IB IL TEMP 4/8 RTD ...
7079_en_05 PHOENIX CONTACT 4
3 Technical data
General data
Housing dimensions (width x height x depth) 48.8 mm x 136.8 mm x 71.5 mm (with connectors)
Weight 125 g (without connectors), 190 g (with connectors)
Operating mode Process data mode with 5 words/1 word PCP
Connection method for sensors 2 and 3-wire technology
Ambient temperature (operation) -25°C ... +55°C
Ambient temperature (storage/transport) -25°C ... +85°C
Permissible humidity (operation/storage/transport) 10% ... 95%, according to DIN EN 61131-2
Permissible air pressure (operation/storage/transport) 70 kPa ... 106 kPa (up to 3000 m above sea level)
Degree of protection IP20
Class of protection Class III, EN 61131-2, IEC 61131-2
Connection data for Inline connectors
Connection method Spring-cage terminal blocks
Conductor cross section 0.08 mm
2
... 1.5 mm
2
(solid or stranded), 28 - 16 AWG
Interface
Inline local bus Inline data routing
Transmission speed
IB IL TEMP 4/8 RTD,
IB IL TEMP 4/8 RTD-PAC,
IB IL TEMP 4/8 RTD-PAC/CN
500 kbps
IB IL TEMP 4/8 RTD-2MBD,
IB IL TEMP 4/8 RTD-2MBD-PAC
2 Mbps
Power consumption 500 kbps 2 Mbps
Communications power U
L
7.5 V 7.5 V
Current consumption from U
L
75 mA (typical) 100 mA (typical)
I/O supply voltage U
ANA
24 V DC 24 V DC
Current consumption at U
ANA
28 mA (typical) 41 mA (typical)
Total power consumption 1.24 W (typical) 1.75 W (typical)
Supply of the module electronics and I/O through bus coupler/power terminal
Connection method Potential routing
Analog inputs
Number 8 analog RTD inputs
Description of the input Input for resistive temperature sensors
Connection method Spring-cage connection
Connection method 2, 3-wire (shielded)
Linear resistance range 0 Ω ... 400 Ω, 0 Ω ... 20 kΩ
Sensor types that can be used Pt, Ni, KTY, Cu, linear resistors
Standards for characteristic curves According to DIN EN 60751: 07/1996/
according to SAMA RC 21-4-1966
Measuring principle Successive approximation
Measured value representation 16 bits (15 bits + sign bit)
Conversion time of the A/D converter 5 µs, typical; 10 µs, maximum
Process data update 6 ms (up to 230 ms possible depending on the connection method)
Data formats IB IL, IB ST, S7-compatible
Accuracy 0.06% (typical), 0.25% (maximum)
Accuracy ±0.5°C (typical)
IB IL TEMP 4/8 RTD ...
7079_en_05 PHOENIX CONTACT 5
The data contains the offset error, gain error, and linearity error in its respec-
tive basic setting. The data is related to nominal operation (preferred mount-
ing position, U
S
= 24 V) with pre-set 32-sample filter. Please also observe the
values for temperature drift and the tolerances under EMI.
All errors indicated as a percentage are related to the positive measuring
range final value.
The maximum tolerance values represent the worst case measurement in-
accuracy. They contain the theoretically maximum possible tolerances in the
corresponding measuring ranges. In the same way, the theoretical maximum
possible tolerances of the calibration and test equipment have been taken
into account. This data is valid for at least twelve months.
Additional tolerances influenced by electromagnetic fields
Type of electromagnetic interference Typical deviation of the measuring range final value
Relative for the input area
linear R 0 to 400 Ω
Relative for the input area
linear R 0 to 20 kΩ
Electromagnetic fields;
field strength 10 V/m
according to EN 61000-4-3/IEC 61000-4-3
< ±4.8% < ±0.5%
Conducted interference
Class 3 (test voltage 10 V)
according to EN 61000-4-6/IEC 61000-4-6
< ±3.5% < ±0.3%
Tolerances at T
A
= +25°C
Sensor type Range T
A
= +25°C T
A
= +25°C
Lower limit Upper limit Absolute deviation Relative deviation
Typical Maximum Typical Maximum
Pt100DIN and SAMA
(3-wire connection)
-200°C +850°C ±0.50°C ±2.13°C ±0.06% ±0.25%
Pt100DIN and SAMA
(2-wire connection)
-200°C +850°C ±1.22°C ±5.64°C ±0.14% ±0.66%
Pt10000 DIN and SAMA
(2 and 3-wire connection)
0°C +70°C ±0.60°C ±1.80°C ±0.86% ±2.57%
Pt10000 DIN and SAMA
(2 and 3-wire connection)
-200°C +180°C ±1.24°C ±3.10°C ±0.69% ±1.72%
Rlin400
(3-wire termination)
0 Ω 400 Ω ±0.20 Ω ±0.83 Ω ±0.05% ±0.21%
Rlin400
(2-wire termination)
0 Ω 400 Ω ±0.48 Ω ±2.20 Ω ±0.12% ±0.55%
Rlin20k
(2 and 3-wire termination)
0 Ω 20000 Ω ±150 Ω ±200 Ω ±0.75% ±1.00%
Temperature and drift response
Sensor type Range T
A
= -25°C to +55°C
Lower limit Upper limit Drift (related to the measuring range final value)
Typical Maximum
Pt100 DIN and SAMA -200°C +850°C 60 ppm/K 220 ppm/K
Pt1000 DIN and SAMA -200°C +850°C 150 ppm/K 500 ppm/K
Pt10000 DIN and SAMA -200°C +180°C 390 ppm/K 1200 ppm/K
Rlin400 0 Ω 400 Ω 60 ppm/K 250 ppm/K
Rlin20k 0 Ω 20000 Ω 280 ppm/K 900 ppm/K
Protective equipment
Short-circuit protection for each input Yes
IB IL TEMP 4/8 RTD ...
7079_en_05 PHOENIX CONTACT 6
Electrical isolation
Common potentials
The 24 V main voltage U
M
, 24 V segment voltage U
S
, and GND have the same potential. FE is a separate potential area.
Separate potentials in the IB IL TEMP 4/8 RTD terminal
Test distance Test voltage
7.5 V supply (bus logic)/
±15 V, ±5 V analog supply (analog I/O)
500 V AC, 50 Hz, 1 min.
7.5 V supply (bus logic)/functional earth ground 500 V AC, 50 Hz, 1 min.
±15 V, ±5 V analog supply (analog I/O)/functional earth ground 500 V AC, 50 Hz, 1 min.
Error messages to the higher-level control or computer system
Failure of the internal I/O voltage supply Yes, I/O error message sent to the bus coupler
Failure of or insufficient communications power U
L
Yes, I/O error message sent to the bus coupler
Error messages via process data
Peripheral fault/user error Yes (see page 15)
Mechanical requirements
Vibration, IEC 60068-2-6; EN 60068-2-6 5g
Shock, IEC 60068-2-27; EN 60068-2-27 30g
Approvals
For the latest approvals, please visit www.phoenixcontact.net/catalog.

2878612