2
REV.G
AD7711
–15–
Table IV. External Series Resistance That Will Not Introduce
16-Bit Gain Error
External Capacitance (pF)
Gain 0 50 100 500 1000 5000
1 184 kW 45.3 kW 27.1 kW 7.3 kW 4.1 kW 1.1 kW
2 88.6 kW 22.1 kW 13.2 kW 3.6 kW 2.0 kW 560 W
4 41.4 kW 10.6 kW 6.3 kW 1.7 kW 970 W 270 W
8–128 17.6 kW 4.8 kW 2.9 kW 790 W 440 W
120 W
Table V. External Series Resistance That Will Not Introduce
20-Bit Gain Error
External Capacitance (pF)
Gain 0 50 100 500 1000 5000
1 145 kW 34.5 kW 20.4 kW 5.2 kW 2.8 kW 700 W
2 70.5 kW 16.9 kW 10 kW 2.5 kW 1.4 kW 350 W
4 31.8 kW 8.0 kW 4.8 kW 1.2 kW 670 W 170 W
8–128 13.4 kW 3.6 kW 2.2 kW 550 W 300 W 80 W
The numbers in Tables IV and V assume a full-scale change
on the analog input. In any case, the error introduced due to
longer charging times is a gain error that can be removed using
the system calibration capabilities of the AD7711, provided
the resultant span is within the limits of the system calibration
techniques.
ANALOG INPUT FUNCTIONS
Analog Input Ranges
Both analog inputs are programmable gain input channels that
can handle either unipolar or bipolar input signals. The AIN1
channel is a differential channel with a common-mode range
from V
SS
to AV
DD
, provided the absolute value of the analog
input voltage lies between V
SS
– 30 mV and AV
DD
+ 30 mV.
The AIN2 input channel is a single-ended input that is referred
to as AGND.
The dc input leakage current is 10 pA maximum at 25C (±1 nA
over temperature). This results in a dc offset voltage developed
across the source impedance. However, this dc offset effect can
be compensated for by a combination of the differential input
capability of the part and its system calibration mode.
Burnout Current
The AIN1(+) input of the AD7711 contains a 4.5 mA current
source that can be turned on/off via the control register. This
current source can be used in checking that a transducer has not
burned out or gone open circuit before attempting to take mea-
surements on that channel. If the current is turned on and
allowed to flow into the transducer and a measurement of the
input voltage on the AIN1 input is taken, it can indicate that the
transducer has burned out or gone open circuit. For normal
operation, this burnout current is turned off by writing a 0 to
the BO bit in the control register.
RTD Excitation Current
The AD7711 also contains two matched 200 mA constant cur-
rent sources that are provided at the RTD1 and RTD2 pins of
the device. These currents can be turned on/off via the control
register. Writing a 1 to the RO bit of the control register enables
these excitation currents.
For 4-wire RTD applications, one of these excitation currents is
used to provide the excitation current for the RTD; the second
current source can be left unconnected. For 3-wire RTD con-
figurations, the second on-chip current source can be used to
eliminate errors due to voltage drops across lead resistances.
Figures 19 to 21 in the Applications section show some RTD
configurations with the AD7711.
The temperature coefficient of the RTD current sources is
typically 20 ppm/C with a typical matching between the tem-
perature coefficients of both current sources of 3 ppm/C. For
applications where the absolute value of the temperature coeffi-
cient is too large, the following schemes can be used to remove
the drift error.
The conversion result from the AD7711 is ratiometric to the
V
REF
voltage. Therefore, if the V
REF
voltage varies with the RTD
temperature coefficient, the temperature drift from the current
source will be removed. For 4-wire RTD applications, the refer-
ence voltage can be made ratiometric to RTD current source
by using the second current with a low TC resistor to generate
the reference voltage for the part. In this case, if a 12.5 kW
resistor is used, the 200 mA current source generates 2.5 V across
the resistor. This 2.5 V can be applied to the REF IN(+) input
of the AD7711 and with the REF IN(–) input at ground, it will
supply a V
REF
of 2.5 V for the part. For 3-wire RTD configura-
tions, the reference voltage for the part is generated by placing a
low TC resistor (12.5 kW for 2.5 V reference) in series with one
of the constant current sources. The RTD current sources can
be driven to within 2 V of AV
DD
. The reference input of the
AD7711 is differential so the REF IN(+) and REF IN(–) of the
AD7711 are driven from either side of the resistor. Both schemes
ensure that the reference voltage for the part tracks the RTD
current sources over temperature and, thereby, remove the
temperature drift error.
Bipolar/Unipolar Inputs
The two analog inputs on the AD7711 can accept either unipo-
lar or bipolar input voltage ranges. Bipolar or unipolar options
are chosen by programming the B/U bit of the control register.
This programs both channels for either unipolar or bipolar
operation. Programming the part for either type of operation
does not change any of the input signal conditioning; it simply
changes the data output coding, using binary for unipolar inputs
and offset binary for bipolar inputs.
The AIN1 input channel is differential and, as a result, the
voltage to which the unipolar and bipolar signals are referenced
is the voltage on the AIN1(–) input. For example, if AIN1(–) is
1.25 V and the AD7711 is configured for unipolar operation
with a gain of 1 and a V
REF
of 2.5 V, the input voltage range on
the AIN1(+) input is 1.25 V to 3.75 V. If AIN1(–) is 1.25 V,
and the AD7711 is configured for bipolar mode with a gain of
1 and a V
REF
of 2.5 V, the analog input range on the AIN1(+)
input is –1.25 V to +3.75 V. For the AIN2 input, the input
signals are referenced to AGND.
REFERENCE INPUT/OUTPUT
The AD7711 contains a temperature compensated 2.5 V reference
that has an initial tolerance of ±1%. This reference voltage is
provided at the REF OUT pin, and it can be used as the reference
voltage for the part by connecting the REF OUT pin to the REF
REV. G
–16–
AD7711
IN(+) pin. This REF OUT pin is a single-ended output, refer-
enced to AGND, which is capable of providing up to 1 mA to
an external load. In applications where REF OUT is connected
directly to REF IN(+), REF IN(–) should be tied to AGND to
provide the nominal 2.5 V reference for the AD7711.
The reference inputs of the AD7711, REF IN(+) and REF IN(–),
provide a differential reference input capability. The common-
mode range for these differential inputs is from V
SS
to AV
DD
.
The nominal differential voltage, V
REF
(REF IN(+) – REF IN(–)),
is 2.5 V for specified operation, but the reference voltage can go
to 5 V with no degradation in performance if the absolute value
of REF IN(+) and REF IN(–) does not exceed its AV
DD
and
V
SS
limits and the V
BIAS
input voltage range limits are obeyed.
The part is also functional with V
REF
voltages down to 1 V but with
degraded performance because the output noise will, in terms
of LSB size, be larger. REF IN(+) must always be greater than
REF IN(–) for correct operation of the AD7711.
Both reference inputs provide a high impedance, dynamic load
similar to the analog inputs. The maximum dc input leakage
current is 10 pA (± 1 nA over temperature), and source resis-
tance may result in gain errors on the part. The reference inputs
look like the analog input (see Figure 7). In this case, R
INT
is
5 kW typ and C
INT
varies with gain. The input sample rate is
f
CLK IN
/256 and does not vary with gain. For gains of 1 to 8, C
INT
is 20 pF; for a gain of 16, it is 10 pF; for a gain of 32, it is 5 pF; for
a gain of 64, it is 2.5 pF; and for a gain of 128, it is 1.25 pF.
The digital filter of the AD7711 removes noise from the refer-
ence input just as it does with the analog input, and the same
limitations apply regarding lack of noise rejection at integer
multiples of the sampling frequency. The output noise perfor-
mance outlined in Tables I and II assumes a clean reference. If
the reference noise in the bandwidth of interest is excessive, it
can degrade the performance of the AD7711. Using the on-chip
reference as the reference source for the part (connecting
REF OUT to REF IN) results in degraded output noise perfor-
mance from the AD7711 for portions of the noise table that are
dominated by the device noise. The on-chip reference noise
effect is eliminated in ratiometric applications where the refer-
ence is used to provide the excitation voltage for the analog
front end. The connection shown in Figure 8 is recommended
when using the on-chip reference. Recommended reference
voltage sources for the AD7711 include the AD580 and AD680
2.5 V references.
AD7711
REF OUT REF IN(+)
REF IN(–)
Figure 8. REF OUT/REF IN Connection
V
BIAS
Input
The V
BIAS
input determines at what voltage the internal analog
circuitry is biased. It essentially provides the return path for
analog currents flowing in the modulator and, as such, it should
be driven from a low impedance point to minimize errors.
For maximum internal headroom, the V
BIAS
voltage should be
set halfway between AV
DD
and V
SS
. The difference between
AV
DD
and (V
BIAS
+ 0.85 ¥ V
REF
) determines the amount of
headroom the circuit has at the upper end, while the difference
between V
SS
and (V
BIAS
– 0.85 ¥ V
REF
) determines the amount
of headroom the circuit has at the lower end. When choosing a
V
BIAS
voltage, ensure that it stays within prescribed limits. For
single 5 V operation, the selected V
BIAS
voltage must ensure that
V
BIAS
± 0.85 ¥ V
REF
does not exceed AV
DD
or V
SS
or that the
V
BIAS
voltage itself is greater than V
SS
+ 2.1 V and less than
AV
DD
– 2.1 V. For single 10 V operation or dual ± 5 V opera-
tion, the selected V
BIAS
voltage must ensure that V
BIAS
¥ 0.85 ¥
V
REF
does not exceed AV
DD
or V
SS
or that the V
BIAS
voltage
itself is greater than V
SS
+ 3 V or less than AV
DD
– 3 V. For
example, with AV
DD
= 4.75 V, V
SS
= 0 V, and V
REF
= 2.5 V, the
allowable range for the V
BIAS
voltage is 2.125 V to 2.625 V. With
AV
DD
= 9.5 V, V
SS
= 0 V, and V
REF
= 5 V, the range for V
BIAS
is 4.25 V to 5.25 V. With AV
DD
= +4.75 V, V
SS
= –4.75 V,
and V
REF
= +2.5 V, the V
BIAS
range is –2.625 V to +2.625 V.
The V
BIAS
voltage does have an effect on the AV
DD
power sup-
ply rejection performance of the AD7711. If the V
BIAS
voltage
tracks the AV
DD
supply, it improves the power supply rejection
from the AV
DD
supply line from 80 dB to 95 dB. Using an
external Zener diode connected between the AV
DD
line and
V
BIAS
as the source for the V
BIAS
voltage gives the improvement
in AV
DD
power supply rejection performance.
USING THE AD7711
SYSTEM DESIGN CONSIDERATIONS
The AD7711 operates differently from successive approxima-
tion ADCs or integrating ADCs. Because it samples the signal
continuously, like a tracking ADC, there is no need for a start
convert command. The output register is updated at a rate
determined by the first notch of the filter, and the output can
be read at any time, either synchronously or asynchronously.
Clocking
The AD7711 requires a master clock input, which may be an
external TTL/CMOS compatible clock signal applied to the
MCLK IN pin with the MCLK OUT pin left unconnected.
Alternatively, a crystal of the correct frequency can be connected
between MCLK IN and MCLK OUT, in which case the clock
circuit will function as a crystal-controlled oscillator. For lower
clock frequencies, a ceramic resonator may be used instead of
the crystal. For these lower frequency oscillators, external
capacitors may be required on either the ceramic resonator or
on the crystal.
The input sampling frequency, the modulator sampling frequency,
the –3 dB frequency, the output update rate, and the calibration
time are all directly related to the master clock frequency,
f
CLK IN.
Reducing the master clock frequency by a factor of 2 will
halve the above frequencies and update rate and will double the
calibration time.
The current drawn from the DV
DD
power supply is also directly
related to f
CLK IN
. Reducing f
CLK IN
by a factor of 2 will halve the
DV
DD
current but will not affect the current drawn from the
AV
DD
power supply.
System Synchronization
If multiple AD7711s are operated from a common master clock,
they can be synchronized to update their output registers simul-
taneously. A falling edge on the SYNC input resets the filter and
2
REV.G
AD7711
–17–
places the AD7711 into a consistent, known state. A common
signal to the AD7711s’ SYNC inputs will synchronize their
operation. This would typically be done after each AD7711 has
performed its own calibration or has had calibration coefficients
loaded to it.
The SYNC input can also be used to reset the digital filter in
systems where the turn-on time of the digital power supply
(DV
DD
) is very long. In such cases, the AD7711 starts operat-
ing internally before the DV
DD
line has reached its minimum
operating level, 4.75 V. With a low DV
DD
voltage, the
AD7711’s internal digital filter logic does not operate correctly.
Thus, the AD7711 may have clocked itself into an incorrect
operating condition by the time DV
DD
has reached its correct
level. The digital filter is reset upon issue of a calibration com-
mand (whether it is self-calibration, system calibration, or back-
ground calibration) to the AD7711. This ensures correct
operation of the AD7711. In systems where the power-on de-
fault conditions of the AD7711 are acceptable, and no calibra-
tion is performed after power-on, issuing a SYNC pulse to the
AD7711 resets the AD7711’s digital filter logic. An R, C on the
SYNC line, with R, C time constant longer than the DV
DD
power-on time, performs the SYNC function.
Accuracy
Sigma-delta ADCs, like VFCs and other integrating ADCs, do
not contain any source of nonmonotonicity and inherently offer
no missing codes performance. The AD7711 achieves excellent
linearity by the use of high quality, on-chip silicon dioxide capaci-
tors, which have a very low capacitance/voltage coefficient. The
device also achieves low input drift through the use of chopper
stabilized techniques in its input stage. To ensure excellent
performance over time and temperature, the AD7711 uses digital
calibration techniques that minimize offset and gain error.
Autocalibration
Autocalibration on the AD7711 removes offset and gain errors
from the device. A calibration routine should be initiated on the
device whenever there is a change in the ambient operating
temperature or supply voltage. It should also be initiated if there
is a change in the selected gain, filter notch, or bipolar/unipolar
input range. However, if the AD7711 is in background calibra-
tion mode, these changes are taken care of automatically (after
the settling time of the filter has been allowed for).
The AD7711 offers self-calibration, system calibration, and
background calibration facilities. For calibration to occur on
the selected channel, the on-chip microcontroller must record
the modulator output for two different input conditions. These
are zero-scale and full-scale points. With these readings, the
microcontroller can calculate the gain slope for the input-to-
output transfer function of the converter. Internally, the part
works with a resolution of 33 bits to determine its conversion
result of either 16 bits or 24 bits.
The AD7711 also provides the facility to write to the on-chip
calibration registers, and in this manner, the span and offset for
the part can be adjusted by the user. The offset calibration regis-
ter contains a value that is subtracted from all conversion
results, while the full-scale calibration register contains a value
that is multiplied by all conversion results. The offset calibration
coefficient is subtracted from the result prior to the multiplica-
tion by the full-scale coefficient. In the first three modes outlined
here, the DRDY line indicates that calibration is complete by
going low. If DRDY is low before (or goes low during) the cali-
bration command, it may take up to one modulator cycle before
DRDY goes high to indicate that calibration is in progress.
Therefore, DRDY should be ignored for up to one modulator
cycle after the last bit of the calibration command is written to
the control register.
Self-Calibration
In the self-calibration mode with a unipolar input range, the
zero-scale point used in determining the calibration coefficients
is with both inputs shorted (that is, AIN1(+) = AIN1(–) =
V
BIAS
for AIN1 and AIN2 = V
BIAS
for AIN2), and the full-scale
point is V
REF
. The zero-scale coefficient is determined by con-
verting an internal shorted input node. The full-scale coefficient
is determined from the span between this shorted input conver-
sion and a conversion on an internal V
REF
node. The self-
calibration mode is invoked by writing the appropriate values
(0, 0, 1) to the MD2, MD1, and MD0 bits of the control regis-
ter. In this calibration mode, the shorted input node is switched
into the modulator first and a conversion is performed; the V
REF
node is then switched in and another conversion is performed.
When the calibration sequence is complete, the calibration coeffi-
cients updated, and the filter resettled to the analog input voltage,
the DRDY output goes low. The self-calibration procedure takes
into account the selected gain on the PGA.
For bipolar input ranges in the self-calibrating mode, the sequence
is very similar to that just outlined. In this case, the two points
that the AD7711 calibrates are midscale (bipolar zero) and
positive full scale.
System Calibration
System calibration allows the AD7711 to compensate for system
gain and offset errors as well as its own internal errors. System
calibration performs the same slope factor calculations as self-
calibration but uses voltage values presented by the system to
the AIN inputs for the zero- and full-scale points. System
calibration is a two-step process. The zero-scale point must
be presented to the converter first. It must be applied to the
converter before the calibration step is initiated and must
remain stable until the step is complete. System calibration is
initiated by writing the appropriate values (0, 1, 0) to the MD2,
MD1, and MD0 bits of the control register. The DRDY output
from the device signals when the step is complete by going low.
After the zero-scale point is calibrated, the full-scale point is
applied and the second step of the calibration process is initiated
by again writing the appropriate values (0, 1, 1) to MD2, MD1,
and MD0. Again the full-scale voltage must be set up before the
calibration is initiated, and it must remain stable throughout the
calibration step. DRDY goes low at the end of this second step to
indicate that the system calibration is complete. In the unipolar
mode, the system calibration is performed between the two end-
points of the transfer function; in the bipolar mode, it is performed
between midscale and positive full scale.
This two-step system calibration mode offers another feature.
After the sequence has been completed, additional offset or gain
calibrations can be performed by themselves to adjust the zero
reference point or the system gain. This is achieved by perform-
ing the first step of the system calibration sequence (by writing
0, 1, 0 to MD2, MD1, MD0). This adjusts the zero-scale or
offset point but does not change the slope factor from that set
during a full system calibration sequence.

AD7711AQ

Mfr. #:
Manufacturer:
Description:
Analog to Digital Converters - ADC CMOS 24B w/ Matched RTD Excitation Crnt
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union