AXL E IOL RTD1 M12 R
8590_en_06 PHOENIX CONTACT 4
IO-Link port supply
Typical current consumption 16.5 mA ±15% (at 24 V DC)
Current consumption max. 135 mA
Protection against polarity reversal Yes
Short-circuit protection Yes
Overload protection Yes
I/O supply voltage 24 V DC
This supply voltage is provided via the IO-Link interface of the IO-Link master.
Analog RTD input
Number of inputs 1 (for resistance temperature detectors)
Connection method M12 connector, A-coded
Connection method 3-wire (optionally 4-wire)
Sensor types (RTD) that can be used Pt100, Pt1000
Linear resistance measuring range IB IL format: 0 Ω ... 500 Ω, 0 Ω ... 5 kΩ
S7-compatible format: 0 Ω ... 600 Ω, 0 Ω ... 6 kΩ
Resolution A/D 12 bit
Data formats IB IL, S7-compatible
Relative accuracy < 0.3 % (from the measuring range final value in the nominal range)
Permissible cable length max. 30 m (shielded cable)
Protection against polarity reversal yes
Short-circuit protection yes
Overload protection yes
Error messages to the higher level control or computer system
Overrange or underrange Message in the diagnostic code (in the IBIL format)
Measured value invalid/no valid measured value present Message in the diagnostic code (in the IBIL format)
Configuration invalid Message in the diagnostic code (in the IBIL format)
Module faulty Message in the diagnostic code (in the IBIL format)
Mechanical tests
Vibration resistance in acc. with EN 60068-2-6/IEC 60068-2-6 Operation: 2g
Vibration resistance in acc. with EN 60068-2-6/IEC 60068-2-6 Storage/transport:5g
Shock in acc. with EN 60068-2-27/IEC 60068-2-27 Operation:30g, 11ms, three shocks in each space direction
Shock in acc. with EN 60068-2-27/IEC 60068-2-27 Storage/transport:50g, 11ms, three shocks in each space direction
Continuous shock according to EN60068-2-27/IEC60068-2-27 10g,16ms, 1000shocks, in all space directions
Approvals
For the latest approvals, please visit phoenixcontact.net/products.