TBFLPNS001BGUCV

Sensing and Productivity Solutions
Datasheet
Basic Force Sensors
TBF Series, Compensated/Unamplified
1 bar to 10 bar | 100 kPa to 1 MPa | 15 psi to 150 psi
Millivolt Analog Output
DESCRIPTION
Honeywell’s TBF Series Force Sensors are small flush
diaphragm pressure sensors designed for customers
who require a simple device for applications where media
compatibility and low trapped volume are important. The TBF
Series has a precisely controlled diaphragm height, making
these products useful in applications where force is applied
by a flexible membrane to the sensor, such as infusion pumps.
The sensor is internally unamplified, providing infinite resolution
and allowing customers to do their own amplification in order to
make use of the maximum resolution of the bare sensor output,
leveraging any algorithm needed for the application. The sensor
is internally temperature compensated and calibrated.
VALUE TO CUSTOMERS
Smallest package size in its class (7 mm x 7 mm x 3,89 mm)
simplifies placement on crowded PCBs
Widest pressure (1 bar to 10 bar |100 kPa to 1 MPa | 15 psi
to 150 psi) enables choice of optimum pressure range to
maximize sensitivity
Many different pressure ranges improve resolution and system
accuracy
Tight accuracy specification of ±0.15 %FSS provides user
with enhanced accuracy in the application
Reliable supply chain
Fast response for quotes and samples
• Designed to Six Sigma standards
• Manufacturing excellence
Supports Lean manufacturing
FEATURES
Cost-effective sensor with many options
• Compensated/unamplified
Wide operating temperature range (0 °C to 50 °C
[32 °F to 122 °F])
Low power consumption allows for potential use in battery
operated applications
Stable offset voltage
Not sensitive to mounting orientation
Small package size [7 mm x 7 mm x 3,89 mm]
RoHS2 compliance
POTENTIAL APPLICATIONS
Medical
Drug delivery systems
Infusion pumps
Kidney dialysis machines
• Robotics
Syringe pumps
• Wearables
PORTFOLIO
The TBF Series joins the 1865 Series, FSG Series, FSS Series,
FSS-SMT Series and FSA Series.
32315675
Issue A
2
Sensing and Productivity Solutions
Basic Force Sensors, TBF Series, Compensated/Unamplified
Table 1. Absolute Maximum Ratings
1
Characteristic Min. Max. Unit
Supply voltage (Vsupply)
2
-12.0 12.0 Vdc
Storage temperature -40 [-40] 125 [257] °C [°F]
Soldering time peak reflow temperature 10 s max. at 240 °C [464 °F]
1
Absolute maximum ratings are the extreme limits the device will withstand without damage.
2
Incorrect application of supply voltage or ground to the wrong pin may cause electrical failure.
Table 2. Operating Specifications
Characteristic Min. Typ. Max. Unit
Supply voltage (Vsupply)
1, 2
1.5 5.0 12.0 Vdc
Supply current (at 5.0 Vdc supply) - 0.6 1 mA
Operating temperature range
3
0 [32] 50 [122] °C [°F]
Compensated temperature range
4
0 [32] 50 [122] °C [°F]
Output resistance 2.5 - kOhm
1
Ratiometricity of the sensor (the ability of the device output to scale to the supply voltage) is achieved within the specified operating
voltage.
2
Incorrect application of supply voltage or ground to the wrong pin may cause electrical failure.
3
Operating temperature range: The temperature range over which the sensor will produce an output proportional to force.
4
Compensated temperature range: The temperature range over which the sensor will produce an output proportional to force within the
specified performance limits.
Table 3. Environmental Specifications
Characteristic Parameter
Humidity
0 %RH to 95 %RH, non-condensing
Vibration
15 g, 10 Hz to 2 kHz
Shock
100 g, 6 ms duration
Life
1
1 million pressure cycles min.
Solder reflow
J-STD-020-D, MSL 1 (unlimited shelf life when stored at less than 30 °C and 85 %RH)
1
Life may vary depending on specific application in which sensor is utilized.
CAUTION
PRODUCT SENSING SURFACE DAMAGE
• The sensing surface of the sensor is composed of a tough silicone gel. Ensure that the sensing surface is not used with media
incompatible with silicones.
• Ensure that the sensing surface does not come into contact with sharp or hard objects.
Failure to comply with these instructions may result in product damage.
NOTICE
In order for the TBF Series sensors to provide a linear and repeatable output, ensure the entire top surface of the gel is exposed
to a uniform pressure. The silicone gel allows direct contact with many liquids or the gel may be protected with a thin, compliant
membrane.
3
Sensing and Productivity Solutions
Basic Force Sensors, TBF Series, Compensated/Unamplified
Table 6. Pressure Range Specifications for 1 bar to 10 bar
Pressure Range
Order Code
(see Figure 1)
Pressure
Range
Unit
Overpressure
Pressure
Accuracy
1
(%FSS)
Offset
2
(mV/V)
Full Scale Span
3
(mV/V)
Thermal Effect
on Offset
4
(%FSS)
Thermal Effect
on Span
5
(%FSS)
Pmin. Pmax. Min. Nom. Max.
0 ºC
to
50 ºC
0 ºC
to
50 ºC
Gage
001BG 0 1 bar 4 ±0.5 ±0.3 4.90 5.10 5.30 ±1.0 ±1.0
1.6BG 0 1.6 bar 4 ±0.5 ±0.3 7.8 4 8.15 8.48 ±1.0 ±1.0
2.5BG 0 2.5 bar 8 ±0.5 ±0.15 6.10 6.35 6.59 ±1.0 ±0.75
004BG 0 4 bar 10 ±0.5 ±0.075 5.57 5.80 6.04 ±1.0 ±0.75
006BG 0 6 bar 17 ±0.5 ±0.075 5.08 5.30 5.54 ±0.75 ±0.75
010BG 0 10 bar 17 ±0.5 ±0.075 8.47 8.85 9.22 ±0.50 ±0.75
1
Accuracy: The maximum deviation in output from a Best Fit Straight Line (BFSL) fitted to the output measured over the pressure
range at 25 °C [77 °F]. Includes all errors due to pressure non-linearity, pressure hysteresis, and non-repeatability.
2
Offset: The output signal obtained when the reference pressure is applied to all available pressure ports. Also known as “null” or
zero”.
3
Full Scale Span: The algebraic difference between the output signal measured at the maximum and minimum limits of the
pressure range (see Figure 1).
4
Thermal effect on offset: The deviation in offset due to changes in temperature over the compensated temperature range, relative
to offset measured at 25 ºC.
5
Thermal effect on span: The deviation in full scale span due to changes in temperature over the compensated temperature range,
relative to full scale span measured at 25 ºC.
Table 5. Material Composition
Component Description
Cover high temperature polyamide
Substrate not exposed - protected by silicone gel
Sensing surface silicone gel
Table 4. Sensor Pressure Type
Pressure Type Description
Gage
Output is proportional to the difference between applied pressure and atmospheric (ambient) pressure.
Reference pressure is atmospheric pressure.

TBFLPNS001BGUCV