SLI-1000 FMK

www.sensirion.com Version 11 July 2016 4/8
Figure 2: Flow meter accuracy and repeatability across the flow range. Relative error in % of measured value (left column) and
absolute error in µl/min or ml/min in case of the SLI-2000 (right column) for IPA
www.sensirion.com Version 11 July 2016 5/8
3 Communication with the Sensor
The SLI flow meter shows bidirectional, linear transfer
characteristics. The product comes fully calibrated for
water and IPA.
Digital Sampling Time, 16 bit 74 ms
Digital Sampling Time, 9 bit 1 ms
3.1 Electrical Specifications
Table 3: DC Characteristics
Parameter
Conditions
Min.
Typ.
Max.
Units
Power Supply
DC, VDD
Sensor only
4
5
6
V
RS485 cable
4
5
6
V
Analog cable
12
24
36
V
Operating
Current
VDD = 4-6 V,
no load
5
5.5
6
mA
RS485 cable
20
70
mA
Analog cable
4.3
mA
3.2 Electrical Connector and Pinout
The flow meter is equipped with a male connector
type M8, 4-pin, threaded lock according to
IEC 61076-2-101 (Ed. 1)/ IEC 60947-5-2, and is
compatible with Sensirion’s SCC1 interface cables.
Table 3: Electrical pinout
Pin
1
SDA (data)
2
GND
3
VDD
4
SCL (clock)
3.3 Communication via USB cable
The Sensirion USB Sensor Cable provides an easy to
use USB Interface for laboratory and desktop use.
For further information please see the SCC1-USB
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.4 Digital Communication via RS485-Bus
The SCC1-RS485 Sensor Cable for flow sensors
allows the communication via RS485 interface for use
in a demanding industrial automation environment. In
addition to the standard commands available in the
I²C interface of the sensor, the incorporated
microcontroller of the cable provides more complex
logic such as a dispense volume totalizer, automatic
dispense detection, automatic heater control and data
buffer for asynchronous read-out.
For further information please see the SCC1-RS485
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.5 Analog Communication
The SCC1-ANALOG Sensor Cable allows simple and
quick readout of Sensirion’s liquid flow meters by
converting the digital sensor reading to a 0...10.5 V
analog voltage output. Additionally, a digital (high/low)
output with two different modes of operation is
available (Flow Switch / Volume Counter)
For further information please see the SCC1-USB
Sensor Cable datasheet, available on
www.sensirion.com/liquidflow-download.
3.6 Digital Communication via I²C-Bus
Digital communication between a master and the SLI
sensor runs via the standard I²C-interface. The
physical interface consists of two bus lines, a data line
(SDA) and a clock line (SCL) which need to be
connected via pull-up resistors to the bus voltage of
the system.
These lines can be used on 3.3V or 5.0V level with a
clock frequency of 100 kHz. For the detailed
specifications of this I²C communication, please refer
to specific I²C Application Notes from Sensirion.
www.sensirion.com Version 11 July 2016 6/8
4 Fluidic Connection
Table 5: Fluidic Specifications and Pressure Rating
Parameter
SLI-0430
SLI-1000
SLI-2000
Wetted Materials:
Internal sensor tube material
Quartz Glass
(Fused Silica)
Borosilicate Glass 3.3
Fitting material
PEEK
Sealing material
None
FEP
Fluid connector ports (Fittings)
¼-28 flat-bottom for 1/16” or 1/8” OD plastic tubing
a
Pressure drop (at full scale flow rate, H
2
O, 23°C)
1 mbar
<1 mbar
<1 mbar
Pressure drop (at full scale flow rate, IPA, 23°C)
7 mbar
5 mbar
2 mbar
Total internal volume
5 µl
25 µl
80 µl
a
1/8” OD tubing with 2 mm minimum ID is recommended for the SLI-2000.
For more information on the fluidic connection please find: “Application Note Sensor Ports and Tubing Connections” in the
Download Center on our webpage (www.sensirion.com/liquidflow-download).
5 Mechanical Specifications
Table 6: Mechanical Specifications
Parameter
SLI-0430
SLI-1000
SLI-2000
Largest dimensions
58 x 42.5 x 20 mm
Total mass
53 g
Inner diameter d
430 µm
1.0 mm
1.8 mm
Protection class
IP 65
All dimensions in mm
SLI-0430
SLI-1000 / SLI-2000

SLI-1000 FMK

Mfr. #:
Manufacturer:
Sensirion
Description:
Multiple Function Sensor Development Tools Media Isolated Microfluidic Flow Meter Kit
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet