘000 h
Switching frequency 225 – 285 kHz (pulse width modulation)
Vibration and thermal shock resistance according to MIL-STD-810F
Safety standards/approvals – Medical equipment ANSI/AAMI ES 60601-1:2005/(R)2012,
IEC/EN 60601-1 3rd edition
– Certification documents www.tracopower.com/overview/thm30
Environmental compliance Reach www.tracopower.com/products/reach-declaration.pdf
RoHS RoHS directive 2011/65/EU
Physical Specifications
Casing material non-conductive plastic
Base material non-conductive plastic
Potting material silicone (UL94 V-0 rated)
Package weight 32 g (1.13 oz)
Soldering temperature max. 265°C / 10 s

THM 30-4815

Mfr. #:
Manufacturer:
TRACO Power
Description:
Isolated DC/DC Converters 30W 36-75Vin 24Vout 1250mA Iso Medical
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union