MAX1153/MAX1154
Stand-Alone, 10-Channel, 10-Bit System Monitors
with Internal Temperature Sensor and V
DD
Monitor
10 ______________________________________________________________________________________
DGND
a) V
OL
TO V
OH
C
LOAD
= 100pF
100μA
DOUT
DGND
V
DD
b) HIGH-Z TO V
OL
AND V
OH
TO V
OL
C
LOAD
= 100pF
100μA
DOUT
Figure 1. Load Circuits for DOUT Enable Time and SCLK to
DOUT Delay Time
DGND
a) V
OH
TO HIGH-Z
C
LOAD
= 100pF
100μA
DOUT
DGND
V
DD
b) V
OL
TO HIGH-Z
C
LOAD
= 100pF
100μA
DOUT
Figure 2. Load Circuit for DOUT Disable Time
Detailed Description
The MAX1153/MAX1154 are precision-monitoring inte-
grated circuit systems specifically intended for stand-
alone operation. They can monitor diverse types of
inputs, such as those from temperature sensors and
voltage signals from pressure, vibration, and accelera-
tion sensors, and digitize these input signals. The digi-
tal values are then compared to preprogrammed
thresholds and, if the thresholds are exceeded, the
processor is alerted by an interrupt signal. No interac-
tion by the CPU or microcontroller (µC) is required until
one of the programmed limits is exceeded (Figures 3
and 4).
Voltages on all the inputs are converted to 10-bit values
sequentially and stored in the current data registers.
Note that eight of these inputs are external and two are
internal. One of the internal inputs monitors the V
DD
voltage supply, while the other monitors the internal IC
temperature. AIN0 to AIN7 can be configured as either
single ended (default) or differential. In addition, these
inputs can be configured for single-ended or differen-
tial temperature measurements. In the temperature
configuration, the device provides the proper bias nec-
essary to measure temperature with a diode-connected
transistor sensor. The user enables which inputs are
measured (both external and internal) and sets the
delay between each sequence of measurements dur-
ing the initial setup of the device.
The values stored in the current data registers are com-
pared to the user-preprogrammed values in the thresh-
old registers (upper and lower thresholds) and, if
exceeded, activate the interrupt output and generate an
alarm condition. If desired, the device can be pro-
grammed to average the results of many measurements
before comparing to the threshold value. This reduces
the sensitivity to external noise in the measured signal.
In addition, the user can set the number of times the
threshold is exceeded (fault cycles) before generating
an interrupt. This feature reduces falsely triggered
alarms caused by undesired, random spurious impulses.
When the fault cycle criterion is exceeded, an alarm
condition is created. The device writes the fault condi-
tion into the alarm register to indicate the alarmed input
channel.
Converter Operation
The MAX1153/MAX1154 ADCs use a fully differential
successive-approximation register (SAR) conversion
technique and an on-chip track-and-hold (T/H) block to
convert temperature and voltage signals into a 10-bit
digital result. Both single-ended and differential config-
urations are supported with a unipolar signal range for
single-ended mode and bipolar or unipolar ranges for
differential mode. Figure 5 shows the equivalent input
circuit for the MAX1153/MAX1154. Configure the input
channels according to Tables 5 and 6 (see the
Input
Configuration Register
section).
In single-ended mode, the positive input (IN+) is con-
nected to the selected input channel and the negative
input (IN-) is connected to GND. In differential mode,
IN+ and IN- are selected from the following pairs:
AIN0/AIN1, AIN2/AIN3, AIN4/AIN5, and AIN6/AIN7.
Once initiated, voltage conversions require 10.6µs (typ)
to complete.
MAX1153/MAX1154
Stand-Alone, 10-Channel, 10-Bit System Monitors
with Internal Temperature Sensor and V
DD
Monitor
______________________________________________________________________________________ 11
During the acquisition interval, IN+ and IN- charge both
a positive (CHOLDP) and a negative (CHOLDN) sam-
pling capacitor. After completing the acquisition inter-
val, the T/H switches open, storing an accurate sample
of the differential voltage between IN+ and IN-. This
charge is then transferred to the ADC and converted.
Finally, the conversion result is transferred to the cur-
rent data register.
Temperature conversions require 46µs (typ) and mea-
sure the difference between two sequential voltage
measurements (see the
Temperature Measurement
section for a detailed description).
Fully Differential Track/Hold (T/H)
The T/H acquisition interval begins with the rising edge
of CS (for manually triggered conversions) and is inter-
nally timed to 1.5µs (typ). The accuracy of the input sig-
nal sample is a function of the input signal’s source
impedance and the T/H’s capacitance. In order to
achieve adequate settling of the T/H, limit the signal
source impedance to a maximum of 1kΩ.
Input Bandwidth
The ADC’s input tracking circuitry has a 1MHz small-
signal bandwidth. To avoid high-frequency signals
aliasing into the frequency band of interest, anti-alias
prefiltering of the input signals is recommended.
Analog Input Protection
Internal protection diodes, which clamp the analog
inputs to V
DD
and GND, allow the channel input pins to
swing from (V
GND
- 0.3V) to (V
DD
+ 0.3V) without dam-
age. However, for accurate conversions near full scale,
the inputs must not exceed V
DD
by more than 50mV or
be lower than V
GND
by 50mV. If the analog input range
must exceed 50mV beyond the supplies, limit the input
current.
Single Ended/Differential
The MAX1153/MAX1154 use a fully differential ADC for
all conversions. Through the input configuration regis-
ter, the analog inputs can be configured for either dif-
ferential or single-ended conversions. When sampling
signal sources close to the MAX1153/MAX1154, single-
ended conversion is generally sufficient. Single-ended
conversions use only one analog input per signal
source, internally referenced to GND.
MUX
AIN 0
V
DD
AIN 1
AIN 2
AIN 3
AIN 4
AIN 5
AIN 6
AIN 7
TEMP
SENSE
12-BIT
ADC WITH T/H
DIGITAL
COMPARATOR
INPUT REGISTERS 0–10
CURRENT DATA
UPPER THRESHOLD
LOWER THRESHOLD
# FAULT CYCLES
AVERAGE
CONFIGURATION/
STATUS
REGISTERS
SCAN AND
CONVERSION
CONTROL
SERIAL
INTERFACE
DIN
INT
DOUT
SCLK
CS
Figure 3. Simplified Alarm Block Diagram of the MAX1153/MAX1154
MAX1153/MAX1154
Stand-Alone, 10-Channel, 10-Bit System Monitors
with Internal Temperature Sensor and V
DD
Monitor
12 ______________________________________________________________________________________
IS CHANNEL
ENABLED?
SAMPLE
CHANNEL
AVERAGE
CONVERTED
CHANNEL DATA
IS
AVG DATA
> UPPER?
SAME FAULT
AS PREVIOUS?
RESET FAULT
COUNTER
INCREMENT
FAULT COUNTER
IS
AVG DATA
< LOWER?
INCREMENT
CHANNEL
COUNTER
IS
FAULT CNT
>
FAULT REG?
SET ALARM
REGISTER
NO
YES
CONVERT
CHANNEL
YES
NO
YES
NO
NO
NO
YES
YES
Figure 4. Alarm Flowchart

MAX1153BEUE+T

Mfr. #:
Manufacturer:
Maxim Integrated
Description:
Analog to Digital Converters - ADC Stnd-Alne 10Ch 10Bit System Monitor
Lifecycle:
New from this manufacturer.
Delivery:
DHL FedEx Ups TNT EMS
Payment:
T/T Paypal Visa MoneyGram Western Union

Products related to this Datasheet