Si7013
16 Rev. 0.95
4.1. Relative Humidity Sensor Accuracy
To determine the accuracy of a relative humidity sensor, it is placed in a temperature and humidity controlled
chamber. The temperature is set to a convenient fixed value (typically 25–30 °C) and the relative humidity is swept
from 20 to 80% and back to 20% in the following steps: 20% – 40% – 60% – 80% – 80% – 60% – 40% – 20%. At
each set-point, the chamber is allowed to settle for a period of 60 minutes before a reading is taken from the
sensor. Prior to the sweep, the device is allowed to stabilize to 50%RH. The solid trace in Figure 9 shows the result
of a typical sweep.
Figure 9. Measuring Sensor Accuracy Including Hysteresis
The RH accuracy is defined as the dotted line shown in Figure 9, which is the average of the two data points at
each relative humidity set-point. In this case, the sensor shows an accuracy of 0.25%RH. The Si7013 accuracy
specification (Table 4) includes:
Unit-to-unit and lot-to-lot variation
Accuracy of factory calibration
Margin for shifts that can occur during solder reflow
The accuracy specification does not include:
Hysteresis (typically ±1%)
Effects from long term exposure to very humid conditions
Contamination of the sensor by particulates, chemicals, etc.
Other aging related shifts ("Long-term stability")
Variations due to temperature
Si7013
Rev. 0.95 17
4.2. Hysteresis
The moisture absorbent film (polymeric dielectric) of the humidity sensor will carry a memory of its exposure
history, particularly its recent or extreme exposure history. A sensor exposed to relatively low humidity will carry a
negative offset relative to the factory calibration, and a sensor exposed to relatively high humidity will carry a
positive offset relative to the factory calibration. This factor causes a hysteresis effect illustrated by the solid trace
in Figure 9. The hysteresis value is the difference in %RH between the maximum absolute error on the decreasing
humidity ramp and the maximum absolute error on the increasing humidity ramp at a single relative humidity
setpoint and is expressed as a bipolar quantity relative to the average error (dashed trace). In the example of
Figure 9, the measurement uncertainty due to the hysteresis effect is +/-1.0%RH.
4.3. Prolonged Exposure to High Humidity
Prolonged exposure to high humidity will result in a gradual upward drift of the RH reading. The shift in sensor
reading resulting from this drift will generally disappear slowly under normal ambient conditions. The amount of
shift is proportional to the magnitude of relative humidity and the length of exposure. In the case of lengthy
exposure to high humidity, some of the resulting shift may persist indefinitely under typical conditions. It is generally
possible to substantially reverse this affect by baking the device (see Section “4.6. Bake/Hydrate Procedure” ).
4.4. PCB Assembly
4.4.1. Soldering
Like most ICs, Si7013 devices are shipped from the factory vacuum-packed with an enclosed desiccant to avoid
any drift during storage and to prevent any moisture-related issues during solder reflow. The following guidelines
should be observed during PCB assembly:
Si7013 devices are compatible with standard board assembly processes. Devices should be soldered
using reflow per the recommended card reflow profile. See Section “10. PCB Land Pattern and Solder
Mask Design” for the recommended card reflow profile.
A “no clean” solder process is recommended to minimize the need for water or solvent rinses after
soldering. Cleaning after soldering is possible, but must be done carefully to avoid impacting the
performance of the sensor. See application note “AN607: Si70xx Humidity Sensor Designer’s Guide” for
more information on cleaning.
It is essential that the exposed polymer sensing film be kept clean and undamaged. This can be
accomplished by careful handling and a clean, well-controlled assembly process. When in doubt or for
extra protection, a heat-resistant, protective cover such as Kapton

KPPD-1/8 can be installed during PCB
assembly.
Si7013s may be ordered with a factory-fitted, solder-resistant protective cover. This cover provides protection
during PCB assembly or rework but without the time and effort required to install and remove the Kapton tape. It
can be left in place for the lifetime of the product, preventing liquids, dust or other contaminants from coming into
contact with the polymer sensor film. See Section “8. Ordering Guide” for a list of ordering part numbers that
include the cover.
4.4.2. Rehydration
The measured humidity value will generally shift slightly after solder reflow. A portion of this shift is permanent and
is accounted for in the accuracy specifications in Table 4. After soldering, an Si7013 should be allowed to
equilibrate under controlled RH conditions (room temperature, 45–55%RH) for at least 48 hours to eliminate the
remainder of the shift and return the device to its specified accuracy performance.
Si7013
18 Rev. 0.95
4.4.3. Rework
To maintain the specified sensor performance, care must be taken during rework to minimize the exposure of the
device to excessive heat and to avoid damage/contamination or a shift in the sensor reading due to liquids, solder
flux, etc. Manual touch-up using a soldering iron is permissible under the following guidelines:
The exposed polymer sensing film must be kept clean and undamaged. A protective cover is
recommended during any rework operation (Kapton® tape or the factory installed cover).
Flux must not be allowed to contaminate the sensor; liquid flux is not recommended even with a cover in
place. Conventional lead-free solder with rosin core is acceptable for touch-up as long as a cover is in
place during the rework.
If possible, avoid water or solvent rinses after touch-up. Cleaning after soldering is possible, but must be
done carefully to avoid impacting the performance of the sensor. See AN607 for more information on
cleaning.
Minimize the heating of the device. Soldering iron temperatures should not exceed 350 °C and the contact
time per pin should not exceed 5 seconds.
Hot air rework is not recommended. If a device must be replaced, remove the device by hot air and solder
a new part in its place by reflow following the guidelines above.
*Note: All trademarks are the property of their respective owners.
Figure 10. Si70xx with Factory-Installed Protective Cover

SI7013-A10-GM

Mfr. #:
Manufacturer:
Silicon Labs
Description:
Board Mount Humidity Sensors Digital RH (+/-2%) & 2-Zone temp senso
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union